Optimization of Parameters for Simultaneous Multielemental Analysis by Atomic Absorption Spectrometry (원자흡수분광법에 의한 다원소 동시분석시 조건의 최적화)

  • Kim, Hyo Jin;Kang, Jong Seong
    • Analytical Science and Technology
    • /
    • v.6 no.4
    • /
    • pp.359-362
    • /
    • 1993
  • To find compromise experimental conditions that will allow a group of elements to be run on flame atomic absorption spectrometer without changing the burner height or gas flow rates, measurements were carried out. The optimum absorbance for simultaneous analysis of ten elements was observed at high flow rate of air-$C_2H_2$ as fuel and at 2mm of burner height. At the condition, 73% of mean relative absorbance were achieved.

  • PDF

Characterization and Studies of Compromised Experimental Conditions for the Simultaneous Multielemental Analysis by Flame Atomic Absorption Spectrometry (원자흡수법을 이용한 다원소 동시 분석 실험 조건에 관한 기초 특성 연구)

  • 최성규;이현주;박양순;김효진
    • YAKHAK HOEJI
    • /
    • v.39 no.2
    • /
    • pp.148-152
    • /
    • 1995
  • For comparison between the compromised experimental condition and the best conditions for each element by flame atomic absorption spectrometer, the absorbances were measured with changing the burner height and HCL current. The optimum absorbance for simultaneous analysis of 6 elements was observed at 1 mA of HCL current and at 4 mm of burner height except Cr and Co. The stability of flame atomic absorption spectrometer on each element was found to be better than 5%. To investigated the accuracy of this instrument, POSCO RM-07 as a real sample was analysed at these compromised experimental conditions and found to be better than 0.04%.