• Title/Summary/Keyword: Retinoscopist

Search Result 1, Processing Time 0.103 seconds

Evaluation of scale accuracy and effect of off-the-visual-axis on schematic retinoscope training eye (검영기 훈련 모형안의 굴절상태 눈금과 벗어난 시축의 평가)

  • Ryu, Dong-Kyu
    • Journal of the Korea Academia-Industrial cooperation Society
    • /
    • v.11 no.10
    • /
    • pp.3789-3793
    • /
    • 2010
  • To investigate accuracy of graduated scale of schematic retinoscope training eye(schematic eye) and induced effect when measured at off-axis from visual axis. Two skilled retinoscopist measured refractive power using retinoscope in random order. Seven schematic eyes from a single manufacturer were recruited and set to mark +4.00 to -6 diopter(+4, +3, +2, +1, 0, -1, -2, -3, -4, -5, -6). After introducing +2.00 diopter trial lens(50cm working lens), neutral distance was measured at 180 degree to estimate accuracy of scale, and refractive power measured at 0, 5, 10, 15 and 20 degree off-axis to see if any error was induced. According to the results measured by two specialists, in six of seven schematic eye, scale setting varied (p<0.05) and measured refractive power at 5, 10, 15 and 20 degree off-axis from visual axis were $-0.13{\pm}0.06$, $-0.29{\pm}0.06$, $-0.58{\pm}0.11$, and $-0.83{\pm}0.16$ diopter respectively. In some schematic eye, scale graduated on the schematic eye and scale measured by retinoscopy could be different and if retinoscopy is performed off-axis from visual axis, any measuring error can be caused.