• Title/Summary/Keyword: Nonrestraint method

Search Result 1, Processing Time 0.014 seconds

Implementation of a Respiration Measurement System Based on a Nonrestraint Approach (무구속 방식의 호흡 측정 시스템 구현)

  • Cho, Seok-Hyang;Cho, Seung-Ho
    • Journal of the Korea Society of Computer and Information
    • /
    • v.19 no.11
    • /
    • pp.33-41
    • /
    • 2014
  • In this paper, we implemented a system to measure respiration rate with nonrestraint sensors comfortable for people to do their everyday life. The proposed system consists of a pad covered with a Piezoelectric sensor, a respiration measuring device able to send the signal data after amplifying and filtering the source signals to the viewer, a viewer providing sensor data visualization and implementing the respiration measuring algorithm. The algorithm is based on a breathing cycle with the local peak points extracted from threshold on sensor data. Respiration measurements on 3 subjects were performed by changing moving averages and thresholds. The proposed system showed less than 5% error rate when proper moving averages are N=50~60 and a range of thresholds is 800~1300. The system will contribute to preventing suffocation during sleep for infants and the elderly living alone.