Evaluating Accuracy according to the Evaluator and Equipment Using Electronic Apex Locators

  • Yu, Beom-Young;Son, Keunbada;Lee, Kyu-Bok
    • Journal of Korean Dental Science
    • /
    • 제13권2호
    • /
    • pp.52-58
    • /
    • 2020
  • Purpose: Using two types of electronic apex locators, this study aimed to investigate the differences in accuracy according to the evaluator and equipment. Materials and Methods: Artificial teeth of the lower first premolars and two mandibular acrylic models (A and B) were used in this study. In the artificial teeth, the pulp chamber was opened and the access cavity was prepared. Using calibrated digital Vernier calipers, the distance from the top of the cavity and the root apex was measured to assess the actual distance between two artificial teeth. The evaluation was conducted by 20 dentists, and each evaluator repeated measurements for each electronic apex locator five times. The difference between the actual distance from the top of the cavity to the root apex and the distance measured using electronic measuring equipment was compared. For statistical analysis, the Friedman test the Mann-Whitney U-test were conducted and the differences between groups were analyzed (α=0.05). Result: As for the accuracy of measurement according to the two types of electronic apex locators, the value of the measurement error was 0.4753 mm in Dentaport ZX and 0.3321 mm in E-Cube Plus. Moreover, electronic apex locators Dentaport ZX and E-Cube Plus showed statistically significant differences (P<0.05). As for the difference in the accuracy of the two types of electronic apex locators according to the evaluator, the resulting values differed depending on the evaluator and showed a statistically significant difference (P<0.001). Conclusion: Electronic apex locator E-Cube Plus showed higher accuracy than did Dentaport ZX. Nevertheless, both types of electronic apex locators showed 100% accuracy in finding the region within root apex ±0.5 mm zone. Furthermore, according to the evaluator, the two electronic apex locators showed different resulting values.

하악 대구치 근심근관에 대한 전자근관장 측정기의 정확도 (AN ACCURACY OF THE SEVERAL ELECTRONIC APEX LOCATORS ON THE MESIAL ROOT CANAL OF THE MANDIBULAR MOLAR)

  • 조영린;손욱희;황호길
    • Restorative Dentistry and Endodontics
    • /
    • 제30권6호
    • /
    • pp.477-485
    • /
    • 2005
  • 근단부위 형성이 완료되고 우식이 없는 하악 대구치 근심치근 50개를 대상으로 주파수 의존형 전자근관장 측정기를 이용하여 측정한 근관장의 정확도를 평가하기 위해 치수강 개방 후 동일한 치아에서 방사선 사진으로 확인한 실측 근관장을 대조군, 각종 전자근관장 측정기로 측정한 근관장을 실험군 ( I군: Root-ZX, II군: Bingo, III군: Propex, IV군: Diagnostic)으로 분류하여 근심협측과 설측의 근관장을 근관의 입구 측으로부터 치근단 1/3까지 확대여부 및 서로 다른 측정시기에 측정하여 비교 평가한 결과 다음과 같은 결론을 얻었다. 하악대구치 근심치근의 협측과 설측 전자근관장은 측정기의 종류, 근관의 입구 측으로부터 치근단 1/3까지의 확대여부 및 측정시기에 상관없이 0.5mm이내의 오차범위를 나타내었으며, 근관치료시 오차한계인 ${\pm}1mm$에서 정확성은 근관의 입구 측으로부터 치근단 1/3까지 미리 확대한 후에 근관장을 측정할 경우와 측정시기가 반복될 경우에 증가되는 경향을 보여 임상에 유용하게 활용할 수 있을 것으로 사료된다.

APEX FINDER A.F.A.의 정확성에 관한 연구 (EVALUATION OF THE ACCURACY OF THE APEX FINDER A.F.A.)

  • 양미영;유현미;오태석
    • Restorative Dentistry and Endodontics
    • /
    • 제23권2호
    • /
    • pp.670-675
    • /
    • 1998
  • Recently electronic apex locators have been used widely in root canal treatment, but, accuracy of electronic apex locators is controversial. The purpose of this study was to evaluate the accuracy of Apex Finder A.F.A(EIE Analytic Technology, U.S.A.) in vivo compared with Root-Zx and radiograph. The root canal lengths were determined with Root-Zx(32 tooth) in before pulp extirpation and after pulp extirpation. Then the radiographs were taken with a file in the canal. The root canal lengths were determined with Apex Finder A.F.A.(21 tooth) in before pulp extirpation and after pulp extirpation and under NaOCl. Then the radiographs were taken with a file in the canal. The results were as follows: 1. There was no significant statistical difference in Root-Zx between before pulp extirpation and after pulp extirpation(p > 0.05). 2. There was no significant statistical difference in Apex Finder A.F.A. between before pulp extirpation and after pulp extirpation(p > 0.05). But, there was significant statistical difference under NaOCl(p < 0.05). 3. There was no significant statistical difference in accuracy between Root-Zx and Apex Finder A.F.A.

  • PDF

2종 전자근관장측정기의 일관성에 관한 in vitro 연구 (IN VITRO EVALUATION OF THE CONSISTENCY OF TWO ELECTRONIC APEX LOCATORS)

  • 황규용;노병덕;김의성;이승종
    • Restorative Dentistry and Endodontics
    • /
    • 제33권1호
    • /
    • pp.20-27
    • /
    • 2008
  • 정확한 근관장 측정은 성공적인 근관치료를 위해 필수적이다. Kuttler는 근관협착부는 주근단공에서 0.5 mm 상방에 위치한다고 하였고 Lee는 상아백악질 경계점부위보다 주근단공이 재현성이 높고 임상적으로 관찰하기 용이하다고 보고하였다. 본 연구는 alginate model상에서 2개의 전자근관장측정기를 사용하여 얻은 측정치의 정확성(accuracy)을 평가하고 각각 0.5 mark와 Apex mark 중에서 어느 지점에서 더 일관성(consistency)을 보이는지를 비교하고 또한 근관장을 측정할 때 파일을 전진시키면서 Apex mark를 측정한 값과 치근단부위를 지나친 후 다시 후퇴하면서 측정한 값의 차이를 비교하고자 하였다. 52개의 발거된 건전한 제 1, 2소구치를 대상으로 하였으며 Root ZX와 E-Magic Finder Deluxe를 이용하여 각각 26개중 13개는 파일을 전진하면서 0.5 mark에서 근관장을 측정하고 Apex mark에서 근관장을 측정한 후 0.5 mark에서 고정하였다. 나머지 13개는 0.5 mark에서 근관장을 측정한 후 Apex mark에서 근관장을 측정하고 다시 치근단 부위를 넘어선 후 후퇴하면서 Apex mark에서 근관장을 측정하고 고정한 후 치근단부 4 mm를 삭제하여 현미경상에서 파일 끝부터 주근단공까지의 거리를 측정하였다. 그 결과 Root ZX와 E-Magic Finder 모두, 실험군 100%에서 주근단공과 file tip간의 거리가 임상적 허용범위인 ${\pm}0.5\;mm$내에 있었으며 0.5 mark보다 Apex mark에서 더 높은 일관성을 보였다. 주근단공에서 fie tip사이의 거리는 Root ZX의 0.5 mark에서 -0.18 mm, Apex mark에서 -0.07 mm이고 E-Magic Finder의 0.5 mark에서 -0.25 mm, Apex mark에서 -0.02 mm를 나타내었다. Apex hand는 0.04 mm였다. 따라서 Alginate model을 사용한 본 실험조건에서는 Apex mark에서 근관장을 측정한 결과 거의 주근단공과 일치하고 다른 여러 연구에서 주근단공에서 0.5 mm상방에 근관협착부가 위치한다고 보고하였으므로 임상적인 측면에서 Apex mark에서 측정한 근관장에서 0.5 mm를 빼서 근관장으로 사용하는 것이 임상적으로 더 좋으리라고 사료된다.

In vivo assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen

  • Tampelini, Fernanda Garcia;Coelho, Marcelo Santos;de Azevedo Rios, Marcos;Fontana, Carlos Eduardo;Rocha, Daniel Guimaraes Pedro;Pinheiro, Sergio Luiz;da Silveira Bueno, Carlos Eduardo
    • Restorative Dentistry and Endodontics
    • /
    • 제42권3호
    • /
    • pp.200-205
    • /
    • 2017
  • Objectives: The aim of this in vivo study was to assess the accuracy of 2 third-generation electronic apex locators (EALs), Propex II (Dentsply Maillefer) and Root ZX II (J. Morita), and radiographic technique for locating the major foramen (MF). Materials and Methods: Thirty-two premolars with single canals that required extraction were included. Following anesthesia, access, and initial canal preparation with size 10 and 15 K-flex files and SX and S1 rotary ProTaper files, the canals were irrigated with 2.5% sodium hypochlorite. The length of the root canal was verified 3 times for each tooth using the 2 apex locators and once using the radiographic technique. Teeth were extracted and the actual WL was determined using size 15 K-files under a ${\times}25$ magnification. The Biostat 4.0 program (AnalystSoft Inc.) was used for comparing the direct measurements with those obtained using radiographic technique and the apex locators. Pearson's correlation analysis and analysis of variance (ANOVA) were used for statistical analyses. Results: The measurements obtained using the visual method exhibited the strongest correlation with Root ZX II (r = 0.94), followed by Propex II (r = 0.90) and Ingle's technique (r = 0.81; p < 0.001). Descriptive statistics using ANOVA (Tukey's post hoc test) revealed significant differences between the radiographic measurements and both EALs measurements (p < 0.05). Conclusions: Both EALs presented similar accuracy that was higher than that of the radiographic measurements obtained with Ingle's technique. Our results suggest that the use of these EALs for MF location is more accurate than the use of radiographic measurements.

근관 성형 전후의 네 가지 전자근관장측정기의 측정 정확성의 비교 (In vitro comparison of measurement accuracy In pre-enlarged and enlarged canals with four apex locators)

  • 성상엽;박정길;허복;김현철
    • Restorative Dentistry and Endodontics
    • /
    • 제31권5호
    • /
    • pp.371-377
    • /
    • 2006
  • 이 연구의 목적은 확대하기 전 근관에서 전자근관장 측정 시의 정확성을 평가하고 확대된 근관에서 파일의 크기에 따른 네 가지 전자근관장 측정기의 정확성을 비교하기 위한 것이다. 발치된 치아 10개를 #10파일을 이용하여 치아의 실제길이를 측정하였다 현미경에서 25배 확대 하에 #10 파일이 치근단공을 넘어 해부학적 치근단공에 이를 때까지 전진하여 디지털 캘리퍼로 측정하였다. Root ZX, Bingo 1020, SmarPex, e-Magic Finder를 알지네이트를 이용한 인체 재현 모델을 이용하여 #10 파일로 apex 표시등에서 측정하였다 (S10). #45까지 핸드 파일로 확대한 다음 #10, #40 파일을 이용하여 전자근관장을 측정하였다. 실제 길이와 각각 측정값의 차이를 계산하여 L10, L40으로 정의하고 기계간의 비교는 one-way ANOVA통계처리하고, Scheffe's multiple range test로 사후 검증하였고, 같은 기계 안의 L10, L40은 Student's t-test로 비교하였다. 본 연구의 실험 조건 하에서, 측정값의 정확도가 기계간에 차이를 보였고, 파일의 크기는 측정값의 정확도에 영향을 주지 못하며, e-Magic Finder는 이 실험 조건하에서 다른 기계에 비해 가장 정확하였다.

An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files

  • Gehlot, Paras Mull;Manjunath, Vinutha;Manjunath, Mysore Krishnaswamy
    • Restorative Dentistry and Endodontics
    • /
    • 제41권1호
    • /
    • pp.6-11
    • /
    • 2016
  • Objectives: The purpose of this in vitro study was to evaluate the accuracy of working length (WL) determination of four electronic apex locators (EALs), namely, Root ZX (RZX), Elements diagnostic unit and apex locator (ELE), SybronEndo Mini Apex locator (MINI) and Propex pixi (PIXI) using Stainless steel (SS) and nickel-titanium (NiTi) hand files. The null hypothesis was that there was no difference between canal length determination by SS and NiTi files of 4 EALs. Materials and Methods: Sixty extracted, single rooted human teeth were decoronated and the canal orifice flared. The actual length (AL) was assessed visually, and the teeth were embedded in an alginate model. The electronic length (EL) measurements were recorded with all four EALs using SS and NiTi files at '0.5' reading on display. The differences between the AL and EL were compared. Results: The results obtained with each EAL with SS and NiTi files were compared with AL. A paired sample t test showed that there was a statistical significant difference between EAL readings with SS and NiTi files for RZX and MINI (p < 0.05). The accuracy of RZX, ELE, MINI and PIXI within ${\pm}0.5 mm$ of AL with SS/NiTi files were 93.3%/70%, 90%/91.7%, 95%/68.3%, and 83.3%/83.3%, respectively. Conclusions: The results of this study indicate that Root ZX was statistically more accurate with NiTi files compared to SS files, while MINI was statistically more accurate with SS files compared to NiTi files. ELE and PIXI were not affected by the alloy type of the file used to determine WL.

4종 전자근관장측정기의 정확성과 일관성에 관한 in vitro 연구 (IN VITRO EVALUATION OF ACCURACY AND CONSISTENCY OF FOUR DIFFERENT ELECTRONIC APEX LOCATORS)

  • 조재현;금기연;이승종
    • Restorative Dentistry and Endodontics
    • /
    • 제31권5호
    • /
    • pp.390-397
    • /
    • 2006
  • 이번 연구는 서로 다른 4개의 전자근관장측정기의 정확성을 측정하고 각각 0.5지점과 Apex지점에서의 일관성을 비교하고자 하였다. 40개의 발치된 상하악 소구치를 대상으로 치수강 개방 후 alginate model에 고정시키고 근관장을 측정하였다. 사용된 전자근관장측정기는 Root ZX (Merits, Tokyo, Japan), SmarPex (META, Seoul, Korea). Elements Diagnostic Unit (SybronEndo, CA, USA), E-Magic Finder Deluxe (S-Denti, Seoul, Korea)이다. 먼저 모든 치아에서 4개의 전자근관장측정기를 사용하여 0.5지점과 Apex지점에서 근관장을 측정하여 한 치아당 8개의 측정값을 얻었다. 다음으로 치아를 각 전자근관장측정기당 10개씩 4개의 그룹으로 나누어, 각각 제조사의 지시대로 Root ZX, Elements Diagnostic Unit 및 E-Magic Finder Deluxe는 "0.5"지점에서, SmarPex는 "Apex"지점에서 file을 치아에 cement로 고정시켰다. 이후 치근단부 4 mm를 삭제하여 100배율의 Image Proplus로 관찰하여 file 끝에서 주근단공의 외연까지의 실제거리를 측정한 후, 4개의 전자근관장측정기의 0.5지점 및 Apex지점에서 file끝과 주근단공 사이의 거리를 계산하여 비교하였다. 그 결과 Root ZX와 E-Magic Finder는 실험군 100%, SmarPex는 90%, Elements Diagnostic Unit는 70%에서 주근단공과의 거리가 임상적 허용범위인 ${\pm}0.5 mm$이내에 있었다. 또한 각 전자근관장측정기 마다 0.5지점과 Apex지점에서의 근관장의 표준편차와 사분위 범위를 구하여 두 지점간의 일관성을 비교한 결과, Root ZX, E-Magic Finder는 0.5지점과 Apex지점에서 비슷한 일관성을 보였으며 SmarPex와 Elements Diagnostic unit는 Apex지점에서 0.5지점보다 더 높은 일관성을 보였다. 전자근관장측정기는 근관 내의 조건에 관계없이 근첨협착부에서 항상 일정한 거리를 재현해 낼 수 있는 일관성이 중요하므로, 이렇게 0.5지점 또는 Apex지점에서의 일관성이 증명된다면 실제 임상에서 사용할 때 전자근관장에서 일정한 거 리를 가감하여 사용할 수 있다.

미세단층촬영기(Micro-CT)를 이용한 전자 근관장 측정기의 정확성에 관한 연구 (A Study on the Accuracy of the Electronic Apex Locator Using a Micro-Computed Tomography)

  • 전경진;김양수;남태계
    • 한국정밀공학회지
    • /
    • 제24권8호통권197호
    • /
    • pp.116-121
    • /
    • 2007
  • The length of root canal has to be measured for endodontic treatment. Several electronic apex locators were developed to measure the length of root canal by other researchers. And their accuracies were verified by X-ray or micrometer method. But these methods did not consider the non-linear bends of pulp and had ${\pm}0.5mm$ error which was large enough to measure the length of root canal. The purpose of this study is the introduction of a new method to measure the length of root canal and the verification of the accuracy of an electronic apex locator using a Micro-CT. The length of root canal of 6 teeth were measured with the electronic apex locator. When the electronic apex locator reads 0.5, 0.6, 0.7, 0.8, 0.9 and 1.0 mm length of the file which was inserted in the hole of the tooth to measure the length of root canal. The average (${\pm}$Standard deviation) length of root canal of 6 teeth measured by the Micro-CT was $0.49{\pm}0.03,\;0.59{\pm}0.04,\;0.68{\pm}0.03,\;0.78{\pm}0.03,\;0.90{\pm}0.04\;and\;1.01{\pm}0.03mm$, respectively. The maximum error of the electronic apex locator was 0.06 mm.

주파수 의존형 전자근관장측정기 개발 및 임피던스차를 이용한 자동보정에 의한 성능향상 (Development of Frequency Dependent Type Apex-Locator Improvementation by Auto-calibration Using Impedance Difference)

  • 남기창;김수찬;김덕원;이승종
    • 대한의용생체공학회:학술대회논문집
    • /
    • 대한의용생체공학회 1997년도 추계학술대회
    • /
    • pp.399-402
    • /
    • 1997
  • It is important to find a accurate root canal length or successful endodontic therapy. By X-ray method takes a long time and difficult in a curved canal. After developed electronic apex locator has allowed to measure the root canal length with easy, in a short time also in a curved canal. But most electronic apex locators have the disadvantage which is too short reading or sometimes the measurement itself becomes impossible if there are electrolytes in the canal. To overcome this drawback, impedance ratio method has been developed. In this study, we have developed frequency dependent electronic apex locator to minimize the interference of electrolytes. And based on that also some error in clinic use, we added the other method. Difference of two signals which are used in calculation of impedance ratio was can be represent the status of root canal fluid. As a result, using impedance ratio method and auto-calibration by voltage difference method can reduce the measurement error.

  • PDF