A Study on Calibration Procedures for Ir-192 High Dose Rate Brachytherapy Sources (고선량률(HDR) 근접치료의 동위원소 Ir-192에 대한 측정방법에 관한 고찰)

  • Baek, Tae-Seong;Lee, Seung-Wook;Na, Soo-Kyong
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.19 no.1
    • /
    • pp.19-26
    • /
    • 2007
  • Purpose: To compare of the accuracy among various measurement procedures of HDR Brachytherapy, and to evaluate the clinical suitability and usefulness of alternative PMMA (polymethylmethacrylateplastics: $C_5H_8O_2$) plate phantom without any additional cost due to the purchase of measuring apparatus. Materials and Methods: We made a comparative study on three types of measuring systems: well type chamber, source calibration jig, and PMMA plate phantom. Farmer type chamber was used for source calibration jig method and PMMA plate phantom method. Measurement was done 5 times each in comparison with the measurement values from manufacturer. Measurement results from experiment were compared with that from the manufacturer which is offered with the source whenever a source is substituted by a new one and evaluate the accuracy of source activity. Results: As a consequence of Ir-192 source measurement using well type chamber, source calibration jig and PMMA plate phantom, RMS (Root Mean Square) values for the relative error are 0.6%, 1.57%, 2.1%, respectively, compared with the data from manufacturer. And the mean errors with standard deviation are given $-0.2{\pm}0.5%$, $0.97{\pm}1.23%$, $-0.89{\pm}1.87%$ respectively. Conclusion: From the results shown by the three types of measurement system (well type chamber, source calibration jig, and PMMA plate phantom), the measurement with well type chamber produced the best accuracy. It turns out that we can also use the alternative system of PMMA plate phantom clinically without purchasing any additional particular apparatus since the system does not exceed the recommendation of AAPM (American Association of Physicists in Medicine), which requires the error range of within ${\pm}5%$.

  • PDF

Performance Evaluation of Lead (II) Oxide Dosimeter for Digital Quality Assurance in Brachytherapy (방사선 근접치료의 디지털 정도관리를 위한 Lead (II) Oxide 선량계 성능 평가)

  • Han, Moo-Jae;Yang, Seung-Woo;Park, Sung-Kwang
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.4
    • /
    • pp.429-435
    • /
    • 2021
  • In intracavitary radiotherapy, incorrect location of the source can cause excessive dose to normal tissue, so it is essential to evaluate the location accuracy of the source. In this study, basic research was performed on digital line dosimeter based on lead (II) oxide (PbO) to improve analog verification method. Therefore, a polycrystalline PbO unit cell dosimeter was manufactured and the measurement performance for Ir-192 sources was evaluated. As a result, the reproducibility satisfies the evaluation criteria of 1.5% with a relative standard deviation of 0.85%. Linearity showed excellent results with a linear coefficient of R2 of 0.9998. In the case of distance dependence evaluation, the power function R2 showed 0.9855 for PbO and 0.9974 for diode, and the overall average difference was 1.66% for PbO and 2.18% for diode. This study presents the basic detection performance of the polycrystalline PbO dosimeter for the Ir-192 source and can provide basic data in the field of radiation measurement.

Development of resistance welding technology for producing Ir-192 industrial radiation sources (Ir-192 산업용 방사선원의 생산을 위한 저항용접기술 개발)

  • Han, In-Su;Son, Kwang-Jae;Lee, Jun-Sig;Jang, Kyung-Duk;Park, Ul-Jae
    • Proceedings of the KWS Conference
    • /
    • 2007.11a
    • /
    • pp.288-290
    • /
    • 2007
  • Ir-192 source is one of the most widely used radioisotopes in the field of non-destructive testing applications. To obtain radiation safety it is necessary to take into consideration integrity of welded joint in the production of sealed radiation source. Generally, the quality of a resistance welded joint is strongly influenced by process parameters during the welding process such as current, welding time and applied force. In this study, resistance welding technology and system were developed for sealing of Ir-192 industrial radiation source capsules. In order to evaluate the weld quality in real time, quantitative relationships between process parameters and electrode displacement were also established.

  • PDF

High Dose Rate Ir-192 Source Calibration Method with Newly Designed Calibration Jig (고선량 Ir-192선원 교정기의 제작 및 특성)

  • Yi, Byong-Yong;Choi, Eun-Kyung;Chang, Hye-Sook
    • Radiation Oncology Journal
    • /
    • v.7 no.2
    • /
    • pp.299-303
    • /
    • 1989
  • Authors have developed highly reproducible calibration method for the Micro-Selectron HDR Ir-192 system (Nucletron, Motherland). The new jig has a 10cm radius circular hole in the $30cm{\times}30cm{\times}0.2cm$ acrylic plate, and 5F flexible bronchial tubes are attached around the hole. The source moves along the circle in the tubes and the ionization chamber is placed verticaly at the center of the circular hole (center of the jig). Dose distribution near the center was derived theoretically, and measured with the film dosimetry system. Theoretical calculation and measurement show the error margin below $0.1\%$ for 1mm or 2mm position deviation. We have measured at 12 and 24 points of circle with 1, 6, 11 and 21 second dwell time of source in order to calculate the activity of the source. Measurements have been repeated daily for 50 days. The accuracy and the reproducibility are below $1\%$ error margin. The half life of the source from our measurement is estimated $73.4\pm0.4$ days.

  • PDF

Dose Comparison of Treatment Plans Using Different Ir-192 Sources and Treatment Planning Systems for Intracavitary HDR Brachytherapy (고선량률 강내 근접치료에 사용되는 Ir-192 선원과 치료계획 시스템간의 계산선량 비교)

  • Park, Dong-Wook;Kim, Young-Seok;Park, Sung-Ho;Choi, Eun-Kyung;Kim, Jong-Hoon;Lee, Sang-Wook;Song, Si-Yeol;Ahn, Seung-Do;Noh, Young-Joo
    • Progress in Medical Physics
    • /
    • v.20 no.1
    • /
    • pp.1-6
    • /
    • 2009
  • For HDR intracavitary brachytherapy with ovoids and a tandem, we compared the dose discrepancy of treatment plans using two different Ir-192 sources (microSelectron, Varian) and generated on two different treatment planning systems (PLATO, BrachyVision). The treatment plans of ten patient treated from Oct. 2007 to Jan. 2008 were selected for these comparisons. For the comparison of dose calculation using different sources, the average discrepancies were $-0.91{\pm}0.09%$, $-0.27{\pm}0.07%$, $0.22{\pm}0.39%$, and $0.88{\pm}0.37%$ in total treatment time and at B-point and ICRU bladder and rectum reference point, respectively. Comparing the two systems, the average dose discrepancies between treatment planning programs were $-0.22{\pm}0.42%$, $-0.25{\pm}0.29%$, $-0.23{\pm}0.63%$, and $-0.17{\pm}0.76%$, and the average dose discrepancies between positioning methods (PLATO with film and BrachyVision with digitial image) were $-0.61{\pm}0.59%$, $-0.77{\pm}0.45%$, $-0.72{\pm}1.70%$, and $0.35{\pm}2.82%$ at A-point, B-point, and ICRU bladder and rectum reference points, respectively. The rectal dose discrepancies between two systems were reached 5.87%. The difference in the dwell position expected by each TPS are mainly affected by the differences in the positioning method in TPSs and have an effect on dose calculations of rectal and bladder located in AP direction.

  • PDF

The Effect of Process Parameters on Sealing Quality for Ir-192 Radiation Source Capsule using Resistance Spot Welding (Ir-192 방사선원의 밀봉 용접부 품질에 미치는 저항용접 공정변수의 영향)

  • Han, In-Su;Son, Kwang-Jae;Lee, Young-Ho;Lee, You-Hwang;Lee, Jun-Sig;Jang, Kyung-Duk;Park, Ul-Jae;Park, Chun-Deuk
    • Journal of Welding and Joining
    • /
    • v.27 no.1
    • /
    • pp.65-70
    • /
    • 2009
  • Ir-192 radiation sealed sources are widely employed to the therapeutic applications as well as the non-destructive testing. Production of Ir-192 sources requires a delicate but robust welding technique because it is employed in a high radioactive working environment. A GTA(Gas Tungsten Arc) welding technique is currently well established for this purpose. However, this welding method requires a frequent replacement of the electrode, which results in the delay of the production to take a preparatory action such as to isolate the radiation sources from the working place before getting access to the welding machine. Hence, a resistance welding technique is considered as an alternative method of the GTA welding technique. The advantages of resistance welding are high welding speed and high-rate production. Also it has very long life of electrode comparing to GTA welding. In this study, the resistance welding system and proper welding conditions were established for sealing Ir-192 source capsule. As a results of various experiments, it showed that electrode displacement can be employed as a indicator to predict welding quality. We proposed two mathematical models(linear and curvilinear) to estimate electrode displacement with process parameters such as applied force, welding current and welding time by using regression analysis method. Predicting results of both linear and curvilinear model were relatively good agreement with experiment.

Analysis of Accuracy of Apparent Activity According to Calibration Method for High Dose Rate Brachytherapy Source (측정용 전리함에 따른 고 선량율 근접치료용 방사성 선원의 겉보기 활성도의 정확도 비교 분석)

  • Huh, Hyun-Do;Choi, Jin-Ho;Choi, Sang-Hyoun;Kim, Seong-Hoon;Kim, Woo-Chul;Kim, Hun-Jeong;Lee, Re-Na;Kim, Kum-Bae;Hong, Seong-Eon;Shin, Dong-Oh
    • Progress in Medical Physics
    • /
    • v.19 no.4
    • /
    • pp.305-312
    • /
    • 2008
  • The aim of this study is to introduce the accuracy of Ir-192 source's apparent activity using the well-type chamber and the Farmer-type ionization chamber in the high dose rate brachytherapy. We measured the apparent activity of Ir-192 that each medical center in the country has and the apparent activity of calibration certificate provided by manufacturer is compared with that by our experimental measurement. The number of sources used for the activity comparison was 5. The accuracy of the measured activity was in the range of -2.8% to -1.0% and -2.1% to 0.2% for the Farmer-type chamber system (Jig) and for the well-type, respectively. The maximum difference was within 1.0% for comparison with two calibration's tool. Our results demonstrate that well-type chamber as wall as Farmer-type chamber is a appropriate system as the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, by periods, a source calibration should be carried out.

  • PDF