Evaluation of Lead Oxide Dosimeter for Quality Assurance of Electron Beam in Radiotherapy (방사선치료 전자선의 정도관리를 위한 Lead Oxide 선량계 평가)

  • Yang, Seungwoo;Han, Moojae;Park, Sungkwang
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.1
    • /
    • pp.79-83
    • /
    • 2021
  • In radiation therapy, electron beam is often used in the treatment of superficial lesion. Accurate measurements are required because electron beam interacts with them in the beam path and affects dose measurements. However, no research has been conducted on electron beam quality assurance. in this study, PbO-based dosimeter was fabricated as a basic study for electron beam quality assurance. Thus, the reproducibility and linearity of the energy of 6, 9, and 12 MeV were analyzed to evaluate measurement accuracy and precision. Reproducibility measurements show RSD value of 1.024%, 1.019% and 0.890%, respectively, at 6, 9, and 12 MeV. linearity measurements show 0.9999 R2 at 6, 9, and 12 MeV altogether. Both evaluations show that the PbO dosimeter has very good measurement accuracy and precision with excellent results.

Solar Radiation Measurement and Data Quality Management (태양자원 측정 및 데이터 품질관리)

  • Jo, Dok-Ki;Yun, Chang-Yeol;Kim, Kwang-Deuk;Kang, Young-Heak
    • 한국신재생에너지학회:학술대회논문집
    • /
    • 2010.11a
    • /
    • pp.63.2-63.2
    • /
    • 2010
  • Solar radiation data are used in several forms and for a variety of purposes. The most detailed information we have is beam and diffuse solar radiation on a horizontal surface, by hours, which is useful in simulations of solar processes. Daily data are more often available and monthly total solar radiation on horizontal surface can be used in some process design methods. However, as performance is generally not linear with solar radiation, the use of average may lead to serious errors if non-linearities ara not taken into account. It is also possible to reduce radiation data to more manageable forms by statistical methods. The control of the quality of most measurements is relegated to the control of the measuring instruments and measuring processes themselves. An accurate measurement will usually result from the use of a high-quality instrument that has been accurately calibrated and is properly used by a qualified individual.

  • PDF

Study of photon beam quality tool at magnetic field change in bending magnet (편향전자석의 자장변화에 따른 광자선 선질평가에 관한 고찰)

  • Kim, Jeong-Ho;Yoo, Se-Jong;Park, Myeong-Cheol;Bae, Seok-Hwan;Kim, Ki-Jin
    • Journal of the Korea Safety Management & Science
    • /
    • v.16 no.4
    • /
    • pp.433-439
    • /
    • 2014
  • Beam quality is changed about magnetic field of bending magnet. Evaluation of beam quality using PDD(Percentage Depth Dose) at 10cm depth at recommendation of AAPM(America Academy of Pain Medicine). However this evaluation shows fragmentary element. Therefore this study is applied to three value, 10cm divided by 5cm depth PDD, 20cm divided by 10cm depth PDD, 30cm divided by 20cm depth PDD, at change the magnetic field. PDD is measured at magnetic field changed ${\pm}1%$, ${\pm}2%$ at 6MV(Mega Voltage), 10MV photon. The plan technique is 3 portal plan using Core-Plan at human pelvic phantom. Conventional and presented methods are compared at maximum and minimum dose. The presented method increased discernment of relieve the unequal distribution and energy area than conventional method. Henceforth, application of presented method will be considered. Development of energy measurement method and detector miniaturization will be needed about continuous study.

A Study on the Characteristics of Therapy Radiation Detector with Diode (다이오드를 이용한 치료방사선 검출기의 특성에 관한 연구)

  • 이동훈;지영훈
    • Journal of Biomedical Engineering Research
    • /
    • v.16 no.2
    • /
    • pp.129-138
    • /
    • 1995
  • High-energy and high-dose X-ray and electron beam have been used in radiation therapy after developing particle accelerators. It is recommended to irradiate patients exect real dose for improving therapy effectiveness by International Committee on Radiation Units and Measurement. The radiation detector for daily beam checks of medical accelerators is described. Using thirteen silicon diodes, we have designed the diode detector providing information about calibration, beam symmetry, flatness, stability variation according to radiation damage, time and general quality assurance for both photon and eletron beams. we also compared these measurement values with those of using ionization chamber, film and semiconductor dosimeter.

  • PDF

Reference X-ray Irradiation System for Personal Dosimeter Testing and Calibration of Radiation Detector

  • Lee, Seung Kyu;Chang, Insu;Kim, Sang In;Lee, Jungil;Kim, Hyoungtaek;Kim, Jang-Lyul;Kim, Min Chae
    • Journal of Radiation Protection and Research
    • /
    • v.44 no.2
    • /
    • pp.72-78
    • /
    • 2019
  • Background: In the calibration and testing laboratory of Korea Atomic Energy Research Institute, the old X-ray generator used for the production of reference X-ray fields was replaced with a new one. For this newly installed X-ray irradiation system, beam alignment as well as the verification of beam qualities was conducted. Materials and Methods: The existing X-ray generator, Phillips MG325, was replaced with YXLON Y.TU 320-D03 in order to generate reference X-ray fields. Theoretical calculations and Monte Carlo simulations were used to determine initial filter thickness. Beam alignment was performed in three steps to deliver a homogeneous radiation dosage to the target at different distances. Finally, the half-value layers were measured for different X-ray fields to verify beam qualities by using an ion chamber. Results and Discussion: Beam alignment was performed in three steps, and collimators and other components were arranged to maintain the uniformity of the mean air kerma rate within ${\pm}2.5%$ at the effective beam diameter of 28 cm. The beam quality was verified by using half-value layer measurement methods specified by American National Standard Institute (ANSI) N13.11-2009 and International Organization for Standardization (ISO)-4037. For each of the nine beams than can be generated by the new X-ray irradiation system, air kerma rates for X-ray fields of different beam qualifies were measured. The results showed that each air kerma rate and homogeneity coefficient of the first and second half-value layers were within ${\pm}5%$ of the recommended values in the standard documents. Conclusion: The results showed that the new X-ray irradiation system provides beam qualities that are as high as moderate beam qualities offered by National Institute of Standards and Technology in ANSI N13.11-2009 and those for narrow-spectrum series of ISO-4037.

Evaluations of a Commercial CLEANBOLUS-WHITE for Clinical Application

  • Geum Bong Yu;Jung-in Kim;Jaeman Son
    • Progress in Medical Physics
    • /
    • v.35 no.1
    • /
    • pp.10-15
    • /
    • 2024
  • Purpose: This study aimed to comprehensively investigate the diverse characteristics of a novel commercial bolus, CLEANBOLUS-WHITE (CBW), to ascertain its suitability for clinical application. Methods: The evaluation of CBW encompassed both physical and biological assessments. Physical parameters such as mass density and shore hardness were measured alongside analyses of element composition. Biological evaluations included assessments for skin irritation and cytotoxicity. Dosimetric properties were examined by calculating surface dose and beam quality using a treatment planning system (TPS). Additionally, doses were measured at maximum and reference depths, and the results were compared with those obtained using a solid water phantom. The effect of air gap on dose measurement was also investigated by comparing measured doses on the RANDO phantom, under the bolus, with doses calculated from the TPS. Results: Biological evaluation confirmed that CBW is non-cytotoxic, nonirritant, and non-sensitizing. The bolus exhibited a mass density of 1.02 g/cm3 and 14 shore 00. Dosimetric evaluations revealed that using the 0.5 cm CBW resulted in less than a 1% difference compared to using the solid water phantom. Furthermore, beam quality calculations in the TPS indicated increased surface dose with the bolus. The air gap effect on dose measurement was deemed negligible, with a difference of approximately 1% between calculated and measured doses, aligning with measurement uncertainty. Conclusions: CBW demonstrates outstanding properties for clinical utilization. The dosimetric evaluation underscores a strong agreement between calculated and measured doses, validating its reliability in both planning and clinical settings.

Development of Diode Based High Energy X-ray Spatial Dose Distribution Measuring Device

  • Lee, Jeonghee;Kim, Ikhyun;Park, Jong-Won;Lim, Yong-Kon;Moon, Myungkook;Lee, Sangheon;Lim, Chang Hwy
    • Journal of Radiation Protection and Research
    • /
    • v.43 no.3
    • /
    • pp.97-106
    • /
    • 2018
  • Background: A cargo container scanner using a high-energy X-ray generates a fan beam X-ray to acquire a transmitted image. Because the generated X-rays by LINAC may affect the image quality and radiation protection of the system, it is necessary to acquire accurate information about the generated X-ray beam distribution. In this paper, a diode-based multi-channel spatial dose measuring device for measuring the X-ray dose distribution developed for measuring the high energy X-ray beam distribution of the container scanner is described. Materials and Methods: The developed high-energy X-ray spatial dose distribution measuring device can measure the spatial distribution of X-rays using 128 diode-based X-ray sensors. And precise measurement of the beam distribution is possible through automatic positioning in the vertical and horizontal directions. The response characteristics of the measurement system were evaluated by comparing the signal gain difference of each pixel, response linearity according to X-ray incident dose change, evaluation of resolution, and measurement of two-dimensional spatial beam distribution. Results and Discussion: As a result, it was found that the difference between the maximum value and the minimum value of the response signal according to the incident position showed a difference of about 10%, and the response signal was linearly increased. And it has been confirmed that high-resolution and two-dimensional measurements are possible. Conclusion: The developed X-ray spatial dose measuring device was evaluated as suitable for dose measurement of high energy X-ray through confirmation of linearity of response signal, spatial uniformity, high resolution measuring ability and ability to measure spatial dose. We will perform precise measurement of the X-ray beamline in the container scanning system using the X-ray spatial dose distribution measuring device developed through this research.

Reliability estimation about quality assurance method of radiotherapy planning (방사선치료계획 정도관리 방법에 따른 신뢰도 평가)

  • Kim, Jeong-Ho;Kim, Gha-Jung;Yoo, Se-Jong;Kim, Ki-Jin
    • Journal of the Korea Safety Management & Science
    • /
    • v.17 no.1
    • /
    • pp.119-124
    • /
    • 2015
  • According as radiation therapy technique develops, standardization of radiation therapy has been complicated by the plan QA(Quality Assurance). However, plan QA tools are two type, OADT (opposite accumulation dose tool) and 3DADT (3 dimensional accumulation dose tool). OADT is not applied to evaluation of beam path. Therefore tolerance error of beam path will establish measurement value at OADT. Plan is six beam path, five irradiation field at each beam path. And beam path error is 0 degree, 0.2 degree, 0.4 degree, 0.6 degree, 0.6 degree, 0.8 degree. Plan QA accomplishes at OADT, 3DADT. The more path error increases, the more plan QA error increases. Tolerance error of OADT path is 0.357 using tolerance error of conventional plan QA. Henceforth plan QA using OADT will include beam path error. In addition, It will increase reliability through precise and various plan technique.

Uncertainty Assessment: Relative versus Absolute Point Dose Measurement for Patient Specific Quality Assurance in EBRT

  • Mahmood, Talat;Ibrahim, Mounir;Aqeel, Muhammad
    • Progress in Medical Physics
    • /
    • v.28 no.3
    • /
    • pp.111-121
    • /
    • 2017
  • Verification of dose distribution is an essential part of ensuring the treatment planning system's (TPS) calculated dose will achieve the desired outcome in radiation therapy. Each measurement have uncertainty associated with it. It is desirable to reduce the measurement uncertainty. A best approach is to reduce the uncertainty associated with each step of the process to keep the total uncertainty under acceptable limits. Point dose patient specific quality assurance (QA) is recommended by American Association of Medical Physicists (AAPM) and European Society for Radiotherapy and Oncology (ESTRO) for all the complex radiation therapy treatment techniques. Relative and absolute point dose measurement methods are used to verify the TPS computed dose. Relative and absolute point dose measurement techniques have a number of steps to measure the point dose which includes chamber cross calibration, electrometer reading, chamber calibration coefficient, beam quality correction factor, reference conditions, influences quantities, machine stability, nominal calibration factor (for relative method) and absolute dose calibration of machine. Keeping these parameters in mind, the estimated relative percentage uncertainty associated with the absolute point dose measurement is 2.1% (k=1). On the other hand, the relative percentage uncertainty associated with the relative point dose verification method is estimated to 1.0% (k=1). To compare both point dose measurement methods, 13 head and neck (H&N) IMRT patients were selected. A point dose for each patient was measured with both methods. The average percentage difference between TPS computed dose and measured absolute relative point dose was 1.4% and 1% respectively. The results of this comparative study show that while choosing the relative or absolute point dose measurement technique, both techniques can produce similar results for H&N IMRT treatment plans. There is no statistically significant difference between both point dose verification methods based upon the t-test for comparing two means.

Radiochromic film dosimetry for linac-based stereotactic radiosurgery

  • Han, Seung-Hee;Park, Suk-Won;Oh, Do-Hoon
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2002.09a
    • /
    • pp.302-304
    • /
    • 2002
  • In linac-based stereotactic radiosurgery, assuring the quality of the planning and delivery of external photon beam requires accurate evaluation of beam parameters, usually including output factors, tissue-phantom ratio and off-axis ratios, and measurement of actual dose distributions from simulated treatment. We're going to test the use of calibrated radio chromic film (Gafchromic film; type MD-55, Nuclear associate) using a Lumiscan 75 digitizer to measure absolute dose and relative dose distributions for linac-based radiosurgery unit Relative dose distribution of a human-style spherical acryl phantom were measured using radiochromic film and calculated by treatment planning system. The absolute dose at the sphere center was measured by radiochromic film and micro chamber (Exradin A-14, 0.009cc). What we want to demonstrate in this work, the 'well selected' radiochromic films when external photon beam are used in linac-based stereotactic radiosurgery are very accurate detector for dosimetry.