• Title/Summary/Keyword: Retro recon

Search Result 1, Processing Time 0.015 seconds

Evaluation of Retro recon for SRS planning correction according to the error of recognize to coordinate (SRS의 좌표 인식 오류 시 Retro recon을 이용한 수정 방법에 관한 평가)

  • Moon, hyeon seok;Jeong, deok yang;Do, gyeong min;Lee, yeong cheol;Kim, sun myung;Kim, young bum
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.28 no.2
    • /
    • pp.101-108
    • /
    • 2016
  • Purpose : The purpose of this study was to evaluate the Retro recon in SRS planning using BranLAB when stereotactic location error occurs by metal artifact. Materials and Methods : By CT simulator, image were acquired from head phantom(CIRS, PTW, USA). To observe stereotactic location recognizing and beam hardening, CT image were approved by SRS planning system(BrainLAB, Feldkirchen, Germany). In addition, we compared acquisition image(1.25mm slice thickness) and Retro recon image(using for 2.5 mm, 5mm slice thickness). To evaluate these three images quality, the test were performed by AAPM phantom study. In patient, it was verified stereotactic location error. Results : All the location recognizing error did not occur in scanned image of phantom. AAPM phantom scan images all showed the same trend. Contrast resolution and Spatial resolution are under 6.4 mm, 1.0 mm. In case of noise and uniformity, under 11, 5 of HU were measured. In patient, the stereotactic location error was not occurred at reconstructive image. Conclusion : For BrainLAB planning, using Retro recon were corrected stereotactic error at beam hardening. Retro recon may be the preferred modality for radiation treatment planning and approving image quality.

  • PDF