A Study on Design and Application of Tissue Compensator for 6MV X-rays (6MV X-선에 대한 조직 보상체의 제작 및 응용에 관한 연구)

  • Chai Kyu Young;Choi Eun Kyung;Chung Woong Ki;kang Wee Saing;Ha Sung Whan;Park Charn Il
    • Radiation Oncology Journal
    • /
    • v.7 no.1
    • /
    • pp.123-132
    • /
    • 1989
  • A radiation beam incident on an irregular or sloping surface produces the non-uniformity of absorded dose. The use of a tissue compensator can partially correct this dose inhomogeneity. The tissue compensator is designed based on the patient's three dimensional contour. After required compensator thickness was determined according to tissue deficit at $25cm\pm25cm$ field size, 10cm depth for 6MV x-rays, tissue deficit was mapped by isoheight technique using laser beam system. Compensator was constructed along the designed model using 0.8mm lead sheet or 5mm acryl plate. Dosimetric verification were peformed by film dosimetry using humanoid phantom. Dosimetric measurements were normalized to central axis full phantom readings for both compensated and non-compensated field. Without compensation, the percent differences in absorbed dose ranged as high as $12.1\%$ along transverse axis, $10.8\%$ along vertical axis. With the tissue compensators in place, the difference was reduced to $0\~43\%$ Therefore, it can be concluded that the compensator system constructed by isoheihnt technique can produce good dose distribution with acceptible inhomogeneity, and such compensator system can be effectively applied to clinical radiotherapy.