• Title/Summary/Keyword: Multi-sample loading device

Search Result 1, Processing Time 0.018 seconds

Development of Multi-sample Loading Device for TEM Characterization of Hydroxyapatite Nanopowder

  • Lee, Jong-Moon;Kim, Jung-Kyun;Jeong, Jong-Man;Kim, Jin-Gyu;Lee, Eunji;Kim, Youn-Joong
    • Bulletin of the Korean Chemical Society
    • /
    • v.34 no.3
    • /
    • pp.788-792
    • /
    • 2013
  • A shortcoming of using transmission electron microscopy (TEM) for structural analysis via electron diffraction is the relatively large error of the measurements as compared to X-ray diffraction. To reduce these errors, various internal standard methods from earlier studies have been widely used. We developed a new device to facilitate the application of internal standard methods in preparation of TEM grids used for nanopowder analysis. Through the application of a partial mask on the TEM grid, both the internal standards and the research materials can be loaded on the same grid. Through this process, we conducted a TEM analysis that compared synthetic hydroxyapatite (HAp) nanopowder to bone apatite from a bovine femur. We determined that the accuracy of the d-spacing measurements of the HAp and bone powders could be improved to better than 1% after statistical treatments of the experimental data. By applying a quarter mask, we loaded four different nanoparticles on a single TEM grid, with one section designated for the internal standard.