• Title/Summary/Keyword: kinetic phosphorescence analyzer

Search Result 1, Processing Time 0.014 seconds

A study on the analysis of uranium isotopes in environmental samples using a kinetic phosphorescence analyzer (반응속도론적 인광 분석기를 이용한 환경 시료 중 우라늄 동위원소 분석에 대한 연구)

  • Lee, Myung-Ho;Park, Ye-Eun;Nam, Jong-Soo;Sohn, Se-Chul;Song, Kyu-Seok
    • Analytical Science and Technology
    • /
    • v.23 no.6
    • /
    • pp.518-523
    • /
    • 2010
  • In this paper, the study of analysis of uranium isotopes in environmental samples with a kinetic phosphorescence analyzer (KPA) was described. After leaching uranium fraction from soil and glass material with microwave acid digestion technology, uranium isotopes were purified with UTEVA column, and then measured using KPA. Linearity and repeatability tests for measurement of uranium isotopes were carried out in the uranium standard solution with KPA. The reliability for analytical method of uranium with KPA was validated by its application to uranium standard solution, ground water, IAEA and NIST reference samples.