This paper described the estimator and eliminator far three kinds of artifacts in electrocardiogram. The most efficient estimation of baseline drift could be obtain in the cubic spline interpolation method with the PQ and TP segment which are considered to be isoelectric, from the experimental results obtained from the applied 4 types of algorithms. The time loss and distortion could be avoided with the aid of detection criteria by checking if baseline drifts exist or not. The AIEF proposed in this paper was verified as having the best removal performance with less distortion in the QRS complex through the comparison of 5 proposed algorithms. furthermore, the AIEF are most suitable far the ECG analyzer which was only needed relatively short time data due to the fast conversion into the stable state. The proposed parabolic filter with 11 points width was identified as having the best performance for the elimination of muscle artifacts. Also we could obtain 99.7% detection accuracy of spike component and minimize the error identifying QRS complex as spike.