• Title/Summary/Keyword: Layered Rescanning

Search Result 1, Processing Time 0.18 seconds

Comparison of Doses of Single Scan PBS and Layered Rescanning PBS Using Moving Phantom in Proton Therapy (양성자 치료에서 Moving Phantom을 이용한 Single Scan PBS와 Layered Rescanning PBS의 선량비교)

  • Kim, Kyeong Tae;Kim, Seon Yeong;Kim, Dae Woong;Kim, Jae Won;Park, Ji Yeon;Jeon, Sang Min
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.31 no.1
    • /
    • pp.43-49
    • /
    • 2019
  • Purpose : We apply the Layered Rescanning PBS designed to complement the Pencil Beam Scanning(PBS), which is vulnerable to moving organs with the Moving Phantom, and compare the homogeneity with the single scan PBS. Methods and materials: Matrix X (IBA, Belgium) and Moving Phantom (standard imaging, USA) were used. A dose of 200 cGy was measured in the AP direction on a hypothetical tumor $10{\times}10{\times}5cm$. The plan type was planned as 4 kinds of sinlge scan PBS, rescan number 4, 8, 12 times. Were measured three times for each types. During the measurement, the respiratory cycle of the Moving Phantom was generally set to 4 seconds per cycle, and the movement radius in the S-I direction was set to 2 cm. In addition, beam on time was measured. Results : The mean values of $D_{max}$ in the PTV were $246.47{\pm}18.8cGy$, $223.43{\pm}8.92cGy$, and $222.47{\pm}7.7cGy$, $213.9{\pm}6.11cGy$ and the mean values of $D_{min}$ were $165.53{\pm}4.32cGy$, $173.13{\pm}11.94cGy$, $184.13{\pm}8.04cGy$, $182.67{\pm}4.38cGy$ and the mean values of $D_{mean}$ $192.77{\pm}6.98cGy$, $196.7{\pm}4.01cGy$, $198.17{\pm}4.96cGy$, $195.77{\pm}3.15cGy$ respectively. As the number of rescanning increased, the Homogeneity Index converged to 1. The beam on time was measured as 2:15, 3:15, 4:30, 5:37 on average. In the measurement process, in the low dose layer of the MU, the problem was found that it was not rescanned as many times as the set number of rescan. Conclusions : In the treatment of tumors with long-term movements, the application of layered rescanning PBS showed a more uniform dose distribution than single scan PBS. And as the number of rescan increase, the distribution of homogeneity is uniform. Compared with single scan plan and 12 rescan plan, HI value was improved by 0.32. Further studies are expected to be applicable to patients who can not be treated with respiratory synchronous radiation therapy.