Daily Setup Uncertainties and Organ Motion Based on the Tomoimages in Prostatic Radiotherapy (전립선암 치료 시 Tomoimage에 기초한 Setup 오차에 관한 고찰)

  • Cho, Jeong-Hee;Lee, Sang-Kyu;Kim, Sei-Joon;Na, Soo-Kyung
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.19 no.2
    • /
    • pp.99-106
    • /
    • 2007
  • Purpose: The patient's position and anatomy during the treatment course little bit varies to some extend due to setup uncertainties and organ motions. These factors could affected to not only the dose coverage of the gross tumor but over dosage of normal tissue. Setup uncertainties and organ motions can be minimized by precise patient positioning and rigid immobilization device but some anatomical site such as prostate, the internal organ motion due to physiological processes are challenge. In planning procedure, the clinical target volume is a little bit enlarged to create a planning target volume that accounts for setup uncertainties and organ motion as well. These uncertainties lead to differences between the calculated dose by treatment planning system and the actually delivered dose. The purpose of this study was to evaluate the differences of interfractional displacement of organ and GTV based on the tomoimages. Materials and Methods: Over the course of 3 months, 3 patients, those who has applied rectal balloon, treated for prostatic cancer patient's tomoimage were studied. During the treatment sessions 26 tomoimages per patient, Total 76 tomoimages were collected. Tomoimage had been taken everyday after initial setup with lead marker attached on the patient's skin center to comparing with C-T simulation images. Tomoimage was taken after rectal balloon inflated with 60 cc of air for prostate gland immobilization for daily treatment just before treatment and it was used routinely in each case. The intrarectal balloon was inserted to a depth of 6 cm from the anal verge. MVCT image was taken with 5 mm slice thickness after the intrarectal balloon in place and inflated. For this study, lead balls are used to guide the registration between the MVCT and CT simulation images. There are three image fusion methods in the tomotherapy, bone technique, bone/tissue technique, and full image technique. We used all this 3 methods to analysis the setup errors. Initially, image fusions were based on the visual alignment of lead ball, CT anatomy and CT simulation contours and then the radiation therapist registered the MVCT images with the CT simulation images based on the bone based, rectal balloon based and GTV based respectively and registered image was compared with each others. The average and standard deviation of each X, Y, Z and rotation from the initial planning center was calculated for each patient. The image fusions were based on the visual alignment of lead ball, CT anatomy and CT simulation contours. Results: There was a significant difference in the mean variations of the rectal balloon among the methods. Statistical results based on the bone fusion shows that maximum x-direction shift was 8 mm and 4.2 mm to the y-direction. It was statistically significant (P=<0.0001) in balloon based fusion, maximum X and Y shift was 6 mm, 16mm respectively. One patient's result was more than 16 mm shift and that was derived from the rectal expansions due to the bowl gas and stool. GTV based fusion results ranging from 2.7 to 6.6 mm to the x-direction and 4.3$\sim$7.8 mm to the y-direction respectively. We have checked rotational error in this study but there are no significant differences among fusion methods and the result was 0.37$\pm$0.36 in bone based fusion and 0.34$\pm$0.38 in GTV based fusion.

  • PDF

Present Status and Future Aspects of Radiation Oncology in Korea (방사선 치료의 국내 현황과 미래)

  • Huh, Seung-Jae
    • Radiation Oncology Journal
    • /
    • v.24 no.4
    • /
    • pp.211-216
    • /
    • 2006
  • $\underline{Purpose}$: An analysis of the infrastructure for radiotherapy in Korea was performed to establish a baseline plan in 2006 for future development. $\underline{Materials\;and\;Methods}$: The data were obtained from 61 radiotherapy centers. The survey covered the number of radiotherapy centers, major equipment and personnel. Centers were classified into technical level groups according to the IAEA criteria. $\underline{Results}$: 28,789 new patients were treated with radiation therapy in 2004. There were 104 megavoltage devices in 61 institutions, which included 96 linear accelerators, two Cobalt 60 units, three Tomotherapy units, two Cyberknife units and one proton accelerator in 2006. Thirty-five high dose rate remote after-loading systems and 20 CT-simulators were surveyed. Personnel included 132 radiation oncologists, 50 radiation oncology residents, 64 medical physicists, 130 nurses and 369 radiation therapy technologists. All of the facilities employed treatment-planning computers and simulators, among these thirty-two percent (20 facilities) used a CT-simulator. Sixty-six percent (40 facilities) used a PET/CT scanner, and 35% (22 facilities) had the capacity to implement intensity modulated radiation therapy. Twenty-five facilities (41%) were included in technical level 3 group (having one of intensity modulated radiotherapy, stereotactic radiotherapy or intra-operative radiotherapy system). $\underline{Conclusion}$: Radiation oncology in Korea evolved greatly in both quality and quantity recently and demand for radiotherapy in Korea is increasing steadily. The information in this analysis represents important data to develop the future planning of equipment and human resources.

Methodologic Aspect of LINAC-based Stereotactic Radiosurgery (선형가속기 기반 뇌정위 방사선 수술기법)

  • Choi, Tae Jin
    • Progress in Medical Physics
    • /
    • v.23 no.3
    • /
    • pp.127-137
    • /
    • 2012
  • A conversing beam is firstly designed for radiosurgery by a neurosugern Lars Leksell in 1949 with orthogonal x-rays tube moving through horizontal moving arc to focusing the beam at target center. After 2 decades he composits 201 source of the Co-60 for gamma knife which beams focused at locus. Sveral linac-based stereotactic radiosurgery using the circular collimated beam which size range for 0.4~4.0 cm in a diameter by non-coplanar multiarc have been developed over the decades. The irregular lesions can be treated by superimposing with several spherical shots of radiation over the tumour volume. Linac based techniques include the use of between 4 and 11 non-co-planar arcs and a dynamic rotation technique and use photon beam energies in the range of 6~10 MV. Reviews of the characteristics of several treatment techniques can be found in the literature (Podgorsak 1989, Schell 1991). More in recent, static conformal beams defined by custom shaped collimators or a mini- or micro-multileaf collimator (mMLC) have been used in SRS. Finally, in the last few years, intensity-modulated mMLC SRS has also been introduced. Today, many commercial and in-house SRS programs have also introduced non-invasive immobilization systems include the cyberknife and tomotherapy and proton beam. This document will be compared the characteristics of dose distribution of radiosurgery as introduced gamma knife, BrainLab include photon knife in-house SRS program and cyberknife in currently wide used for a cranial SRS.

Evaluations and Comparisons of Body Surface Doses during Breast Cancer Treatment by Tomotherapy and LINAC Radiotherapy Devices

  • Lee, Hyun-Jik;Bae, Sun-Hyun;Cho, Kwang Hwan;Jeong, Jae-Hong;Kwon, Su-Il;Lee, Kil-Dong
    • Progress in Medical Physics
    • /
    • v.28 no.4
    • /
    • pp.218-225
    • /
    • 2017
  • Effects on skin caused by the dose from linear accelerator (LINAC) opposing portal irradiation and TomoDirect 3-D modeling treatment according to the radiation devices and treatment methods were measured, and a comparative analysis was performed. Two groups of 10 patients each were created and measurements were carried out using an optically stimulated luminescence dosimeter. These patients were already receiving radiation treatment in the hospital. Using the SPSS statistical program, the minimum and maximum average standard deviations of the measured skin dose data were obtained. Two types of treatment method were selected as independent variables; the measured points and total average were the dependent variables. An independent sample T-test was used, and it was checked whether there was a significance probability between the two groups. The average of the measured results for the LINAC opposing portal radiation was 117.7 cGy and PDD 65.39% for the inner breast, 144.7 cGy and PDD 80.39% for the outer breast, 143.2 cGy and PDD 79.56% for the upper breast, 151.4 cGy and PDD 84.11% for the lower breast, 149.6 cGy and PDD 83.11% for the axilla, and 141.32 cGy and PDD 78.51% for the total average. In contrast, for TomoDirect 3-D conformal radiotherapy, the corresponding measurement values were 137.6 cGy and PDD 76.44%, 152.3 cGy and PDD 84.61%, 148.6 cGy and PDD 82.56%, 159.7 cGy and PDD 88.72%, and 148.6 cGy PDD 82.56%, respectively, and the total average was 149.36 cGy and PDD 82.98%. To determine if the difference between the total averages was statistically significant, the independent sample T-test of the SPSS statistical program was used, which indicated that the P-value was P=0.024, which was 0.05 lower than the significance level. Thus, it can be understood that the null hypothesis can be dismissed, and that there was a difference in the averages. In conclusion, even though the treatment dose was similar, there could be a difference in the dose entering the body surface from the radiation treatment plan; however, depending on the properties of the treatment devices, there is a difference in the dose affecting the body surface. Thus, the absorbed dose entering the body surface can be high. During breast cancer radiotherapy, radiation dermatitis occurs in almost all patients. Most patients have a difficult time while undergoing treatment, and therefore, when choosing a radiotherapy treatment method, minimizing radiation dermatitis is an important consideration.

Impact of the Respiratory Motion and Longitudinal Profile on Helical Tomotherapy

  • Park, So Hyun;Choi, Jinhyun;Kim, JinSung;Ahn, Sohyun;Kim, Min Joo;Lee, Ho;Choi, Seo Hee;Park, Kwangwoo
    • Progress in Medical Physics
    • /
    • v.29 no.1
    • /
    • pp.1-7
    • /
    • 2018
  • The $TomoTherapy^{(R)}$ beam-delivery method creates helical beam-junctioning patterns in the dose distribution within the target. In addition, the dose discrepancy results in the particular region where the resonance by pattern of dose delivery occurs owing to the change in the position and shape of internal organs with a patient's respiration during long treatment times. In this study, we evaluated the dose pattern of the longitudinal profile with the change in respiration. The superior-inferior motion signal of the programmable respiratory motion phantom was obtained using AbChes as a four-dimensional computed tomography (4DCT) original moving signal. We delineated virtual targets in the phantom and planned to deliver the prescription dose of 300 cGy using field widths of 1.0 cm, 2.5 cm, and 5.0 cm. An original moving signal was fitted to reflecting the beam delivery time of the $TomoTherapy^{(R)}$. The EBT3 film was inserted into the phantom movement cassette, and static, without the movement and with the original movement, was measured with signal changes of 2.0 s, 4.0 s, and 5.0 s periods, and 2.0 mm and 4.0 mm amplitudes. It was found that a dose fluctuation within ${\pm}4.0%$ occurred in all longitudinal profiles. Compared with the original movement, the region of the gamma index above 1 partially appeared within the target and the border of the target when the period and amplitude were changed. Gamma passing rates were 95.00% or more. However, cases for a 5.0 s period and 4.0 mm amplitude at a field width of 2.5 cm and for 2.0 s and 5.0 s periods at a field width of 5.0 cm have gamma passing rates of 92.73%, 90.31%, 90.31%, and 93.60%. $TomoTherapy^{(R)}$ shows a small difference in dose distribution according to the changes of period and amplitude of respiration. Therefore, to treat a variable respiratory motion region, a margin reflecting the degree of change of respiration signal is required.

Preliminary results of entire pleural intensity-modulated radiotherapy in a neoadjuvant setting for resectable malignant mesothelioma

  • Hong, Ji Hyun;Lee, Hyo Chun;Choi, Kyu Hye;Moon, Seok Whan;Kim, Kyung Soo;Hong, Suk Hee;Hong, Ju-Young;Kim, Yeon-Sil;Multidisciplinary Team of Lung Cancer in Seoul St. Mary's Hospital
    • Radiation Oncology Journal
    • /
    • v.37 no.2
    • /
    • pp.101-109
    • /
    • 2019
  • Purpose: The purpose of this study is to evaluate the safety and efficacy of the multimodality treatment with neoadjuvant intensity-modulated radiotherapy (IMRT) for resectable clinical T1-3N0-1M0 malignant pleural mesothelioma (MPM). Materials and Methods: A total of eleven patients who received neoadjuvant chemotherapy and radiotherapy between March 2016 and June 2018 were reviewed. Patients received 25 Gy in 5 fractions to entire ipsilateral hemithorax with helical tomotherapy. Results: All of patients were men with a median age of 56 years. Epithelioid subtype was found in 10 patients. All patients received neoadjuvant chemotherapy with pemetrexed-cisplatin regimen. Ten patients (90.9%) completed 25 Gy/5 fractions and one (9.0%) completed 20 Gy/4 fractions of radiotherapy. IMRT was well tolerated with only one acute grade 3 radiation pneumonitis. Surgery was performed 1 week (median, 8 days; range, 1 to 15 days) after completing IMRT. Extrapleural pneumonectomy was performed in 4 patients (36.3%), extended pleurectomy/decortication in 2 (18.2%) and pleurectomy/decortications in 5 (63.6%). There was no grade 3+ surgical complication except two deaths after EPP in 1 month. Based on operative findings and pathologic staging, adjuvant chemotherapy was delivered in 7 patients (63.6%), and 2 (18.2%) were decided to add adjuvant radiotherapy. After a median follow-up of 14.6 months (range, 2.8 to 30 months), there were 3 local recurrence (33.3%) and 1 distant metastasis (11.1%). Conclusion: Neoadjuvant entire pleural IMRT can be delivered with a favorable radiation complication. An optimal strategy has to be made in resectable MPM patients who would benefit from neoadjuvant radiation and surgery. Further studies are needed to look at long-term outcomes.

Comparison of Setup Deviations for Two Thermoplastic Immobilization Masks in Glottis Cancer (성문암 세기변조방사선치료에서 두 가지 열가소성 마스크에 대한 환자위치잡이 오차 평가)

  • Jung, Jae Hong
    • Journal of radiological science and technology
    • /
    • v.40 no.1
    • /
    • pp.63-70
    • /
    • 2017
  • The purpose of this study was compare to the patient setup deviation of two different type thermoplastic immobilization masks for glottis cancer in the intensity-modulated radiation therapy (IMRT). A total of 16 glottis cancer cases were divided into two groups based on applied mask type: standard or alternative group. The mean error (M), three-dimensional setup displacement error (3D-error), systematic error (${\Sigma}$), random error (${\sigma}$) were calculated for each group, and also analyzed setup margin (mm). The 3D-errors were $5.2{\pm}1.3mm$ and $5.9{\pm}0.7mm$ for the standard and alternative groups, respectively; the alternative group was 13.6% higher than the standard group. The systematic errors in the roll angle and the x, y, z directions were $0.8^{\circ}$, 1.7 mm, 1.0 mm, and 1.5 mm in the alternative group and $0.8^{\circ}$, 1.1 mm, 1.8 mm, and 2.0 mm in the alternative group. The random errors in the x, y, z directions were 10.9%, 1.7%, and 23.1% lower in the alternative group than in the standard group. However, absolute rotational angle (i.e., roll) in the alternative group was 12.4% higher than in the standard group. For calculated setup margin, the alternative group in x direction was 31.8% lower than in standard group. In contrast, the y and z direction were 52.6% and 21.6% higher than in the standard group. Although using a modified thermoplastic immobilization mask could be affect patient setup deviation in terms of numerical results, various point of view for an immobilization masks has need to research in terms of clinic issue.

Prognostic value of nodal SUVmax of 18F-FDG PET/CT in nasopharyngeal carcinoma treated with intensity-modulated radiotherapy

  • Lee, So Jung;Kay, Chul-Seoung;Kim, Yeon-Sil;Son, Seok Hyun;Kim, Myungsoo;Lee, Sea-Won;Kang, Hye Jin
    • Radiation Oncology Journal
    • /
    • v.35 no.4
    • /
    • pp.306-316
    • /
    • 2017
  • Purpose: To investigate the predictive role of maximum standardized uptake value ($SUV_{max}$) of 2-[$^{18}F$]fluoro-2-deoxy-D-glucose($^{18}F-FDG$) positron emission tomography/computed tomography (PET/CT) in nasopharyngeal cancer patients treated with intensity-modulated radiotherapy (IMRT). Materials and Methods: Between October 2006 and April 2016, 53 patients were treated with IMRT in two institutions and their PET/CT at the time of diagnosis was reviewed. The $SUV_{max}$ of their nasopharyngeal lesions and metastatic lymph nodes (LN) was recorded. IMRT was delivered using helical tomotherapy. All patients except for one were treated with concurrent chemoradiation therapy (CCRT). Correlations between $SUV_{max}$ and patients' survival and recurrence were analyzed. Results: At a median follow-up time of 31.5 months (range, 3.4 to 98.7 months), the 3-year overall survival (OS) and disease-free survival (DFS) rates were 83.2% and 77.5%, respectively. In univariate analysis, patients with a higher nodal pre-treatment $SUV_{max}$ (${\geq}13.4$) demonstrated significantly lower 3-year OS (93.1% vs. 55.5%; p = 0.003), DFS (92.7% vs. 38.5%; p < 0.001), locoregional recurrence-free survival (100% vs. 50.5%; p < 0.001), and distant metastasis-free survival (100% vs. 69.2%; p = 0.004), respectively. In multivariate analysis, high pre-treatment nodal $SUV_{max}$ (${\geq}13.4$) was a negative prognostic factor for OS (hazard ratio [HR], 7.799; 95% confidence interval [CI], 1.506-40.397; p = 0.014) and DFS (HR, 9.392; 95% CI, 1.989-44.339; p = 0.005). Conclusions: High pre-treatment nodal $SUV_{max}$ was an independent prognosticator of survival and disease progression in nasopharyngeal carcinoma patients treated with IMRT in our cohort. Therefore, nodal $SUV_{max}$ may provide important information for identifying patients who require more aggressive treatment.

Analysis on Longitudinal Dose according to Change of Field Width (선속 폭(Field Width) 변화에 따른 종축선량 분석)

  • Jung, Won-Seok;Back, Jong-Geal;Shin, Ryung-Mi;Oh, Byung-Cheon;Jo, Jun-Young;Kim, Gi-Chul;Choi, Tae-Gu
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.23 no.2
    • /
    • pp.109-117
    • /
    • 2011
  • Purpose: To analyze the accuracy of tumor volume dose following field width change, to check the difference of dose change by using self-made moving car, and to evaluate practical delivery tumor dose when tomotherapy in the treatment of organ influenced by breathing. Materials and Methods: By using self-made moving car, the difference of longitudinal movement (0.0 cm, 1.0 cm, 1.5 cm, 2.0 cm) was applied and compared calculated dose with measured dose according to change of field width (1.05 cm, 2.50 cm, 5.02 cm) and apprehended margin of error. Then done comparative analysis in degree of photosensitivity of DQA film measured by using Gafchromic EBT film. Dose profile and Gamma histogram were used to measure degree of photosensitivity of DQA film. Results: When field width were 1.05 cm, 2.50 cm, 5.02 cm, margin of error of dose delivery coefficient was -2.00%, -0.39%, -2.55%. In dose profile of Gafchromic EBT film's analysis, the movement of moving car had greater motion toward longitudinal direction and as field width was narrower, big error increased considerably at high dose part compared to calculated dose. The more field width was narrowed, gamma index had a large considerable influence of moving at gamma histogram. Conclusion: We could check the difference of longitudinal dose of moving organ. In order to small field width and minimize organ moving due to breathing, it is thought to be needed to develop breathing control unit and fixation tool.

  • PDF

Usability Assessment of Plan Transfer between TOMO HAD and Radixact : Planning Study (TOMO HDA와 Radixact 치료 계획 간 Transfer에 대한 유용성 평가 : Planning Study)

  • An, Ye Chan;Kim, Jong Sik;Kwon, Dong Yeol;Kim, Jin Man;Choi, Byeong Ki
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.30 no.1_2
    • /
    • pp.117-128
    • /
    • 2018
  • Purpose : To evaluate the usability of plan transfer between TOMO HD and Radixact, we compared the differences of dose in transferred plans by evaluating the dose of normal organ and target. TOMO HDA and Radixact. The completed plans were transferred each other and we compared the differences of dose by evaluating the DVH of each plans. Materials and Methods : We planned 4 different plans assuming the treatment of 2 cases in Head and Neck Cancer and 2 cases Prostate cancer. Each plan was designed so that 95 % of the prescription dose was irradiated over 99 % of the target volume, and the normal organ constraints dose was based on the SMC tolerance dose protocol. Each plan was transferred to each equipment and DVH(dose volume histogram) analysis of the transferred plans was compared and evaluated. Results : The Mean dose of CTV and GTV was increased and decreased in the transferred plans, but there was no significant differences. The target coverage of CTV and GTV was decreased in all cases of transferred plans from TOMO HAD to Radixact, and the change of CI and HI in CTV was within 0.1. Normal organ dose was increased in most cases when transferring from HAD to Radixact in both treatment plans. Conclusion : According to the results of this experiment, the target coverage was above the standard and the normal organ dose was almost same or decreased when transferring the plans from Radixact to HDA equipment. However the target coverage was reduced when transferring the plans from HDA to Radixact and there was an increase in dose in normal organs that could cause sever side effects such as Optic Chiasm ($D_{max}$1.38 Gy), Bladder ($D_{max}$3.07 Gy), Penile Bulb ($D_{max}$1.14 Gy). Therefore, it is necessary to pay attention to the dose change when transferring the plan and one-time transfer due to equipment inspection will be useful for efficient radiation therapy, but if the transferred treatment plans continue for several consecutive days, the treatment plan should be resumed.

  • PDF