The Patient Specific QA of IMRT and VMAT Through the AAPM Task Group Report 119 (AAPM TG-119 보고서를 통한 세기조절방사선치료(IMRT)와 부피적세기조절회전치료(VMAT)의 치료 전 환자별 정도관리)

  • Kang, Dong-Jin;Jung, Jae-Yong;Kim, Jong-Ha;Park, Seung;Lee, Keun-Sub;Sohn, Seung-Chang;Shin, Young-Joo;Kim, Yon-Lae
    • Journal of radiological science and technology
    • /
    • v.35 no.3
    • /
    • pp.255-263
    • /
    • 2012
  • The aim of this study was to evaluate the patient specific quality assurance (QA) results of intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) through the AAPM Task Group Report 119. Using the treatment planning system, both IMRT and VMAT treatment plans were established. The absolute dose and relative dose for the target and OAR were measured by using an ion chamber and the bi-planar diode array, respectively. The plan evaluation was used by the Dose volume histogram (DVH) and the dose verification was implemented by compare the measured value with the calculated value. For the evaluation of plan, in case of prostate, both IMRT and VMAT were closed the goal of target and OARs. In case of H&N and Multi-target, IMRT was not reached the goal of target, but VMAT was reached the goal of target and OARs. In case of C-shape(easy), both were reached the goal of target and OARs. In case of C-shape(hard), both were reached the goal of target but not reached the goal of OARs. For the evaluation of absolute dose, in case of IMRT, the mean of relative error (%) between measured and calculated value was $1.24{\pm}2.06%$ and $1.4{\pm}2.9%$ for target and OAR, respectively. The confidence limits were 3.65% and 4.39% for target and OAR, respectively. In case of VMAT the mean of relative error was $2.06{\pm}0.64%$ and $2.21{\pm}0.74%$ for target and OAR, respectively. The confidence limits were 4.09% and 3.04% for target and OAR, respectively. For the evaluation of relative dose, in case of IMRT, the average percentage of passing gamma criteria (3mm/3%) were $98.3{\pm}1.5%$ and the confidence limits were 3.78%. In case of VMAT, the average percentage were $98.2{\pm}1.1%$ and the confidence limits were 3.95%. We performed IMRT and VMAT patient specific QA using TG-119 based procedure, all analyzed results were satisfied with acceptance criteria based on TG-119. So, the IMRT and VMAT of our institution was confirmed the accuracy.

Evaluation of Beam-Matching Accuracy for 8 MV Photon Beam between the Same Model Linear Accelerator (동일 기종 선형가속기간 8 MV 광자선에 대한 빔 매칭 정확도 평가)

  • Kim, Yon-Lae;Chung, Jin-Beom;Kang, Seong-Hee
    • Journal of radiological science and technology
    • /
    • v.43 no.2
    • /
    • pp.105-114
    • /
    • 2020
  • This study aimed to assess of beam-matching accuracy for an 8 MV beam between the same model linear accelerators(Linac) commissioned over two years. Two models were got the customer acceptance procedure(CAP) criteria. For commissioning data for beam-matched linacs, the percentage depth doses(PDDs), beam profiles, output factors, multi-leaf collimator(MLC) leaf transmission factors, and the dosimetric leaf gap(DLG) were compared. In addition, the accuracy of beam matching was verified at phantom and patient levels. At phantom level, the point doses specified in TG-53 and TG-119 were compared to evaluate the accuracy of beam modelling. At patient level, the dose volume histogram(DVH) parameters and the delivery accuracy are evaluated on volumetric modulated arc therapy(VMAT) plan for 40 patients that included 20 lung and 20 brain cases. Ionization depth curve and dose profiles obtained in CAP showed a good level for beam matching between both Linacs. The variations in commissioning beam data, such as PDDs, beam profiles, output factors, TF, and DLG were all less than 1%. For the treatment plans of brain tumor and lung cancer, the average and maximum differences in evaluated DVH parameters for the planning target volume(PTV) and the organs at risk(OARs) were within 0.30% and 1.30%. Furthermore, all gamma passing rates for both beam-matched Linacs were higher than 98% for the 2%/2 mm criteria and 99% for the 2%/3 mm criteria. The overall variations in the beam data, as well as tests at phantom and patient levels remains all within the tolerance (1% difference) of clinical acceptability between beam-matched Linacs. Thus, we found an excellent dosimetric agreement to 8 MV beam characteristics for the same model Linacs.

Evaluation of the Jaw-Tracking Technique for Volume-Modulated Radiation Therapy in Brain Cancer and Head and Neck Cancer (뇌암 및 두경부암 체적변조방사선치료시 Jaw-Tracking 기법의 선량학적 유용성 평가)

  • Kim, Hee Sung;Moon, Jae Hee;Kim, Koon Joo;Seo, Jung Min;Lee, Joung Jin;Choi, Jae Hoon;Kim, Sung Ki;Jang, In-Gi
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.30 no.1_2
    • /
    • pp.177-183
    • /
    • 2018
  • Purpose : Volumetric Modulated Arc Therapy(VMAT) has the advantage of uniformly and precisely irradiating the tumor to the shape of the tumor while reducing the risk of radiation damage to normal tissues. such as brain cancer, head and neck cancer and prostate cancer, It is being used for treatment. The purpose of this study is to evaluate the usefulness of the Jaw-Tracking technique(JTT) in VMAT for brain and head and neck cancer. Materials and Methods : We selected eight patients with brain and head and neck cancer(4 Brain, 4 head and neck) who were treated with the VMAT treatment technique. Contouring information of the patient's tumor and normal organ was fused to the Rando phantom using the deformable registration of Velocity(Varian, USA). A treatment plan was developed using the Varian Eclipse(ver 15.5, Varian, USA) with the same patient actual beam parameters except for the use of jaw-tracking. As the evaluation index, the maximum dose and mean dose of target and OAR were compared and a portal dosimetry was performed for the treatment plan verification. Results : When using JTT, the relative dose of OAR decreased by 5.24 % and the maximum dose by 7.05 %, respectively, compared with the Static-Jaw technique(SJT). In the various OARs, the mean dose and maximum dose reduction ranges ranged from 0.01 to 3.16 Gy and from 0.12 to 6.27 Gy, respectively. In the case of the target, the maximum dose of GTV, CTV, PTV decreased by 0.17 %, 0.43 %, and 0.37 % in JTT, and the mean dose decreased by 0.24 %, 0.47 % and 0.47 %, respectively. Gamma analysis The JTT and SJT passing rates were $98{\pm}1.73%$ and $97{\pm}1.83%$ on the basis of 3 % / 3 mm, respectively. Comparing the doses of all OARs applied to the experiment, it was found that the use of JTT resulted in a significant decrease in dose due to additional jaw shielding besides MLC than SJT. Conclusion : In radiation therapy using VMAT treatment plan, we can apply JTT in the case of adjacent tumor and normal organs such as brain cancer and head and neck cancer, and in radiotherapy required large field and high energy caused increase leakage dose through MLC. It is considered that the target dose of PTV can be increased by lowering the dose of normal tissue surrounding the tumor.

  • PDF

Evaluation of the Usefulness for VMAT of multiple brain metastasis using jaw tracking (Jaw tracking을 이용한 다발성 뇌 전이의 용적세기조절회전치료에 대한 유용성 평가)

  • Kim, Tae Won;Yoo, Soon Mi;Jeon, Soo Dong;Yoon, In Ha;Back, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.30 no.1_2
    • /
    • pp.73-81
    • /
    • 2018
  • Purpose : The aims of this study were to compare and assess the effectiveness of Volumetric Modulated Arc Therapy(VMAT) using jaw tracking(JT) and fixed jaw(FJ) in radiation therapy of multiple brain metastasis. Methode and material : Among the patients with Multiple Brain Metastasis treated with jaw tracking, 10 patients with more than 6 tumors and with the size of radiation field $14{\times}14cm^2$ or more were included. Each Treatment plans with jaw tracking(JT) and fixed jaw(FJ) was established with Eclipse (Ver. 13.6 Varian, USA). Gamma Index (3 mm, 3 % confidence interval - 95 %) and maximum dose difference were measured with an electronic portal imaging device(EPID). The $D_{max}$ and $D_{mean}$ of Organ At Risk(OAR) were assessed and compared, and the Conformity Index(CI) and Homogeneity Index(HI) were evaluated. Result : Evaluating jaw tracking(JT) and fixed jaw(FJ) outcomes, in all cases, Gamma Index met the permissible standard of 3 mm, 3 % confidence intervals of 95 %. The maximum dose difference value from the areas with leaf end transmission was measured at a maximum of 98.4 % and an average of 43.6 % in clockwise(CW), and 67.9 % and 41.0 % for each in Counter-Clockwise(CCW). With jaw tracking, the maximum value of $D_{max}$ for each normal organ in OAR decreased in 15.36 %~74.59 % with the average value decreasing in 2.84 %~39.80 %. The maximum value of $D_{mean}$ in OAR decreased in 27.90 %~65.23 %, with the average value decreasing in 7.70 %~41.71 %. No change has been found in Conformity Index and Homogeneity Index values. Conclusion : When Jaw tracking is used in treating patients with multiple brain metastasis with VMAT, the unnecessary exposure due to leakage and transmission of radiation in unspecified areas was reduced, without affecting the dose distribution of the planning target volume(PTV), and the availability of radiation therapy with lower doses in normal organs is expected.

  • PDF

The Evaluation of Hybrid-Volumetric Modulated Arc Therapy for Lung Cancer Radiation Therapy (폐암 방사선 치료 시 Hybrid-Volumetric Modulated Arc Therapy의 유용성 평가)

  • Lee, Geon Ho;Kang, Hyo Seok;Choi, Byoung Joon;Park, Sang Jun;Jung, Da Ee;Lee, Du Sang;Ahn, Min Woo;Jeon, Myeong Soo
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.29 no.2
    • /
    • pp.19-26
    • /
    • 2017
  • Objectives: In the Lung, the VMAT rotates continuously and examines radiation. That increases the low doses to normal lung. Due to that, the incidence of radiation pneumonia among radiation side effects may increase. The cause of radiation pneumonia is the lower dose area of the lungs. The H-VMAT was applied to patients who applied to reduce radiation in the lower doses of the lungs. We wanted to assess the usefulness of the H-VMAT by comparing the radiation doses to the low dose areas of the lungs and the normal organs. Materials and Methods: A total of 26 patients who applied for a H-VMAT procedure were applied to the patient. The prescription dose applied to total dose 44 Gy from 22 divisions. For each patient, a plan was implemented with Conventional RT, VMAT and H-VMAT. Conventional RT was carried out in four to five fields each, considering the size, location, shape, and location of the PTV. In the case of a VMAT plan, the two Half ARC, three Half ARC method and the two Full ARC were planned. The H-VMAT was planned by adding two Static fields in the VMAT, taking into account the dose of the lung and the tolerance dose of the organs. Results: In the NSCLC, the lung doses $V_5$ and $V_{10}$ of the lungs except for the treatment plan volume were the lowest with $55.40{\pm}13.39%$ and $32.05{\pm}11.37%$ of H-VMAT. And, in the SCLC, the lung doses of V5 and V10 were the lowest at $64.32{\pm}16.15%$ and $35.50{\pm}9.91%$, respectively. The spinal dose of VMAT in NSCLC was $21.15{\pm}4.02Gy$, which was 7.94 Gy lower than other treatment methods. The lowest spinal dose was delivered at $19.72{\pm}1.82Gy$ for SCLC. The mean dose delivered to the esophagus was also $17.44{\pm}2.04Gy$ and $17.84{\pm}9.20Gy$ in SCLC and NSCLC, respectively. Conclusion: When comparing the value of the surrounding normal organ dose, the VMAT showed that less doses were transmitted from the heart, esophagus and spinal cord than the rest of the treatment plan. However, it was similar to VMAT in normal organs except for the spinal cord. VMAT has increased doses of some normal organs but did not exceed the tolerance dose. It showed a low value in $V_5$, $V_{10}$. When comparing Conventional RT, VMAT, and H-VMAT, If the dose to the heart, esophagus and spinal cord is lower than the tolerance dose, it is thought to reduce the incidence of radiation pneumonia by applying H-VMAT that show the benefits of low doses of the lungs.

  • PDF

A Pilot Research for Real-Time Specific Patient Quality Assurance Using the Hybrid Optimized Vmat Phantom (Hovp) in Volume Modulated Arc Therapy (체적변조회전치료에서 Hybrid Optimized VMAT Phantom (HOVP)을 이용한 실시간 환자 맞춤형 정도관리를 위한 예비연구)

  • Huh, Hyun-Do;Choi, Sang-Hyoun;Kim, Woo-Chul;Kim, Hun-Jeong;Kim, Kum-Bae;Kim, Seong-Hoon;Cho, Sam-Ju;Min, Chul-Kee;Cho, Kwang-Hwan;Lee, Sang-Hoon;Lee, Suk;Shim, Jang-Bo;Shin, Dong-Oh;Ji, Young-Hoon
    • Progress in Medical Physics
    • /
    • v.22 no.4
    • /
    • pp.206-215
    • /
    • 2011
  • The purpose of this was to investigate the measurement of fluence dose map for the specific patient quality assurance. The measurement of fluence map was performed using 2D matrixx detector. The absorbed dose was measured by a glass detector, Gafchromic film and ion chamber in Hybrid Optimized VMAT Phantom (HOVP). For 2D Matrixx, the results of comparison were average passing rate $85.22%{\pm}1.7$ (RT_Target), $89.96%{\pm}2.15$ (LT_Target) and $95.14%{\pm}1.18$ (G4). The dose difference was $11.72%{\pm}0.531$, $-11.47%{\pm}0.991$, $7.81%{\pm}0.857$, $-4.14%{\pm}0.761$ at the G1, G2, G3, G4. In HOVP, the results of comparison for film were average passing rate (3%, 3 mm) $93.64%{\pm}3.87$, $90.82%{\pm}0.99$. We were measured an absolute dose in steep gradient area G1, G2, G3, G4 using the glass detector. The difference between the measurement and calculation are 8.3% (G1), -5.4% (G2), 6.1% (G3), 7.2% (G4). The using an Ion-chamber were an average relative dose error $-1.02%{\pm}0.222$ (Rt_target), $0.96%{\pm}0.294$ (Lt_target). Though we need a more study using a transmission detector. However, a measurement of real-time fluence map will be predicting a dose for real-time specific patient quality assurance in volume modulated arc therapy.

Comparison of dose-variation in skin due to Set-up error in case of radiation therapy for left breast using Volumetric Modulated Arc Therapy(VMAT) (좌측 유방에 대한 용적 변조 회전 방사선 치료 시 자세 오차로 인한 피부 선량)

  • Kwon, Yongjae;Park, Ryeunghwang;Kim, Seyoung;Jung, Dongmin;Baek, Jonggeol;Cho, Jeonghee
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.33
    • /
    • pp.55-62
    • /
    • 2021
  • Purpose: This study aims to contribute to the reduction of complications of breast cancer radiation therapy by analyzing skin dose differences due to Set-up error. Materials and Method: Pseudo breast was produced using a 3D printer, applied to the phantom, and images were acquired through CT. Treatment plan was carried out that the PTV, which contains 95% of the prescription dose, could be more than 95% of the volume, so that Dmax did not exceed 107% of the prescription dose. The Set-up error was evaluated by applying ±1mm/±3mm/±5mm to the X-axis, Y-axis, and Z-axis. Results: The dose-variation in skin due to Set-up error was approximately 106% to 123% compared to prescription dose, and the highest dose in skin was 49.24 Gy at 5mm Set-up error in the lateral direction of the X-axis. More than 107% of the prescription dose was the widest at 6.87 cc in skin lateral. Conclusions: If a Set-up error occurs during left breast cancer VMAT, a great difference in skin dose was shown in the lateral direction of the X-axis. If more effort is made to align the X-axis of the breast treated during CBCT registration, the dose-variation of skin will be reduced.

Evaluation of the Modified Hybrid-VMAT for multiple bone metastatic cancer (다중표적 뼈 전이암의 하이브리드 세기변조(modified hybrid-VMAT) 방사선치료계획 유용성 평가)

  • Jung, Il Hun;Cho, Yoon Jin;Chang, Won Suk;Kim, Sei Joon;Ha, Jin Sook;Jeon, Mi Jin;Jung, In Ho;Kim, Jong Dea;Shin, Dong Bong;Lee, Ik Jae
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.30 no.1_2
    • /
    • pp.161-167
    • /
    • 2018
  • Purpose : This study evaluates the usefulness of the Modified Hybrid-VMAT scheme with consideration of background radiation when establishing a treatment plan for multiple bone metastatic cancer including multiple tumors on the same axis. Materials and Methods : The subjects of this study consisted of five patients with multiple bone metastatic cancer on the same axis. The planning target volume(PTV) prescription dose was 30 Gy, and the treatment plan was established using Ray Station(Ray station, 5.0.2.35, Sweden). In the treatment plan for each patient, two or more tumors were set as one isocenter. A volumetric modulated arc therapy(VMAT) plan, a hybrid VMAT(h) plan with no consideration of background radiation, and a modified hybrid VMAT(mh) with consideration of background radiation were established. Then, using each dose volume histogram(DVH), the PTV maximum dose($D_{max}$), mean dose($D_{mean}$), conformity index(CI), and homogeneity index(HI) were compared among the plans. In addition, the organ at risk(OAR) of each treatment site was evaluated, and the total MU(Monitor Unit) and treatment time were also analyzed. Results : The PTV $D_{max}$ values of VMAT, VMAT(h) and VMAT(mh) were 3188.33 cGy, 3526 cGy, and 3285.67 cGy, the $D_{mean}$ values were 3081 cGy, 3252 cGy, and 3094 cGy; the CI values were $1.35{\pm}0.19$, $1.43{\pm}0.12$, and $1.30{\pm}0.06$; the HI values were $1.06{\pm}0.01$, $1.14{\pm}0.06$, and $1.09{\pm}0.02$; and the VMAT(h) OAR value was increased 3 %, and VMAT(mh) OAR value was decreased 18 %, respectively. Furthermore, the mean MU values were 904.90, 911.73, and 1202.13, and the mean beam on times were $128.67{\pm}10.97$, $167.33{\pm}7.57$, and $190.33{\pm}4.51$ respectively. Conclusions : Applying Modified Hybrid-VMAT when treating multiple targets can prevent overdose by correcting the overlapping of doses. Furthermore, it is possible to establish a treatment plan that can protect surrounding normal organs more effectively while satisfying the inclusion of PTV dose. Long-term follow-up of many patients is necessary to confirm the clinical efficacy of Modified Hybrid-VMAT.

  • PDF

Inter-fractional Target Displacement in the Prostate Image-Guided Radiotherapy using Cone Beam Computed Tomography (전립선암 영상유도 방사선 치료시 골반내장기의 체적변화에 따른 표적장기의 변화)

  • Dong, Kap Sang;Back, Chang Wook;Jeong, Yun Jeong;Bae, Jae Beom;Choi, Young Eun;Sung, Ki Hoon
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.28 no.2
    • /
    • pp.161-169
    • /
    • 2016
  • Purpose : To quantify the inter-fractional variation in prostate displacement and their dosimetric effects for prostate cancer treatment. Materials and Methods : A total of 176 daily cone-beam CT (CBCT) sets acquired for 6 prostate cancer patients treated with volumetric-modulated arc therapy (VMAT) were retrospectively reviewed. For each patient, the planning CT (pCT) was registered to each daily CBCT by aligning the bony anatomy. The prostate, rectum, and bladder were delineated on daily CBCT, and the contours of these organs in the pCT were copied to the daily CBCT. The concordance of prostate displacement, deformation, and size variation between pCT and daily CBCT was evaluated using the Dice similarity coefficient (DSC). Results : The mean volume of prostate was 37.2 cm3 in the initial pCT, and the variation was around ${\pm}5%$ during the entire course of treatment for all patients. The mean DSC was 89.9%, ranging from 70% to 100% for prostate displacement. Although the volume change of bladder and rectum per treatment fraction did not show any correlation with the value of DSC (r=-0.084, p=0.268 and r=-0.162, p=0.032, respectively), a decrease in the DSC value was observed with increasing volume change of the bladder and rectum (r=-0.230,p=0.049 and r=-0.240,p=0.020, respectively). Conclusion : Consistency of the volume of the bladder and rectum cannot guarantee the accuracy of the treatment. Our results suggest that patient setup with the registration between the pCT and daily CBCT should be considered aligning soft tissue.

  • PDF

A Comparative Study of Radiation Therapy Planning between Volumetric-Modulated Arc Therapy and Three-Dimensional Conformal Radiotherapy in Nasopharyngeal Cancer (비인두암의 방사선치료 시 삼차원입체조형 치료기법과 용적세기조절회전치료기법의 비교연구)

  • Kim, Ji-Sung;Lee, Seok-Ho;Lee, Seung-Heon;Kim, Hye-Young;Choi, Jin-Ho;Lee, Kyu-Chan;Kim, Dong-Young
    • Korean Journal of Head & Neck Oncology
    • /
    • v.26 no.2
    • /
    • pp.171-177
    • /
    • 2010
  • 연구목적 : 비인두암 환자들을 대상으로 방사선치료 시 삼차원입체조형치료기법과 용적세기조절회전치료기법을 비교하고 이하선을 포함한 정상조직 보호에 있어 그 차이점을 알아 보고자 본 연구를 시행하였다. 대상 및 방법 : 비인두암 환자 5명을 대상으로 치료계획용 CT(computed tomography)를 시행 후 삼차원입체조형방사선치료계획 과 용적세기조절회전치료계획을 시행하였다. 이를 바탕으로 얻은 선량분포, conformity index(CI) 그리고 선량체적 히스토그램을 통해 손상위험장기(organ at risk)와 계획용표적체적(planning target volume)을 비교 분석하였다. 결 과 : 분석결과 이하선에 조사되는 평균선량이 용적세기조절회전치료계획에서는 43.9%로 삼차원입체조형치료계획에서의 89.4% 보다 유의하게(p=0.043) 감소하였다. 계획용표적체적 conformity index의 경우 용적세기조절회전치료계획 (CI=1.06)에서 삼차원입체조형치료계획(CI=2.55) 보다 유의하게(p=0.043) 향상된 결과를 보였다. 결 론 : 비인두암 환자에서 용적세기조절회전 치료계획 시 삼차원입체조형치료계획 보다 유의하게 이하선에 평균선량이 줄었고 계획용 표적체적에 대한 conformity도 유의하게 향상되는 결과를 보였다. 본 연구가 적은 수의 환자를 대상으로 하였으나 용적세기조절회전치료기법을 시행 시 구강건조증의 발생을 줄일 수 있을 것으로 기대된다. 향후 더 많은 환자군을 대상으로 한 임상연구가 필요할 것으로 사료된다.