• Title/Summary/Keyword: average radiation dose

Search Result 600, Processing Time 0.038 seconds

Changes of Radiation Dose and Image Quality Due to Additional Filtration Material in Computed Radiography (Computed Radiography에서 Additional Filtration Material에 따른 Radiation Dose와 Image Quality의 변화)

  • Kwon, Soon-Mu;Cho, Hyung-Wook;Kang, Yeong-Han;Kim, Boo-Soon;Kim, Jung-Su
    • Journal of radiological science and technology
    • /
    • v.37 no.4
    • /
    • pp.239-246
    • /
    • 2014
  • Filter absorbs low-energy X-ray to increase the average energy and reduces patient exposure dose. This study investigates if the materials of Mo and W could be used for the digital imaging device CR by conducting image assessment and dose measurement of SNR, FOM and histogram. In addition, measurement of beam quality was conducted depending on the material of the filter, and at the same time, a proper combination of filters was examined depending on the change in tube voltage (kVp). In regard to entrance skin dose, Mo filter showed the dose reduction by 42~56%, compared to Cu filter. Moreover, Mo filter showed higher transmission dose by around 1.5 times than that of Cu filter. In image assessment, it was found that W was unsuitable to be used as a filter, whereas Mo could be used as a filter to reduce dose without decline in image quality at the tube voltage of 80 kVp or higher. As tube voltage increased, 2.0 mm Al+0.1 mm Mo almost had a similar histogram width to that of 2.0 mm Al+0.2 mm Cu. Therefore, Mo filter can be used at relatively high tube voltage of 80 kVp, 100 kVp and 120 kVp. The SNR of 2.0 mm Al+0.1 mm Mo did not show any significant difference from those of 2.0 mm Al+0.2 mm Cu and 2.0 mm Al+0.1 mm Cu. As a result, if Mo filter is used to replace Cu filter in general radiography, where 80 kVp or higher is used for digital radiation image, patient exposure dose can be reduced significantly without decline in image quality, compared to Cu filter. Therefore, it is believed that Mo filter can be applied to chest X-ray and high tube voltage X-ray in actual clinical practice.

Independent Verification Program for High-Dose-Rate Brachytherapy Treatment Plans (고선량률 근접치료계획의 정도보증 프로그램)

  • Han Youngyih;Chu Sung Sil;Huh Seung Jae;Suh Chang-Ok
    • Radiation Oncology Journal
    • /
    • v.21 no.3
    • /
    • pp.238-244
    • /
    • 2003
  • Purpose: The Planning of High-Dose-Rate (HDR) brachytherapy treatments are becoming individualized and more dependent on the treatment planning system. Therefore, computer software has been developed to perform independent point dose calculations with the integration of an isodose distribution curve display into the patient anatomy images. Meterials and Methods: As primary input data, the program takes patients'planning data including the source dwell positions, dwell times and the doses at reference points, computed by an HDR treatment planning system (TPS). Dosimetric calculations were peformed in a $10\times12\times10\;Cm^3$ grid space using the Interstitial Collaborative Working Group (ICWG) formalism and an anisotropy table for the HDR Iridium-192 source. The computed doses at the reference points were automatically compared with the relevant results of the TPS. The MR and simulation film images were then imported and the isodose distributions on the axial, sagittal and coronal planes intersecting the point selected by a user were superimposed on the imported images and then displayed. The accuracy of the software was tested in three benchmark plans peformed by Gamma-Med 12i TPS (MDS Nordion, Germany). Nine patients'plans generated by Plato (Nucletron Corporation, The Netherlands) were verified by the developed software. Results: The absolute doses computed by the developed software agreed with the commercial TPS results within an accuracy of $2.8\%$ in the benchmark plans. The isodose distribution plots showed excellent agreements with the exception of the tip legion of the source's longitudinal axis where a slight deviation was observed. In clinical plans, the secondary dose calculations had, on average, about a $3.4\%$ deviation from the TPS plans. Conclusion: The accurate validation of complicate treatment plans is possible with the developed software and the qualify of the HDR treatment plan can be improved with the isodose display integrated into the patient anatomy information.

Comparison of Treatment Planning on Dosimetric Differences Between 192Ir Sources for High-Dose Rate Brachytherapy (고선량률 근접치료에서 이리듐-192 선원의 선량특성 차이에 관한 치료계획 비교)

  • Yang, Oh-Nam;Shin, Seong Soo;Ahn, Woo Sang;Kim, Dae-Yong;Kwon, Kyung-Tae;Lim, Cheong-Hwan;Lee, Sang Ho;Choi, Wonsik
    • Journal of radiological science and technology
    • /
    • v.39 no.2
    • /
    • pp.163-170
    • /
    • 2016
  • To evaluate whether the difference in geometrical characteristics between high-dose-rate (HDR) $^{192}Ir$ sources would influence the dose distributions of intracavitary brachytherapy. Two types of microSelectron HDR $^{192}Ir$ sources (classic and new models) were selected in this study. Two-dimensional (2D) treatment plans for classic and new sources were generated by using PLATO treatment planning system. We compared the point A, point B, and bladder and rectum reference points based on ICRU 38 recommendation. The radial dose function of the new source agrees with that of the classic source except difference of up to 2.6% at the nearest radial distance. The differences of anisotropy functions agree within 2% for r=1, 3, and 5 cm and $20^{\circ}$ < ${\theta}$ < $165^{\circ}$. The largest discrepancies of anisotropy functions reached up to 27% for ${\theta}$ < $20^{\circ}$ at r=0.25 cm and were up to 13%, 10%, and 7% at r=1, 3, and 5 cm for ${\theta}$ > $170^{\circ}$, respectively. There were no significant differences in doses of point A, point B, and bladder point for the treatment plans between the new and classic sources. For the ICRU rectum point, the percent dose difference was on average 0.65% and up to 1.0%. The dose discrepancies between two treatment plans are mainly affected due to the geometrical difference of the source and the sealed capsule.

Daily localization of partial breast irradiation patients with three-dimensional ultrasound imaging

  • Sayan, Mutlay;Vergalasova, Irina;Hard, Daphne;Wrigth, Heather;Archambault, Jessica;Gagne, Havaleh;Nelson, Carl;Heimann, Ruth
    • Radiation Oncology Journal
    • /
    • v.37 no.4
    • /
    • pp.259-264
    • /
    • 2019
  • Purpose: Accurate localization of the lumpectomy cavity during accelerated partial breast radiation (APBR) is essential for daily setup to ensure the prescribed dose encompasses the target and avoids unnecessary irradiation to surrounding normal tissues. Three-dimensional ultrasound (3D-US) allows direct visualization of the lumpectomy cavity without additional radiation exposure. The purpose of this study was to evaluate the feasibility of 3D-US in daily target localization for APBR. Materials and methods: Forty-seven patients with stage I breast cancer who underwent breast conserving surgery were treated with a 2-week course of APBR. Patients with visible lumpectomy cavities on high quality 3D-US images were included in this analysis. Prior to each treatment, X-ray and 3D-US images were acquired and compared to images from simulation to confirm accurate position and determine shifts. Volume change of the lumpectomy cavity was determined daily with 3D-US. Results: A total of 118 images of each modality from 12 eligible patients were analyzed. The average change in cavity volume was 7.8% (range, -24.1% to 14.4%) on 3D-US from simulation to the end-of-treatment. Based on 3D-US, significantly larger shifts were necessary compared to portal films in all three dimensions: anterior/posterior (p = 7E-11), left/right (p = 0.002), and superior/inferior (p = 0.004). Conclusion: Given that the lumpectomy cavity is not directly visible via X-ray images, accurate positioning may not be fully achieved by X-ray images. Therefore, when the lumpectomy cavity is visible on US, 3D-US can be considered as an alternative to X-ray imaging during daily positioning for selected patients treated with APBR, thus avoiding additional exposure to ionizing radiation.

A Method to Calculate a Pass Rate of the ${\gamma}$-index Analysis in Tomotherapy Delivery Quality Assurance (DQA) (단층치료기를 이용한 방사선 치료의 환자별 정도관리 평가를 위한 감마인덱스의 정량화 방법)

  • Park, Dahl;Kim, Yong-Ho;Kim, Won-Taek;Kim, Dong-Won;Kim, Dong-Hyun;Jeon, Ho-Sang;Nam, Ji-Ho;Lim, Sang-Wook
    • Progress in Medical Physics
    • /
    • v.21 no.4
    • /
    • pp.340-347
    • /
    • 2010
  • DQA, a patient specific quality assurance in tomotherapy, is usually performed using an ion chamber and a film. The result of DQA is analysed with the treatment planning system called Tomo Planning Station (TomoPS). The two-dimensional dose distribution of film measurement is compared with the dose distribution calculated by TomoPS using the ${\gamma}$-index analysis. In ${\gamma}$-index analysis, the criteria such as 3%/3 mm is used and we verify that whether the rate of number of points which pass the criteria (pass rate) is within tolerance. TomoPS does not provide any quantitative information regarding the pass rate. In this work, a method to get the pass rate of the ${\gamma}$-index analysis was suggested and a software PassRT which calculates the pass rate was developed. The results of patient specific QA of the intensity modulated radiation therapy measured with I'mRT MatriXX (IBA Dosimetry, Germany) and DQA of tomotherapy measured with film were used to verify the proposed method. The pass rate was calculated using PassRT and compared with the pass rate calculated by OmniPro I'mRT (IBA Dosimetry, Germany). The average difference between the two pass rates was 0.00% for the MatriXX measurement. The standard deviation and the maximum difference were 0.02% and 0.02%, respectively. For the film measurement, average difference, standard deviation and maximum difference were 0.00%, 0.02% and 0.02%, respectively. For regions of interest smaller than $24.3{\times}16.6cm^2$ the proposed method can be used to calculate the pass rate of the gamma index analysis to one decimal place and will be helpful for the more accurate DQA in tomotherapy.

Fluoroscopic the equipment study in accordance with the entrance surface dose study of patients and practitioners (투시 검사 시 장비에 따른 환자와 시술자의 입사표면선량 연구)

  • Yang, Hae-Doo;Hong, Seon-Sook;Seong, Min-Sook;Ha, Dong-Yoon
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.15 no.2
    • /
    • pp.13-18
    • /
    • 2013
  • Purpose : Fluoroscopy equipment, depending on the type of changes that occur in the patient's position ESD and study the patient's scatter ray of ESD Practitioners considered a comparative analysis was to evaluate the correct dose. Materials and Methods : HITACHI four overtube type TU-8000 Flat Detector and Under tube C-Arm Philips' Multi Diagnost Eleva with Flat Detector type were measured by. Each devices is a measure of the patient's esd randophantom position in tabel unfors Xi multi funtion then fixed to the abdomen fluoroscopy and 10 seconds, spot was measured three times, practitioners of the incident surface dose by considering the patient's scatter ray of the table for each device in the average human stomach 21cm thickness acrylic phantom ($25cm{\times}25cm$) Place the practitioner position after position randophantom unfors Xi multi funtion in the thyroid and stomach 1 minute by a fixed one-time fluoroscopy and measured. Results : 10 seconds and the patient perspective of the c-arm ESD 1.2 times smaller on the AP and oblique measurements were measured in the 6-13 times smaller. spot positions to changes in the measured three times on the AP of the abdomen, ESD is 18 times smaller c-arm measurements and the oblique measurement was 19-30 times smaller. And 1 minute at practitioners fluoroscopy esd in the thyroid 2.12 times the c-arm, chest 1.75 times less the dose was measured. On the AP, depending on the device, but the lack of dose difference oblique positions of the two devices depending on changes in the area due to changes in both the AP than on the dose increased, the difference in dose between the two devices, the maximum difference was approximately 27 times. Conclusion : Fluoroscopic equipment at the time of inspection in accordance with changes in dose according to the patient and the patient's positions changes, because the area of the scatter ray considering the change of dose measurements be made, and study of the equipment according to the characteristics of the efficiency and the exposure of the patient and practitioner is considered smooth study equipment manufacturers that can be done is to build the system and think that is also important. Various fluoroscopy when you check future changes in many factors of change in dose for the equipment in the laboratory system by considering the scatter ray radiation shielding for the management to take advantage of reckless undertube have been utilized as more exposure Reduction activities can help is considered as the direction.

  • PDF

Evaluation on Organ Dose and Image Quality by Changing kVp and Ion Chamber Combination while Taking Digital Chest Lateral Decubitus PA Projection (디지털 흉부 측와위 후전방향 검사 시 Ion chamber조합 설정과 관전압 변화에 따른 장기선량 및 화질 평가)

  • Lee, Jin-Soo;Park, Hyong-Hu
    • The Journal of the Korea Contents Association
    • /
    • v.15 no.1
    • /
    • pp.316-323
    • /
    • 2015
  • In this study, we analyzed radiation dose and MTF with setting of Ion chamber and changing kVp so that we are able to suggest acquiring optimized diagnostic images and minimizing patient dose. we assumed right lateral decubitus position among chest decubitus projection and set 7 combination of Ion chamber. By changing kVp(100, 110, 120, 130kVp), we exposed x-ray five times respectively and calculated average value after measuring entrance dose. we input the entrance dose value to PCXMC Monte carlo simulation tool and calculated organ dose and effective dose. Then we did physical image evaluation with MTF for the purpose to compare image quality. As a result, the high kVp, entrance dose is reduced. As change of ion chamber, when selecting second ion chamber, both organ dose and effective dose were the lowest. In contrast, selecting first ion chamber was the highest. MTF is superior to set second Ion chamber and using 120 kVp. Consequently, when taking chest right lateral decubitus using Digital radiography, the optimized combination which have both reducing dose efficiently without declining image quality and aquring good qualified image is set 120 kVp and selecting second Ion chamber.

A Study on Electron Beam Dosimetry for Chest Wall Irradiation (흉곽(胸廓)의 전자선(電子線) 조사시(照射時) 선량분포(線量分布)에 관(關)한 연구(硏究))

  • Kang, Wee Saing;Koh, Kyoung Hwan;Ha, Sung Whan;Park, Charn Il
    • Radiation Oncology Journal
    • /
    • v.1 no.1
    • /
    • pp.41-45
    • /
    • 1983
  • To obtain 7 MeV electron beam which is suitable for treatment of the chest wall after radical of modified radical mastectomy, the authors reduced the energy of electron beam by means by Lucite plate inserted in the beam. To determine the proper thickness of the Lucite plate necessary to reduce the energy of 9 MeV electron beam to 6 MeV, dosimetry was made by using a parallel plate ionization chamber in polystyrene phantom. Separation between two adjacent fields, 7 MeV for chest wall and 12 MeV for internal mammary region, was studied by means of film dosimetry in both polytyrene phantom and Humanoid phantom. The results were as follows. 1. The average energy of 9 MeV electron beam transmitted through the Lucite plate was reduced. Reduction was proportional to the thickness of the Lucite plate in the rate of 1.7 MeV/cm. 2. The proper thickness of the Lucite plate necessary to obtain 6 MeV electron beam from 9 MeV was 1.2 cm. 3. 7 MeV electron beam, 80% dose at 2cm depth, is adequate for treatment of the chest wall. 4. Proper separation between two adjacent electron fields, 7 MeV and 12 MeV, was 5mm on both flat surface and sloping surface to produce uniform dose distribution.

  • PDF

Dosimetric Verification for Primary Focal Hypermetabolism of Nasopharyngeal Carcinoma Patients Treated with Dynamic Intensity-modulated Radiation Therapy

  • Xin, Yong;Wang, Jia-Yang;Li, Liang;Tang, Tian-You;Liu, Gui-Hong;Wang, Jian-She;Xu, Yu-Mei;Chen, Yong;Zhang, Long-Zhen
    • Asian Pacific Journal of Cancer Prevention
    • /
    • v.13 no.3
    • /
    • pp.985-989
    • /
    • 2012
  • Objective: To make sure the feasibility with $^{18F}FDG$ PET/CT to guided dynamic intensity-modulated radiation therapy (IMRT) for nasopharyngeal carcinoma patients, by dosimetric verification before treatment. Methods: Chose 11 patients in III~IVA nasopharyngeal carcinoma treated with functional image-guided IMRT and absolute and relative dosimetric verification by Varian 23EX LA, ionization chamber, 2DICA of I'mRT Matrixx and IBA detachable phantom. Drawing outline and making treatment plan were by different imaging techniques (CT and $^{18F}FDG$ PET/CT). The dose distributions of the various regional were realized by SMART. Results: The absolute mean errors of interest area were $2.39%{\pm}0.66$ using 0.6cc ice chamber. Results using DTA method, the average relative dose measurements within our protocol (3%, 3 mm) were 87.64% at 300 MU/min in all filed. Conclusions: Dosimetric verification before IMRT is obligatory and necessary. Ionization chamber and 2DICA of I'mRT Matrixx was the effective dosimetric verification tool for primary focal hyper metabolism in functional image-guided dynamic IMRT for nasopharyngeal carcinoma. Our preliminary evidence indicates that functional image-guided dynamic IMRT is feasible.

Evaluation on Usefulness of Applying Body-fix to Four Dimensional Radiation Therapy (4차원 방사선 치료시 Body Fix의 유용성 평가)

  • Kim, Young-Jae;Jang, Young-Ill;Ji, Yeon-Sang;Han, Jae-Bok;Choi, Nam-Gil;Jang, Seong-Joo
    • The Journal of the Korea Contents Association
    • /
    • v.13 no.10
    • /
    • pp.419-426
    • /
    • 2013
  • This study investigates the usefulness of body fix in 4DRT on Liver cancer trying to find tumor tissue's volume and located variations, absorbed dose on tumor and normal tissues. Test subjects 23 patients were agreed these test. These patient's have a 4 dimensional CT scan. We make an acquisition on patients CT image by two types -put on the body fix or not-. Average tumor volume reduced by 0.17% on GTV and 3.2% on CTV and PTV. Tumor's variation reduces 29.8%(anterior and posterior, AP) and 5.31% (upper and lower, UL). The absorbed tumor doses under put on the body fix was a little higher(1.3%) than other. Normal tissues'(normal liver, stomach, Rt. kidney, spinal cord) absorbed dose could be reduced approximately 5%. Therefore, using body fix on 4DRT for liver cancer patient is considered effectively.