• Title/Summary/Keyword: Proximal bone res option

Search Result 1, Processing Time 0.014 seconds

A Finite Element Analysis of Stress on the Femoral Stem with Resorption of Proximal Medial Femur after Total Hip Replacement (대퇴골 근위부 골흡수가 인공 고관절 대퇴 stem에 미치는 응력에 관한 연구-FEM을 이용한 분석)

  • 김성곤
    • Journal of Biomedical Engineering Research
    • /
    • v.15 no.2
    • /
    • pp.183-188
    • /
    • 1994
  • In clinical orthopaedics, bone resoption in the cortex is often seen post operatively on X-rays or bone densitometry after total hip replacement (THR) in the form of cortical osteoporosis or atropy. Stress shielding of bone occurs, when a load, normally carried by the bone alone, is shared with an implant as a result, the bone stresses are abnormal and with remodelling analysis this may cause extensive proximal bone resoption, possibly weakening the bone bed to the point of failure. The author made finite element models of the cemented and non-cemented type implanted femoral stem with bone resorption of the proximal medial femur and studied the feed back effect of the various degree of bone resoption to THR system by parametric analysis on the stress of the femoral stem and interface. The results of the present finite element analysis implied that the extent of proximal bone resorption has the effect of more increasing stress on the distal stem tip, cement mantle and interface in both type of femoral stem and this high distal stress possibly can cause the mechanical failure of loosening or failure after THR.