An Accidental over Exposure in Mednif Tele-Cobalt Machine in Nepal

  • chaurasia, P.P. (Medical physicist/RSO, Department of Radiation Oncology, B.P.Koirala Memorial Cancer hospital, Bharatpur, Nepal) ;
  • Srivastava, R.P. (Medical Physicist) ;
  • Prasiko, G. (Head Radiation Oncology Department) ;
  • Neupane, B.P. (Radiotherapy Technician)
  • Published : 2002.09.01

Abstract

A radiation incident took place during treatment on MEDNIF Tele cobalt-60 therapy machine in B.P.KOIRALA MEMORIAL CANCER HOSPITAL in Bharatpur, Nepal. This Chinese made machine has activity of 6240 Curies of cobalt -60. This machine has fulfilled safety requirements. ICRP recommendations, safety rules are followed and practiced. The source was struck up during treatment and a technician was exposed to equivalent dose of 13.75 mSv. recorded by Personal film badge. Risks of workers are comparable to other safe industries. All exposures shall be kept as low as reasonably possible. The higher level of safety is achieved only when every one is dedicated to common goal. A lesson is learnt for future. Good practice is essential but not sufficient. A high demand for tele Cobalt therapy convinced management to replace Mednif machine with a new efficient Elite Tele Cobalt theratron Machine.

Keywords