• 제목/요약/키워드: Eye Entrance Surface Dose

검색결과 1건 처리시간 0.015초

컴퓨터 방사선영상에서 고 관전압 기법을 이용한 안구 입사표면선량 감소에 관한 연구 (A Study on Reducing of Entrance Surface Dose with the Eye in the Computed Radiography by Using High Kilo Voltage Peak Technique)

  • 성열훈;임재동
    • 대한안전경영과학회지
    • /
    • 제13권2호
    • /
    • pp.91-96
    • /
    • 2011
  • The purpose of this study was to minimize of entrance surface dose (ESD) at the eye using high kVp technique in the computed radiography. We used REX-650R (Listem, Korea) general X-ray unit, and external detector with ESD dosimeter of Piranha 657 (RTI Electronics, Sweden). We used head of the whole body phantom. The total 64 images of X-ray anterior-posterior of skull were acquired using the film/screen (F/S) method and the digital of computed radiography method. The three radiology professor of more 10 years of clinical career evaluated a X-rays images in the same space by 5-point scale. The external detector was performed measurement of ESD of three times by same condition on the eye of the head phantom. The good image quality in the F/S method (90 kVp, 2.5 mAs) showed at the minimized ESD of 0.310 ${\pm}$ 0.001 mGy. the good image quality in the computed radiography method (90 kVp, 2.0 mAs) showed at the minimized ESD of 0.180 ${\pm}$ 0.002 mGy (P = 0.002). Finally the radiation dose could reduced about 50% in the computed radiography method more than the F/S method. In addition the eye entrance surface dose using high kVp technique with the computed radiography was reduced 92% more than conventional technique (F/S method).