• Title/Summary/Keyword: Resistive electric transfer

Search Result 7, Processing Time 0.024 seconds

Analysis of Output Characteristics and Frequency Variation Design for Personal High Frequency Electrical Stimulation Medical Devices (개인용 고주파 전기자극기의 주파수 가변 설계 및 출력 특성의 분석)

  • Jang, Kyeong-Wook;Leem, Ji-Hyun;Baek, Seung-Myoung;Shon, Jin-Geun
    • The Transactions of the Korean Institute of Electrical Engineers P
    • /
    • v.65 no.1
    • /
    • pp.25-30
    • /
    • 2016
  • In this paper, personal electrical stimulation medcial devices using bioelectrical stimulating was developed. Therapy effect of RET(resistive electric transfer) was more effective than CET(capacitive electric transfer), but CET was adopted because of safety issue. Then, the optimum parameters that may be effective in thin skin or facial wrinkles was set. For example, the frequency of the pulse voltage for stimulation is 1.8[MHz], burst frequency is 7[kHz] or 400[Hz], the development of devices was to have ON/OFF control and frequency control. When burst frequency was adjusted 7[kHz], heat was generated in the electrode. The case of 400[Hz] the heat was little generated. The microcontroller ATmega128-based experimental results show that the proposed personal high frequency electrical stimulation devices can be applied to medical equipment using therapy effect successfully.

Quench Characteristics of Resistive Superconducting Fault Current Limiters (저항형 초전도 한류소자의 퀜치 특성)

  • Kim, Hye-Rim;Hyun, Ok-Bae;Choi, Hyo-Sang;Hwang, Si-Dole;Kim, Sang-Joon
    • 한국초전도학회:학술대회논문집
    • /
    • v.9
    • /
    • pp.214-217
    • /
    • 1999
  • We investigated the quench characteristics of meander line type resistive superconducting fault current limiters based on YBCO thin films grown on 2" diameter LaAlO$_3$ substrates. A gold layer was deposited onto the 0.4 ${\mu}$ m thick YBCO film to disperse the heat generated at hot spots, prior to patterning into 1 mm wide meander lines by photolithography. The limiters were tested with simulated fault currents of various amplitudes. The quench started at 10 A and was completed within 1 msec at the fault current of 65 A$_{peak}$. The dynamic quench characteristics were explained based on the heat conduction within the film and the heat transfer between the film and the surrounding liquid nitrogen. The heat transfer coefficient per unit area was estimated to be 3.0 W/cm$^2$K.

  • PDF

The Effectiveness of 448-kHz Capacitive Resistive Monopolar Radiofrequency for Subcutaneous Fat Reduction in a Porcine Model

  • Kwon, Tae-Rin;Lee, Sung-Eun;Kim, Jong Hwan;Jeon, Yong Jae;Jang, You Na;Yoo, Kwang Ho;Kim, Beom Joon
    • Medical Lasers
    • /
    • v.8 no.2
    • /
    • pp.64-73
    • /
    • 2019
  • Background and Objectives The effectiveness of many physiotherapy modalities in reducing subcutaneous fat has been investigated in numerous previous studies. However, to the best of our knowledge, there have been no attempts to determine the effectiveness of physiotherapy modalities in body contouring. The present report determined the effect of 448-kHz capacitive resistive monopolar radiofrequency (CRMRF) in a porcine model. Materials and Methods This study investigated the effect of selective destruction of the subcutaneous fat layer in abdominal fat tissue using CRMRF. The effects of two types of CRMRF (capacitive electric transfer (CET) and resistive electric transfer (RET)) treatment were evaluated using regular digital photography in addition to thermal imaging evaluation, ultrasound measurement, hematological evaluation, and histologic analyses (H&E (hematoxylin and eosin), Oil red O, and immunohistochemistry staining). Results Preclinical evaluation was performed to obtain the data for comparison of the safety and efficacy of the subcutaneous fat reduction after applying CRMRF using CET and RET. After treatment, the thermal transmission was effective, and a 42-47℃ temperature change was observed in the fat layer while an approximately temperature of 42℃ was confirmed on the skin surface. Moreover, after the application of both types of CRMRF treatment, fibrotic septa were observed in the adipose tissue induced by heat at the treatment sites. TUNEL staining was also performed to confirm the process of apoptosis in the adipocytes. Conclusion These results suggest that both CET and RET for CRMRF treatment are safe and effective for subcutaneous fat reduction in a porcine model.

Fabrication of high-frequency therapy device for deep part and temperature distribution characteristic according to electrode condition of RET (심부투열용 고주파 치료기의 제작과 RET 전극조건에 따른 온도 분포 특성)

  • Jung, Jae-Won;Kim, Beong-Ju;Kim, Ki-Seon
    • Journal of Advanced Engineering and Technology
    • /
    • v.11 no.4
    • /
    • pp.267-271
    • /
    • 2018
  • A high-frequency therapy device with improved output by modifying a high-frequency stimulator was fabricated. The details of the design include generating part design, high-frequency transformer design, large output FET installation, DC voltage input part design and gate input driver design. Based on the real test using the pork meat, the temperature distributions according to the current electric transfer method, penetration depth, electrode diameter size were measured. In the CET method, the penetration depth was 0.5 cm and in the RET method, the penetration depth was 20 cm or more. In addition, it was confirmed that the temperature rise according to the penetration depth in the RET system was substantially constant, and the temperature rise was remarkable as the electrode diameter was small. As a result, it has been confirmed that the high frequency therapy device is highly affected by various conditions of the electrode.

In-Situ Heat Cooling using Thick Graphene and Temperature Monitoring with Single Mask Process

  • Kwack, Kyuhyun;Chun, Kukjin
    • Journal of Sensor Science and Technology
    • /
    • v.24 no.3
    • /
    • pp.155-158
    • /
    • 2015
  • In this paper, in-situ heat cooling with temperature monitoring is reported to solve thermal issues in electric vehicle (EV) batteries. The device consists of a thick graphene cooler on top of the substrate and a platinum-based resistive temperature sensor with an embedded heater above the graphene. The graphene layer is synthesized by using chemical vapor deposition directly on the Ni layer above the Si substrate. The proposed thick graphene heat cooler does not use transfer technology, which involves many process steps and does not provide a high yield. This method also reduces the mechanical damage of the graphene and uses only one photomask. Using this structure, temperature detection and cooling are conducted simultaneously using one device. The temperature coefficient of resistance (TCR) of a $1{\times}1mm^2$ temperature sensor on 1-$\grave{i}m$-thick graphene is $1.573{\times}10^3ppm/^{\circ}C$. The heat source cools down $7.3^{\circ}C$ from $54.4^{\circ}C$ to $47.1^{\circ}C$.

Effects of High-Frequency Treatment using Radiofrequency on Autonomic Nervous System and Pain in Women with Dysmenorrhea

  • Sungeon Park;Seungwon Lee;Inok Kim
    • Physical Therapy Rehabilitation Science
    • /
    • v.11 no.4
    • /
    • pp.493-501
    • /
    • 2022
  • Objective: The purpose of this study is to present basic data for appropriate therapeutic intervention by confirming changes in the autonomic nervous system and pain by applying high-frequency deep diathermy to the lower abdomen in patients with primary dysmenorrhea. Design: A randomized controlled clinical trial. Methods: Thirty-eight women aged 18-50 years who complained of regular menstrual cycles (24-32 days) and primary dysmenorrhea symptoms were randomly assigned to a high-frequency therapy group (5, 7, or 9 mins) and a superficial heat therapy group (20 min). High frequency treatment group: The subject was in a supine position, and radio frequency was applied to the lower abdomen below the umbilicus. The radio frequency therapy device used in this study uses a 300 kHz capacitive electrode and a 500 kHz resistive electric transfer to deliver deep heat. Superficial heat treatment Group: Subjects applied a hot pack to the lower abdomen for 20 minutes while lying on their back. Evaluations were made of Heart rate variability and Visual Analogue Scale. Results: In subjects with menstrual pain, there was a significant difference in pain between the high-frequency therapy group and the superficial heat therapy group (p=0.026). However, there was no significant difference between the autonomic nervous system and the stress resistance (p>0.05). Conclusions: As a result of this study, high-frequencytreatment using radiofrequency was effective in relieving pain because it can penetrate deeper tissues than conventional hot packs using superficial heat. In particular, it was found that the optimum effect was obtained when high frequency was applied forfive-seven minutes.