Multichannel Liquid Phase Microextraction System (다채널 액상 미세 추출 시스템 설계 및 제작)

  • Zhang, XinJie;Cheng, Shuo;Piao, Xiang Fan
    • Journal of the Korean Society of Manufacturing Process Engineers
    • /
    • v.19 no.10
    • /
    • pp.1-7
    • /
    • 2020
  • In this study, a multichannel gas-liquid microextraction system is designed by integrating the automatic elution of extraction line and multichannel gas-purging liquid phase microextraction. The system uses an injection pump and inert gas to push the extraction solvent to a sample bottle of a gas-phase color autosampler and then implements multichannel gas-liquid microextraction and gas chromatography-mass spectrometry. The system also employs a three-way integrated micro-high-temperature heater, syringe pump, and microflow controller to realize the simultaneous processing of multiple groups of samples, thus improving the sample pretreatment speed and reproducibility and reducing human error. Autoinjection experiments were implemented with polycyclic aromatic hydrocarbon standard samples. The experiments show that the average recovery rate of the system exceeds 70%, and the relative standard among the channels is less than 15%.