Sensor Calibration of a Helmet MEG System (헬멧형 뇌자도 장치의 센서 교정)

  • Kwon, H.;Kim, K.;Yu, K.K.;Kim, J.M.;Lee, Y.H.
    • Progress in Superconductivity
    • /
    • v.12 no.1
    • /
    • pp.57-61
    • /
    • 2010
  • We have developed a whole-head MEG system for basic brain research and clinical application. The sensor system consists of a 152 SQUID gradiometer array oriented and located in a suitable way to cover a whole head of the human. The system measures magnetic fields generated by neuronal currents in the brain to get information on the brain activities. For this purpose, the field sensitivity determined by the position, orientation and geometry of the pickup coil as well as amplification factor of the electronic circuits should be known precisely. However, the position and orientation of the pickup coil might be changed from the designed specifications during cool down of the dewar and it is necessary to characterize the field sensitivity. In this study, we made calibration systems to determine the actual position and orientation of the 152 pickup coils and compared the localization results of the N100m source in the auditory cortex.

Characteristics of Superconductive Pb shield for a Whole Head MEG System (헬멧형 뇌자도 장치로의 활용을 위한 Pb 초전도 차폐의 특성)

  • Yu, K.K.;Kim, K.;Lee, Y.H.;Kwon, H.
    • Progress in Superconductivity
    • /
    • v.11 no.1
    • /
    • pp.30-35
    • /
    • 2009
  • We have investigated the characteristics of a superconductive Pb shield for hemispherical shape and plate to improving signal-to-noise ratio(SNR) of biomagnetism. We measured the shielding factor for the position of helmet shape Pb and for changing the distance from Pb surface. To make a uniform magnetic field, a $1.5m{\times}1.5m$ set of the helmholtz coils activated at several frequencies. The shielding factor of hemispherical shape Pb was from 20 to 57 dB and of Pb plate was about $6{\sim}26dB$ as a function of distance from the lead surface. The shielding factor was rapidly reduced as increasing the distance from Pb surface. The white noise of superconductive quantum interference device(SQUID) with a superconductive shield was about $12fT/Hz^{1/2}$ at 1 Hz, $7fT/Hz^{1/2}$ at 100 Hz. The white noise was more increased about two times than conventional SQUID system without Pb shielding. An auditory signal was measured by first order gradiometer and magnetometer with Pb superconductive shield and compared the SNR. The SQUID system with Pb shield had better performance at low frequency noise level.

  • PDF