Evaluation of Setup Error Correction for Patients Using On Board Imager in Image Guided Radiation Therapy (Image Guided Radiation Therapy (IGRT) 시 On Board Imager를 이용한 환자 Setup Error 보정평가)

  • Kang, Soo-Man
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.20 no.2
    • /
    • pp.69-81
    • /
    • 2008
  • Purpose: To reduce side effects in image guided radiation therapy (IGRT) and to improve the quality of life of patients, also to meet accurate SETUP condition for patients, the various SETUP correction conditions were compared and evaluated by using on board imager (OBI) during the SETUP. Materials and Methods: Each 30 cases of the head, the neck, the chest, the belly, and the pelvis in 150 cases of IGRT patients was corrected after confirmation by using OBI at every 2∼3 day. Also, the difference of the SETUP through the skin-marker and the anatomic SETUP through the OBI was evaluated. Results: General SETUP errors (Transverse, Coronal, Sagittal) through the OBI at original SETUP position were Head & Neck: 1.3 mm, Brain: 2 mm, Chest: 3 mm, Abdoman: 3.7 mm, Pelvis: 4 mm. The patients with more that 3 mm in the error range were observed in the correction devices and the patient motions by confirming in treatment room. Moreover, in the case of female patients, the result came from the position of hairs during the Head & Neck, Brain tumor. Therefore, after another SETUP in each cases of over 3 mm in the error range, the treatment was carried out. Mean error values of each parts estimated after the correction were 1 mm for the head, 1.2 mm for the neck, 2.5 mm for the chest, 2.5 mm for the belly, and 2.6 mm for the pelvis. Conclusion: The result showed the correction of SETUP for each treatment through OBI is extremely difficult because of the importance of SETUP in radiation treatment. However, by establishing the average standard of the patients from this research result, the better patient satisfaction and treatment results could be obtained.

  • PDF

Application of Modified Mupit for the Recurrent Vulva Cancer in Brachytherapy

  • Kim, Jong-Sik;Jung, Chun-Young;Oh, Dong-Gyoon;Song, Ki-Won;Park, Young-Hwan
    • 대한방사선치료학회:학술대회논문집
    • /
    • 2005.06a
    • /
    • pp.23-26
    • /
    • 2005
  • Introduction: To evaluate whether modified MUPIT applicator can effectively eradicate recurrent tumor in uterine cervix cancer and reduce rectal complication after complete radiation treatment. Methods and Materials: Modified MUPIT applicator basically consists of an acrylic cylinder with flexible brain applicator , an acrylic template with a predrilled array of holes that serve as guides for interstitial needles and interstitial needles. CT scan was performed to determine tumor volume and the position of interstitial needles. Modified MUPIT applicator was applied to patient in operation room and the accuracy for position of interstitial needles in tumor volume was confirmed by CTscan. Brachytherapy was delivered using modified MUPIT applicator and RALS (192-Ir HDR) after calculated computer planning by orthogonal film. The daily dose was 600cGy and the total dose was delivered 3000cGy in tumor volume by BID. Rectal dose was measured by TLD at 5 points so that evaluated the risk of rectal complication. Result: The application of modified MUPIT applicator improved dramatically dose distributions in tumor volume and follow-up of 3 month for this patient was clinically partial response without normal tissue complication, Rectal dose was measured 34.1cGy, 57.1cGy, 103.8cGy, 162.7cGy, 165.7cGy at each points, especially the rectal dose including previous EBRT and ICR was 34.1cGy, 57.1cGy Conclusion: Patients with locally recurrent tumor in uterine cervix cancer treated with modified MIUPIT applicator can expect reasonable rates of local control. The advantages of the system are the fixed geometry Provided by the template and cylinders, and improved dose distributions in irregular tumor volume without rectal complication

  • PDF