• Title/Summary/Keyword: Golden STT (Segmented Treatment Table)

Search Result 2, Processing Time 0.014 seconds

Simplistic QA for an Enhanced Dynamic Wedge using the Reversed Wedge Pair Method (역방향 조사방식을 통한 동적쐐기의 품질관리)

  • Lee Jeong Woo;Hong Semie;Suh Tae Suk
    • Progress in Medical Physics
    • /
    • v.15 no.3
    • /
    • pp.161-166
    • /
    • 2004
  • A simplistic quality assurance (QA) method was designed for a Linac built-in enhanced dynamic wedge (EDW), which can be utilized to make wedged beam distributions. For the purpose of implementing the EDW symmetry QA, a film dosimetry system, low speedy dosimetry film, film densitometer and 3D RTP system were used, and the films irradiated by means of a 60$^{\circ}$ Reversed wedge pair (REWP) method. The profiles were then analyzed in terms of their symmetries, including partial treatment, which is the case of stopping it abruptly during EDW irradiation, and the measured and calculated values compared using the Cad Plan Golden Segmented Treatment Table (Golden STT). The result of this experiment was in good agreement, within 1 %, of the 'reversed wedge pair counterbalance effect'. For the QA of the effective wedge factor (EWF), the authors measured EWFs in relation to the 10$^{\circ}$, 15$^{\circ}$, 20$^{\circ}$, 25$^{\circ}$, 30$^{\circ}$, 45$^{\circ}$ and 60$^{\circ}$ EDW, which were compared with the calculated values using the correction factor derived from the Golden STT and the log files produced automatically during the process of EDW irradiation. By means of this method it was capable of check up the safety of effective wedge factor without any other dosimetry system. The EDW QA was able to be completed within 1 hour from irradiation to analysis as a consequence of the simplified QA procedure, with maximized effectiveness. Unlike the metal wedge system, the EDW system was heavily dependent on the dose rates and jaw movements; therefore, its features could potentially cause inaccuracy. The frequent simplistic QA for the EDW is essential, and could secure against the flaw of dynamic treatment that uses the EDW.

  • PDF

Comparison of Enhanced Dynamic Wedge with Physical Metal Wedge based on the Basic Dosimetric Parameters (선량계측인자에 따른 기능강화동적쐐기와 금속쐐기의 비교)

  • Lee Jeong-Woo;Hong Semie;Choi Kyoung-Sik;Chung Jin-Beom;Choe Bo-Young;Jang Hong Seok;Suh Tae-Suk
    • Progress in Medical Physics
    • /
    • v.16 no.2
    • /
    • pp.70-76
    • /
    • 2005
  • For clinical implementation of Enhanced Dynamic Wedge (EDW), it is necessary to adequately analyze and commission its dosimetric properties in comparison to common physical metal wedge (MTW). This study was implemented with the essential measurements of parameters for clinical application, such as percentage depth dose, peripheral dose, surface dose, effective wedge factor, and wedge profile. In addition, through the comparison study of EDW with open and MTW, the analysis was performed to characterize the EDW. We also compared EDW dose profiles of measured values using chamber array 24 (CA24) with calculated values using radiation treatment planning system. PDDs of EDW showed good agreements between $0.2\~0.5\%$ of open beam, but $2\%$ differences with MTW. In the result of the measurements of peripheral dose, it was shown that MTW was about $1\%$ higher than open field and EDW. The surface doses of $60^{\circ}$ MTW showed 10% lower than the others. We found that effective wedge factor of EDW had linear relationships according to Y jaw sizes and was independent of X jaw sizes and was independent of X jaw sizes and asymmetric Y jaw opening. In comparison with measured values and calculate values from Golden-STT based radiation treatment planning system (RTP system), it showed very good agreement within difference of $1\%$. It could be concluded that EDW is a very reliable and useful tool as a beam modification substitute for conventional MTW.

  • PDF