• Title/Summary/Keyword: Stereotactic Radiosugery (SRS)

Search Result 2, Processing Time 0.016 seconds

Evaluation of Dosimetric Characteristics of Small Field in Cone Versus Square Fields Based on Linear Accelerators(LINAC) for Stereotactic Radiosugery(SRS) (선형가속기를 기반으로 한 뇌정위 방사선 수술 시 전용 콘과 정방형 소조사면의 선량 특성에 관한 고찰)

  • Yoon, Joon;Lee, Gui-Won;Park, Byung-Moon
    • Journal of radiological science and technology
    • /
    • v.33 no.1
    • /
    • pp.61-66
    • /
    • 2010
  • In this paper we evaluated small field dose characteristics of exclusive cone fields versus square fields for stereotactic radiosugery (SRS) which is based on linear accelerators (LINAC). For this test, we used a small beam detector (stereotactic fields detector : SFD) with a 6 MV photon beam and a water phantom system (IBA, Germany). Percentage depth dose (PDD) was measured for different field sets (cones : ${\Phi}1\;cm$, ${\Phi}2\;cm$, ${\Phi}3\;cm$ ; square fields : $1{\times}1\;cm^2$, $2{\times}2\;cm^2$, $3{\times}3\;cm^2$) at a source skin distance (SSD) of 100 cm. We measured the point depths at 1.5 cm, 5 cm, 10 cm, 20 cm, and 30 cm. The output factors were measured under the same geometrical conditions of the PDD and normalized at the maximum dose depth. To analyze the penumbra, we measured the dose profile with 95 cm of SSD, 5 cm of depth for each field sizes (${\Phi}1\;cm$, ${\Phi}3\;cm$, $1{\times}1\;cm^2$, and $3{\times}3\;cm^2$) using SFD. We obtained the values for every 1 mm interval in the physical field (90%) and 0.5 mm interval in the penumbra region (20 to 80%). The PDD variation of exclusive cones and square fields were 4.3 to 7.9% lesser than the standard field size ($10{\times}10\;cm^2$. The variation of PDD was reduced while the field size was increased. To compare the beam quality, we analyzed the $PDD_{20,10}$ and the results showed under the 1% of variations for all experiments except for ${\Phi}1\;cm$ cone and $1{\times}1\;cm^2$ fields. Output factors of exclusive cone were increased 3.1~4.6% than the square fields, and the penumbra region of exclusive cone was reduced 20% as compared to the square fields. As the previous researches report, it is very important for SRS and SFD that precise dosimetry in small beam fields. In this paper, we showed the effectiveness of exclusive cone, compared to square field. And we will study on the various detector characteristics for small beam fields.

Evaluation of Detector Dependency on Collimator in SRS: Compared Detectors; CC01, CC13, SFD (뇌정위적 방사선수술 시 콜리메이터 크기 변화에 따른 검출기 의존성 평가)

  • Bae, Yong-Ki;Bang, Dong-Wan;Park, Byung-Moon;Kang, Min-Yeong;Kim, Yeon-Rye
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.20 no.2
    • /
    • pp.109-113
    • /
    • 2008
  • Purpose: To evaluate the detector dependency in the various collimator size for Stereotactic Radiosugery (SRS). Materials and Methods: This study was performed with 6 MV photon beam (Varian 21EX, Varian, US) and the measurement detectors are used by ion chamber CC01, CC13 (Wellhofer, Germany) and stereotactic diode detector (SFD, Wellhofer, Germany). SRS collimator size was used by ${\varphi}$5, 10, 20, 30 mm (Brain Lab, Germany). Percentage depth dose (PDD) was measured at SSD 100 cm and field size 10×10 cm from individual detectors. Ouput factor was measured by using same setup of PDD and with maximum dose depth. Data was normalized at field size $10{\times}10\;cm$. Beam profile was measured at SSD 100 cm in SRS collimator ${\varphi}$10, 30 mm and field $10{\times}10\;cm$ and a comparison of FWHM (full width half maximum), penumbra width (20~80%). Results: The CC13 detector was overestimated 16% than other detectors from the PDD in the 5 mm collimator. Output factors were underestimated CC01 28%, CC13 72% in the 5 mm collimator and CC01 9.6%, CC13 25% in the 10 mm collimator than the SFD. Maximum difference was 3% at the FWHM of the dose profile in the 10 mm collimator and difference of the 30 mm collimator was 0% at the FWHM. Penumbra width was increased CC01 122%, CC13 194% in the 10 mm collimator and CC01 68%, CC13 185% in the 30 mm collimator than the SFD. Conclusion: It is very important for accurate dosimetry to select a detector in small field. The SFD was considered with the most accurate dosimeter for small collimator dosimetry in this study.

  • PDF