Magnetocardiogram Measurement of Laboratory Rat (백서를 이용한 심자도 신호 측정)

  • Kim, I.S.;Ahn, San;Kwon, H.C.;Song, J.H.
    • Progress in Superconductivity
    • /
    • v.11 no.2
    • /
    • pp.147-151
    • /
    • 2010
  • We have developed a high-$T_c$ SQUID magnetocardiogram (MCG) system for small laboratory animals. White noise of the measurement system was about 30 fT/$Hz^{1/2}$ when measured in a magnetically shielded room. We optimized the measurement position to obtain clear MCG wave from rat's small heart by using grid measurements. With the optimization, the MCG signal was successfully detected with the peak amplitude of about 30 pT. We could observe well defined P-, QRS-, and T-waves from the rat MCG. The results suggest that the developed system has a strong potential to monitor the progress of the heart disease model by using a laboratory rat.

Rat Magnetocardiograms Measured by YBCO SQUID Magnetometer (YBCO SQUID를 이용한 실험쥐의 심자도 측정)

  • Ahn, San;Kim, I.S.;Song, J.H.
    • Progress in Superconductivity
    • /
    • v.12 no.2
    • /
    • pp.124-128
    • /
    • 2011
  • We have developed a high-$T_C$ SQUID magnetometer system to measure magnetocardiograms of laboratory rats. White noise of the measurement system was about 50 fT/$Hz^{1/2}$ when measured in a magnetically shielded box. We optimized the measurement position to obtain clear MCG wave from rat's small heart by using grid measurements. With the optimization, the MCG signal was successfully detected with the peak amplitude of about 50 pT. We could observe well defined P-, QRS-, and T-wave from the rat MCG. The results suggest that the developed system has a strong potential to monitor the progress of heart disease model using laboratory rat.