Study on the partially premixed flames produced by a coflow burner as temperature calibration source
동축류 버너에서 생성된 부분 예혼합 화염을 이용한 화염 온도 측정 검정원 연구
- Published : 2000.12.08
Abstract
We investigated a uniform temperature zone, produced by double flame structure of a coflow CH4/air partially premixed flame, to be used as a temperature calibration source for laser diagnostics. A broadband N2 CARS(coherent anti-Stokes Raman spectroscopy) system with a modeless laser was used for temperature measurement. When the stoichiometric ratio was 1.5, we found the uniform temperature zone in radial direction of the flame of which the averaged temperature was 2110 K with standard deviation 24 K. In the stoichiometric ratio range between 2.0 and 2.5, we found very stable temperature-varying zones in vertical direction at the center of the flame. The size of the zone was approximately 15 mm and it covered a temperature range from 300 K to 1900 K. We also suggest that this zone can be used as a calibration source for 2-D PLIF(planar laser induced flurescence) temperature measurement.
Keywords
- calibration source;
- CARS;
- coherent anti-Stokes Raman spectroscopy;
- combustion;
- temperature measurement
- 검정원;
- 코헤런트 반 스톡스-라만 분광법;
- 연소;
- 온도 측정;