• 제목/요약/키워드: Interspinous Process Implant

검색결과 2건 처리시간 0.02초

퇴행성 요추부 척추관 협착증 치료를 위한 극돌기간 고정기구의 설계 및 생체역학적 분석 (Design of the Interspinous Process Fixator Using Biomechanical Analysis for the Treament of Degenerative Lumbar Spinal Stenosis)

  • 허순;손권;이성재;문병영
    • 한국정밀공학회:학술대회논문집
    • /
    • 한국정밀공학회 2005년도 춘계학술대회 논문집
    • /
    • pp.1963-1966
    • /
    • 2005
  • Degenerative lumbar spinal stenosis(DLSS) is a disease inducing low back pain, leg pain, convulsion, numbness, and neurogenic claudication from compression of nerve root. Intervertebra fixation was reported to increase the degenerative of neighbor region after treatment. Recently, a new surgical technique of inserting a fixator between interspinous processes has been introduced. The purpose of this study is to design of the interspinous process fixator with flexibility to complement the trouble of using fixator in DLSS. This study evaluated the existing fixator through the mechanical test and modified fixators using the finite element analysis(FEA). Displacement, stiffness and Von-Mises stress were found to have similar values to those obtained from the mechanical test and the FEA in the biomechanical loading condition. Effects of variation in length and thickness were investigated to design an optimal fixator.

  • PDF