• Title/Summary/Keyword: acrylic assistant equipment (ACR)

Search Result 1, Processing Time 0.016 seconds

Feasibility Study of Applying the Acrylic Assistant Equipment (ACR) to Reduce Patient's Discomfort in Lower Abdomen MRI Scan (하복부 MRI 검사 시 환자의 불편함을 줄이기 위한 아크릴 보조 장치 사용의 타당성 조사)

  • Park, Eunhye;Lee, Minsik
    • Journal of the Korean Society of Radiology
    • /
    • v.12 no.4
    • /
    • pp.475-480
    • /
    • 2018
  • In lower abdominal MRI scan, patients have been tested by physically contacting with the body array coil. In this study, we have designed the acrylic assistant equipment (ACR) which allows the contactless scan of the patient to the coil and evaluated the feasibility by comparing the acquired images with ACR to those obtained without ACR. We tested 10 cases (F: 5, m: 5) by using the Ingenia $3.0T^{TM}$ MR system and dStreamTM torso coil (Philips Healthcare, Netherlands). We implemented T1 AX TSE and eTHRIVE (GRE) techniques. The scanned images were quantitatively and qualitatively assessed. In qualitatively, the TSE shows 4.44 and 4.56 mean values with and without the ACR and 4.34 and 4.28 at the GRE, respectively. In quantitatively, the TSE shows 12.15 CNR, 17.95 SNR and 12.71 CNR, 18.96 SNR with and without the ACR. And GRE shows 17.72 CNR, 22.59 SNR and 18.26 CNR, 24.47 SNR with and without the ACR, respectively. We have designed and implemented the acrylic assistant equipment to lower abdominal patients. Our data indicate that it is possible to obtain similar image qualities to current lower abdominal MRI scan without the physical contact to the patient.