• Title/Summary/Keyword: X-ray fluorescene spectrometer

Search Result 1, Processing Time 0.013 seconds

X-ray Fluorescence Analysis of Chemical Ingredients for Portland Cement (X-선 형광분석법에 의한 포틀란드 시멘트의 정량분석)

  • 임헌진;백연봉;김도생;윤준수;이경원
    • Journal of the Korean Ceramic Society
    • /
    • v.33 no.8
    • /
    • pp.928-934
    • /
    • 1996
  • Quantitative analysis each portland cement components was conducted by standard calibration method using X-ray fluorescence spectrometer. Standard sample and the unknown sample were prepared by fused cast bead method. In checking the errors of standard cement 227A372 the relative errors of constituents such as CaO, SiO2, Al2O3, Fe2O3, MgO and SrO were less than 1% and the relative errors of constituents such as So3, K2O, TiO2, Na2O P2O5, and Mn2O3 were less than 5% Sample preparation was mainly contributed to the errors. Compared with general wet chemical method the results of X-ray fluorescence analyses were more precise and accurate. Moreover it is possible to analyze precisely a little amount of the constituents such as SrO, Mn2O3 and P2O5 the analyses of which were very difficult using the wet chemical method.

  • PDF