Commissioning Experience of Tri-Cobalt-60 MRI-guided Radiation Therapy System (자기공명영상유도 Co-60 기반 방사선치료기기의 커미셔닝 경험)

  • Park, Jong Min;Park, So-Yeon;Wu, Hong-Gyun;Kim, Jung-in
    • Progress in Medical Physics
    • /
    • v.26 no.4
    • /
    • pp.193-200
    • /
    • 2015
  • The aim of this study is to present commissioning results of the ViewRay system. We verified safety functions of the ViewRay system. For imaging system, we acquired signal to noise ratio (SNR) and image uniformity. In addition, we checked spatial integrity of the image. Couch movement accuracy and coincidence of isocenters (radiation therapy system, imaging system and virtual isocneter) was verified. Accuracy of MLC positioing was checked. We performed reference dosimetry according to American Association of Physicists in Medicine (AAPM) Task Group 51 (TG-51) in water phantom for head 1 and 3. The deviations between measurements and calculation of percent depth dose (PDD) and output factor were evaluated. Finally, we performed gamma evaluations with a total of 8 IMRT plans as an end-to-end (E2E) test of the system. Every safety system of ViewRay operated properly. The values of SNR and Uniformity met the tolerance level. Every point within 10 cm and 17.5 cm radii about the isocenter showed deviations less than 1 mm and 2 mm, respectively. The average couch movement errors in transverse (x), longitudinal (y) and vertical (z) directions were 0.2 mm, 0.1 mm and 0.2 mm, respectively. The deviations between radiation isocenter and virtual isocenter in x, y and z directions were 0 mm, 0 mm and 0.3 mm, respectively. Those between virtual isocenter and imaging isocenter were 0.6 mm, 0.5 mm and 0.2 mm, respectively. The average MLC positioning errors were less than 0.6 mm. The deviations of output, PDDs between mesured vs. BJR supplement 25, PDDs between measured and calculated and output factors of each head were less than 0.5%, 1%, 1% and 2%, respectively. For E2E test, average gamma passing rate with 3%/3 mm criterion was $99.9%{\pm}0.1%$.

Computer Assisted EPID Analysis of Breast Intrafractional and Interfractional Positioning Error (유방암 방사선치료에 있어 치료도중 및 분할치료 간 위치오차에 대한 전자포탈영상의 컴퓨터를 이용한 자동 분석)

  • Sohn Jason W.;Mansur David B.;Monroe James I.;Drzymala Robert E.;Jin Ho-Sang;Suh Tae-Suk;Dempsey James F.;Klein Eric E.
    • Progress in Medical Physics
    • /
    • v.17 no.1
    • /
    • pp.24-31
    • /
    • 2006
  • Automated analysis software was developed to measure the magnitude of the intrafractional and interfractional errors during breast radiation treatments. Error analysis results are important for determining suitable planning target volumes (PTV) prior to Implementing breast-conserving 3-D conformal radiation treatment (CRT). The electrical portal imaging device (EPID) used for this study was a Portal Vision LC250 liquid-filled ionization detector (fast frame-averaging mode, 1.4 frames per second, 256X256 pixels). Twelve patients were imaged for a minimum of 7 treatment days. During each treatment day, an average of 8 to 9 images per field were acquired (dose rate of 400 MU/minute). We developed automated image analysis software to quantitatively analyze 2,931 images (encompassing 720 measurements). Standard deviations ($\sigma$) of intrafractional (breathing motion) and intefractional (setup uncertainty) errors were calculated. The PTV margin to include the clinical target volume (CTV) with 95% confidence level was calculated as $2\;(1.96\;{\sigma})$. To compensate for intra-fractional error (mainly due to breathing motion) the required PTV margin ranged from 2 mm to 4 mm. However, PTV margins compensating for intefractional error ranged from 7 mm to 31 mm. The total average error observed for 12 patients was 17 mm. The intefractional setup error ranged from 2 to 15 times larger than intrafractional errors associated with breathing motion. Prior to 3-D conformal radiation treatment or IMRT breast treatment, the magnitude of setup errors must be measured and properly incorporated into the PTV. To reduce large PTVs for breast IMRT or 3-D CRT, an image-guided system would be extremely valuable, if not required. EPID systems should incorporate automated analysis software as described in this report to process and take advantage of the large numbers of EPID images available for error analysis which will help Individual clinics arrive at an appropriate PTV for their practice. Such systems can also provide valuable patient monitoring information with minimal effort.

  • PDF

Effect of Gamma Ray on Growth of Juvenile Abalone, Haliotis discus hannai (참전복(Haliotis discus hannai)의 성장에 대한 감마선 영향)

  • Kim, Kyung-Ju;Choe, Mi-Kyung;Yeo, In-Kyu
    • Korean Journal of Environmental Biology
    • /
    • v.25 no.2
    • /
    • pp.107-114
    • /
    • 2007
  • Radiation at very low doses frequently has a stimulating or hermetic effect on growth of organism. Effects of growth and survival rate on various dose of gamma irradiation in the farm culture of juvenile abalone (Haliotis discus hannai) were determined in Hallim (Jeju) from February 2004 to January 2005. The initial shell length of abalone juveniles in this study was average $3.45{\pm}0.4cm$. A change of growth after irradiation $(0{\sim}20Gy)$ was observed for 48 weeks. The highest growth rate was observed in 4 Gy-irradiated group and the lowest growth rate was observed in 20 Gy-irradiated group. The additional research about biochemical changes on juvenile abalone after irradiation should be accomplished. Continuous study for gamma radiation-induced hormesis on growth and metabolism of juvenile abalone will further induce the creation of value in ocean industry.

A Study on a KTP Crystal Laser System for a Cancer Using P.D.T. (KTP 크리스탈을 이용한 PDT용 레이저 시스템 개발)

  • Kim, Byoung-Mun;Nam, Hyo-Duk;Kim, Byoung-Chul
    • Proceedings of the Korean Institute of Electrical and Electronic Material Engineers Conference
    • /
    • 2004.07b
    • /
    • pp.631-634
    • /
    • 2004
  • The method that exists in Photodynamic Therapy uses Photosensibility drug strongly Influencing tumour accumulation together with photochemical laser effect and makes the structure of tumour be localized and become extinct. The intracavity transformation of the Nd :YAP main radiation 1079 nm was Raman converted in barium nitrate crystal and the Stokes frequency (1216 nm) was doubled using KTP or RTA crystals. The LiF or Cr:YAG crystals are used for the Q-switch. The radiation Parameters were obtained at 100 Hz pump repetition frequency. The average power at 608 nm radiation with LiF and KTP was 700 mW at multi-mode generation. The 3-6 single 10-15 ns pulses were generated during one cycle of pumping. The doubling efficiency with RTA was two times more than with KTP. The cells of Ehrlich adenocarcinoma (0.1 ml) were i.m. implanted in hind thighs of ICR white non-imbred mice. The cells were preliminarily diluted in medium 199 in the ratio of 1 to 5. HpD was intravenous administered in a dose of 10 mg/kg. The left clean-shaven hind leg was irradiated with laser light 21-27 hours after the administration of the preparation. The right non-Irradiated leg of each animal served as a control. The animals with the transplanted tumor that were not injected with HpD sewed as a control to estimate the complex effect (HpD+ irradiation). Before the administration of HpD and on 3 and 4 days after irradiation the tumor size was measured and the percent of the tumor growth inhibition was calculated. The results of animal treatments has shown high efficiency of PDT method for cancer treatment by means 0.608 m high power pulse solid state laser.

  • PDF

Evaluation of Images Depending on an Attenuation Correction in a Brain PET/CT Scan

  • Choi, Eun-Jin;Jeong, Mon-Taeg;Dong, Kyung-Rae;Kwak, Jong-Gil;Choi, Ji-Won;Ryu, Jae-Kwang
    • Journal of Radiation Industry
    • /
    • v.12 no.4
    • /
    • pp.267-276
    • /
    • 2018
  • A Hoffman 3D Brain Phantom was used to evaluate two PET/CT scanners, BIO_40 and D_690, according to the radiation dose of CT (low, medium and high) at a fixed kilo-voltage-peak (kVp) with the tube current(mA) varied in 17~20 stages(Bio_40 PET/CT scanner: the tube voltage was fixed to 120 kVp, the effective tube current(mAs) was increased from 33 mAs to 190 mAs in 10 mAs increments, D_690 PET/CT scanner: the tube voltage was fixed to 140 kVp, tube current(mA) was increased from 10 mAs to 200 mAs in 10 mAs increments). After obtaining the PET image, an attenuation correction was conducted based on the attenuation map, which led to an analysis of the difference in the image. First, the ratio of white to gray matter for each scanner was examined by comparing the coefficient of variation (CV) depending on the average ratio. In addition, a blind test was carried out to evaluate the image. According to the study results, the BIO_40 and D_690 scanners showed a <1% change in CV value due to the tube current conversion. The change in the coefficients of white and gray matter showed that the Z value was negative for both scanners, indicating that the coefficient of gray matter was higher than that of white matter. Moreover, no difference was observed when the images were compared in a blind test.

A Comparison of Peripheral Doses Scattered from a Physical Wedge and an Enhanced Dynamic Wedge (금속쐐기와 기능강화동적쐐기의 조사야 주변부 선량 비교)

  • Park, Jong-Min;Kim, Hee-Jung;Min, Je-Soon;Lee, Je-Hee;Park, Charn-Il;Ye, Sung-Joon
    • Progress in Medical Physics
    • /
    • v.18 no.3
    • /
    • pp.107-117
    • /
    • 2007
  • In order to evaluate the radio-protective advantage of an enhanced dynamic wedge (EDW) over a physical wedge (PW), we measured peripheral doses scattered from both types of wedges using a 2D array of ion-chambers. A 2D array of ion-chambers was used for this purpose. In order to confirm the accuracy of the device we first compared measured profiles of open fields with the profiles calculated by our commissioned treatment planning system. Then, we measured peripheral doses for the wedge angles of $15^{\circ},\;30^{\circ},\;45^{\circ},\;and\;60^{\circ}$ at source to surface distances (SSD) of 80 cm and 90 cm. The measured points were located at 0.5 cm depth from 1 cm to 5 cm outside of the field edge. In addition, the measurements were repeated by using thermoluminescence dosimeters (TLD). The peripheral doses of EDW were (1.4% to 11.9%) lower than those of PW (2.5% to 12.4%). At 15 MV energy, the average peripheral doses of both wedges were 2.9% higher than those at 6MV energy. At a small SSD (80 cm vs. 90 cm), peripheral dose differences were more recognizable. The average peripheral doses to the heel direction were 0.9% lower than those to the toe direction. The results from the TLD measurements confirmed these findings with similar tendency. Dynamic wedges can reduce unnecessary scattered doses to normal tissues outside of the field edge in many clinical situations. Such an advantage is more profound in the treatment of steeper wedge angles, and shorter SSD.

  • PDF

Synthesis of High Affinity Anion Exchanger Using Ultrafine Fibrous PPmb Nonwoven Fabric by Co60 Irradiation Method (방사선 조사에 의한 초극세 폴리프로필렌 섬유부직포를 이용한 고효율 음이온교환체의 합성)

  • Choi, Kuk-Jong;Lee, Choul-Ho;Hwang, Taek-Sung
    • Polymer(Korea)
    • /
    • v.32 no.6
    • /
    • pp.509-515
    • /
    • 2008
  • The aminated polypropylene melt blown ion exchange fibers were synthesized with acrylic acid monomer onto polypropylene melt blown fibers by radiation-induced polymerization and subsequent amination. Degree of grafting was increased with increasing the acrylic acid monomer concentration and total dose. The highest degree of grafting was obtained 140% at a monomer concentration of 20 v/v% acrylic acid and total dose of 4 kGy. Optimum condition of Mohr's salt was 5.0 $\times10^{-3}$ M. Degree of amination was increased with increasing degree of grafting. Water content was about 1.5 times higher than that of trunk polymer. The maximum ion-exchange capacity was 7.3 meq/g which was 2$\sim$3 times higher than a commercial ion exchange fiber. The average pore size was decreased and BET surface area was increased in order of PPmb, PPmb- g- AAc and APPmb- g- AAc. The average pore size and BET surface area of synthesised fibers were $366.1\;{\AA},\;3.71m^2/g,\;143.3\;{\AA},\;4.94m^2/g,\;40.97\;{\AA},\;8.98m^2/g$, respectively.

An analysis of the concentration of radioactivity of natural radionuclides (238U, 232Th, 40K) and gamma-ray emitting artificial radionuclides(137Cs, 60Co) present in the drinking water of the city of Busan, Republic of Korea, and the calculated absorbed dose of the residents

  • Kim, Chang-Soo;Kim, Jung-Hoon
    • International Journal of Contents
    • /
    • v.8 no.2
    • /
    • pp.60-66
    • /
    • 2012
  • This study was designed to detect and measure the concentration of radioactivity of natural radionuclides ($^{238}U$, $^{232}Th$, $^{40}K$) and artificial radionuclides ($^{137}Cs$, $^{60}Co$) present in the drinking water of the city of Busan and surrounding areas in South Korea, and also to measure the absorbed dose of radiation caused by these elements in the residents so as to help better manage the risk that these radionuclides pose in the future. For the purposes of the study, a total of 42 samples of water were collected from three key water sources (19 samples of groundwater, 4 samples of tap water, and 19 samples of surface water) and their contents were analyzed for radioactivity concentration. The results revealed that two natural radionuclides, $^{238}U$ and $^{232}Th$, exist in the groundwater with an average concentration of radioactivity of 3.34 Bq/L and $8.28{\times}10^{-5}Bq/L$ respectively, while the surface water was found to contain the same two radionuclides with mean concentrations of 0.849 Bq/L and $1.103{\times}10^{-4}Bq/L$ respectively. In addition, of the 19 samples of the groundwater, $^{137}Cs$ was found in eight of them and $^{60}Co$ was detected in ten. Of the four samples of the tap water, $^{137}Cs$ was detected in all samples and $^{60}Co$ was detected in three. Both $^{137}Cs$ and $^{60}Co$ were detected in all 12 samples of surface water. As far as $^{40}K$ is concerned, this element was detected in three of the 19 groundwater samples, but was not detected in any surface or tap water sample. In addition, the absorbed dose of $^{238}U$ from the groundwater was $7.94{\times}10^{-8}Sv/y$, while the absorbed dose of $^{232}Th$ from the surface water was $9.33{\times}10^{-13}Sv/y$. The absorbed dose of $^{137}Cs$ from the tap water was $7.33{\times}10^{-5}Sv/y$, while the absorbed dose of $^{60}Co$ from the surface water was the highest at $4.23{\times}10^{-6}Sv/y$.

Quality Assurance of Volumetric Modulated Arc Therapy Using the Dynalog Files (다이나로그 파일을 이용한 부피세기조절회전치료의 정도관리)

  • Kang, Dong-Jin;Jung, Jae-Yong;Shin, Young-Joo;Min, Jung-Whan;Kim, Yon-Lae;Yang, Hyung-jin
    • Journal of radiological science and technology
    • /
    • v.39 no.4
    • /
    • pp.577-585
    • /
    • 2016
  • The purpose of this study is to evaluate the accuracy of beam delivery QA software using the MLC dynalog file, about the VMAT plan with AAPM TG-119 protocol. The Clinac iX with a built-in 120 MLC was used to acquire the MLC dynalog file be imported in MobiusFx(MFX). To establish VMAT plan, Oncentra RTP system was used target and organ structures were contoured in Im'RT phantom. For evaluation of dose distribution was evaluated by using gamma index, and the point dose was evaluated by using the CC13 ion chamber in Im'RT phantom. For the evaluation of point dose, the mean of relative error between measured and calculated value was $1.41{\pm}0.92%$(Target) and $0.89{\pm}0.86%$(OAR), the confidence limit were 3.21(96.79%, Target) and 2.58(97.42%, OAR). For the evaluation of dose distribution, in case of $Delta^{4PT}$, the average percentage of passing rate were $99.78{\pm}0.2%$(3%/3 mm), $96.86{\pm}1.76%$(2%/2 mm). In case of MFX, the average percentage of passing rate were $99.90{\pm}0.14%$(3%/3 mm), $97.98{\pm}1.97%$(2%/2 mm), the confidence limits(CL) were in case of $Delta^{4PT}$ 0.62(99.38%, 3%/3 mm), 6.6(93.4%, 2%/2 mm), in case of MFX, 0.38(99.62%, 3%/3 mm), 5.88(94.12%, 2%/2 mm). In this study, we performed VMAT QA method using dynamic MLC log file compare to binary diode array chamber. All analyzed results were satisfied with acceptance criteria based on TG-119 protocol.

Inhomogeneity correction in on-line dosimetry using transmission dose (투과선량을 이용한 온라인 선량측정에서 불균질조직에 대한 선량 보정)

  • Wu, Hong-Gyun;Huh, Soon-Nyung;Lee, Hyoung-Koo;Ha, Sung-Whan
    • Journal of Radiation Protection and Research
    • /
    • v.23 no.3
    • /
    • pp.139-147
    • /
    • 1998
  • Purpose: Tissue inhomogeneity such as lung affects tumor dose as well as transmission dose in new concept of on-line dosimetry which estimates tumor dose from transmission dose using the new algorithm. This study was carried out to confirm accuracy of correction by tissue density in tumor dose estimation utilizing transmission dose. Methods: Cork phantom (CP, density $0.202\;gm/cm^3$) having similar density with lung parenchyme and polystyrene phantom (PP, density $1.040\;gm/cm^3$) having similar density with soft tissue were used. Dose measurement was carried out under condition simulating human chest. On simulating AP-PA irradiation, PPs with 3 cm thickness were placed above and below CP, which had thickness of 5, 10, and 20 cm. On simulating lateral irradiation, 6 cm thickness of PP was placed between two 10 cm thickness CPs additional 3 cm thick PP was placed to both lateral sides. 4, 6, and 10 MV x-ray were used. Field size was in the range of $3{\times}3$ cm through $20{\times}20$ cm, and phantom-chamber distance (PCD) was 10 to 50 cm. Above result was compared with another sets of data with equivalent thickness of PP which was corrected by density. Result: When transmission dose of PP was compared with equivalent thickness of CP which was corrected with density, the average error was 0.18 (${\pm}0.27$) % for 4 MV, 0.10 (${\pm}0.43$) % for 6 MV, and 0.33 (${\pm}0.30$) % for 10 MV with CP having thickness of 5 cm. When CP was 10 cm thick, the error was 0.23 (${\pm}0.73$) %, 0.05 (${\pm}0.57$) %, and 0.04 (${\pm}0.40$) %, while for 20 cm, error was 0.55 (${\pm}0.36$) %, 0.34 (${\pm}0.27$) %, and 0.34 (${\pm}0.18$) % for corresponding energy. With lateral irradiation model, difference was 1.15 (${\pm}1.86$) %, 0.90 (${\pm}1.43$) %, and 0.86 (${\pm}1.01$) % for corresponding energy. Relatively large difference was found in case of PCD having value of 10 cm. Omitting PCD with 10 cm, the difference was reduced to 0.47 (${\pm}$1.17) %, 0.42 (${\pm}$0.96) %, and 0.55 (${\pm}$0.77) % for corresponding energy. Conclusion When tissue inhomogeneity such as lung is in tract of x-ray beam, tumor dose could be calculated from transmission dose after correction utilizing tissue density.

  • PDF