Evaluation by thickness of a linear accelerator target at 6-20 MeV electron beam in MCNP6

  • Dong-Hee Han ;Kyung-Hwan Jung;Jang-Oh Kim ;Da-Eun Kwon ;Ki-Yoon Lee;Chang-Ho Lee
    • Nuclear Engineering and Technology
    • /
    • v.55 no.6
    • /
    • pp.1994-1998
    • /
    • 2023
  • This study quantitatively evaluated the source term of a linear accelerator according to target thickness for a 6-20 MeV electron beam using MCNP6. The elements of the target were tungsten and copper, and a composite target and single target were simulated by setting different thickness parameters depending on energy. The accumulation of energy generated through interaction with the collided target was evaluated at 0.1-mm intervals, and F6 tally was used. The results indicated that less than 3% reference error was maintained according to the MCNP recommendations. At 6, 8, 10, 15, 18, and 20 MeV, the energy accumulation peaks identified for each target were 0.3 mm in tungsten, 1.3 mm in copper, 1.5 mm in copper, 0.5 mm in tungsten, 0.5 mm in tungsten, and 0.5 mm in tungsten. For 8 and 10 MeV in a single target consisting only of copper, the movement of electrons was confirmed at the end of the target, and the proportion of escaped electrons was 0.00011% and 0.00181%, respectively.

Fs-THz beam line electron accelerator of pulse modulator design and fabrication (펨토 초-테라헤르츠 빔 라인용 펄스 모듈레이터의 설계와 제작)

  • Son, Yoon-Kyoo;Kwon, Sei-Jin;Suh, Jae-Hak;Jang, Sung-Duck;Kang, Heung-Sik;Lee, Kyung-Tae;Ro, Sung-Chae
    • Proceedings of the KIEE Conference
    • /
    • 2008.07a
    • /
    • pp.1322-1323
    • /
    • 2008
  • 기존의 펨토 초 레이저를 이용하여 발생시킨 테라 헤르츠 광원의 한계를 극복하기 위하여 대용량의 가속기를 이용한 테라 헤르츠 광원의 발생에 관한 연구가 활발히 진행되고 있다. 포항방사광 가속기연구소에서도 펨토 초 테라 헤르츠 빔을 이용한 분광학적인 방법을 사용하여 단백질 접힘과 DNA-단백질 간 상호작용, 화학적, 생물학적인 반응 동력학 등에 관한 연구와 영상 기술개발 등을 할 계획을 가지고 펨토초 테라 헤르츠 빔 라인을 건설 중에 있다. 펨토 초-테라 헤르츠 빔 라인의 마이크로웨이브를 가속하는 장치에 사용되는 전원장치의 설계와 제작 및 시험과정을 외국기술에 의존하지 않고 순수 국내기술로 실현하였다. 본 논문에서는 펄스 모듈레이터의 설계와 실험결과를 보이고자 한다.

  • PDF

Design of cryostat for superconducting quadrupole magnets in In-Flight fragmentation separator

  • Choi, Y.S.;Chang, H.M.;Baudouy, B.;Kim, D.G.;Kim, J.W.
    • Progress in Superconductivity and Cryogenics
    • /
    • v.17 no.3
    • /
    • pp.62-66
    • /
    • 2015
  • The cryostat is designed for the superconducting quadrupole magnets to be used in a heavy-ion accelerator facility. The main accelerator is superconducting linac, which can accelerate a $^{238}U$ beam to 200 MeV/u (Mega electron voltage per nucleon). The cryostat for the magnet employs an innovative design primarily driven by the requirement of the compactness, user-friendliness and reliability. Also, several ancillary requirements such as background field, space restriction due to the beam line and cryostat structure need technical attentions. The development of the cryostat for three quadrupole magnets in the in-flight fragmentation separator is presented in the paper. The concept of cryogenic design is reported and the amount of cryogenic load is estimated by a relevant analysis. The structure of the cryostat to endure the heavy iron yoke including three quadrupole magnets is presented. In addition, the design as well as the performance test of the support link for the cold mass is described.

Inactivation of Agrobacterium tumefaciens Inoculated on Fresh Radix Ginseng by Electron Beam Irradiation and Aqueous Chlorine Dioxide Treatment

  • Chun, Ho-Hyun;Kim, Ju-Yeon;Song, Kyung-Bin
    • Journal of Applied Biological Chemistry
    • /
    • v.51 no.3
    • /
    • pp.117-122
    • /
    • 2008
  • Inactivation of Agrobacterium tumefaciens was evaluated on the inoculated fresh Radix Ginseng by electron beam irradiation or aqueous chlorine dioxide ($ClO_2$) treatment. Two groups of fresh ginsengs were prepared and inoculated with A. tumefaciens. One group was then irradiated at 0, 2, and 4 kGy using an electron beam accelerator, and the other group was treated with 0, 50, and 100 ppm of aqueous $ClO_2$. Microbiological data indicated that populations of A. tumefaciens significantly decreased with increasing irradiation dose or aqueous $ClO_2$ concentration. In particular, A. tumefaciens was eliminated by irradiation at 4 kGy, and 100 ppm $ClO_2$ treatment reduced the populations of A. tumefaciens by 1.44 log CFU/g. These results suggest that electron beam irradiation or aqueous $ClO_2$ treatment can be useful in improving the microbial safety of fresh ginsengs during storage.

Monte Carlo Simulation for Electron Contamination of Photon Beam (치료용 광자선의 전자오염에 대한 몬테카를로 시뮬레이션)

  • Chung, Kap-Soo;Ko, Shin-Kwan;Yang, Han-Joon;Han, Chang-Yul
    • Journal of radiological science and technology
    • /
    • v.22 no.2
    • /
    • pp.47-51
    • /
    • 1999
  • We calculated the energy distribution and the percentage depth-dose at 10 cm in a $10{\times}10\;cm^2$ with a photon beam at SSD of 100 cm by using a Monte Carlo Simulation. PDD is used as a beam-quality specifier for radiotherapy beams. It is better than the commonly used values of TPR or nominal accelerating potential. The presence of electron contamination affects the measurement of PDD, but can be removed by the use of a 0.1 cm lead filter. It reduces surface dose from contaminant electrons from the accelerator by more than 90% for radiotherapy beams. The filter performs best when it is placed immediately below the head. An electron-contamination correction factor is introduced to correct for electron contamination from the filter and air. It converts PDD which includes the electron contamination with the filter in place into PDD for the photons in the filtered beam. The correction factor can be used to determine stopping-power ratio. Calculations show that the values of water-to-air slopping power ratio in the unfiltered beam are related to PDD.

  • PDF

Disinfection of Total Coliforms in Sewage Treatment Effluent using Electron Beam (전자선을 이용한 하수처리장 방류수내 대장균군 살균)

  • Kim, Yuri;Han, Bumsoo;Kim, Jinkyu;Kang, Ho
    • Journal of Korean Society on Water Environment
    • /
    • v.20 no.4
    • /
    • pp.376-381
    • /
    • 2004
  • The use of electron beam irradiation was investigated to disinfect total coliforms in the secondary sewage treatment effluent. Unchlorinated secondary effluent was irradiated at different dose of 0.2~1.0 kGy by 1 MeV, ELV-4 Model electron beam accelerator. It is interesting to note that a 100 % reduction in total coliforms and total colonies were achieved until a dose of approximately 0.8 kGy. Even at low dose of 0.2 kGy, the total coliforms and total colonies were successfully inactivated to the level of satisfying the new effluent discharge guideline. Besides disinfection of total coliforms, approximately a 50% removal in biochemical oxygen demand was pronounced at a dose of 0.2 kGy. More than 20 % removal in suspended solids and turbidity was also observed at a dose of 1.0 kGy. The application of electron beam irradiation appeared to be one of options to reuse sewage treatment effluent as agricultural or industrial water.

Dose Characteristics for IORT Applicator of ML-15MDX Electron Beam (ML-15MDX 술중조사용 Applicator에 의한 전자선선량 특성)

  • Choi, Tae-Jin;Lee, Ho-Joon;Kim, Yeung-Ae;Kim, Jin-Hee;Kim, Ok-Bae
    • Radiation Oncology Journal
    • /
    • v.11 no.2
    • /
    • pp.455-461
    • /
    • 1993
  • Experimental measurements of dose characteristics with pentagonal applicator at nominal energy of 4, 6, 9, 12 and 15 MeV electron beam were performed for intraoperative radiotherapy (IORT) in ML-15MDX linear accelerator. This paper presents the percent depth dose, surface dose, beam flatness and output factors of using the IORT applicator in different electron beam energy. The output factor showed as a 24 percent higher in IORT applicator than that of reference $10{\times}10cm^2$ applicator. The surface dose of using the IORT applicator showed 7.7 and 2.7 percent higher than that of reference field in 4 and 15 MeV electron beam, respectively. In our experiments, the variation of percent depth dose was very small but the output factor and flatnees at 0.5 cm depth have showed a large value in IORT applicator.

  • PDF

Dose Effect of Phytosanitary Irradiation on the Postharvest Quality of Cut Flowers

  • Kwon, Song;Kwon, Hye Jin;Ryu, Ju Hyun;Kim, Yu Ri
    • Journal of People, Plants, and Environment
    • /
    • v.23 no.2
    • /
    • pp.171-178
    • /
    • 2020
  • The present study was conducted to determine the effects of electron beam irradiation on the postharvest quality of cut flowers. Cut flowers were irradiated with electron beam at 100, 200, 400, 600, 800, 1,000, and 2,000 Gy with a 10 MeV linear electron beam accelerator to evaluate their irradiation tolerance. Postharvest quality was determined by monitoring fresh weight loss, flower longevity, flower diameter, flowering rate, visual quality of flowers and leaves, and chlorophyll content. Cut flowers showed a radiation-induced damage with increasing the irradiation dose. Flower longevity and fresh weight of cut flowers decreased when the irradiation dose was increased. Flower bud opening was also inhibited in a dose-dependent manner. The effective irradiation doses for 10% reduction of postharvest quality (ED10) values were 144.4, 451.6, and 841.2 Gy in the 'Medusa' lily, 'Montezuma' carnation, and 'Rosina White' eustoma, respectively. Although tolerance of cut flowers to electron beam irradiation vary according to species, cultivars, or maturity stage conditions, it is conceivable that 'Montezuma' carnation and 'Rosina White' eustoma could be tolerated and maintained overall postharvest quality up to 400 Gy, the generic irradiation dose approved by the Animal and Plant Health Inspection Service (APHIS) and the International Plant Protection Convention (IPPC) for postharvest phytosanitary treatments.

Study of Pulse Generator used Inverter HV Power Supply (XFEL를 위한 기존의 펄스전원공급장치 개선 연구)

  • Park, S.S.;Kim, S.H.;Kim, S.C.;Hwang, J.Y.;Han, Y.J.;Chio, J.H.;Kim, H.G.
    • Proceedings of the KIEE Conference
    • /
    • 2005.07c
    • /
    • pp.2146-2148
    • /
    • 2005
  • The 2.5GeV linac of the Pohang Light Source(PLS) is planed to be converted to a XFEL. The PAL XFEL requires a new 1.2-GeV linac that will be combined to the existing linac to increase a beam energy upto 3.7GeV. This stability is mainly determined by a low level RF drive system and klystron-modulators. The stability level of the modulator has to be improved 10 times better to meet the pulse stability of 0.02 %. The regulation methods such as traditional de-Qing and precision inverter charging technology are reviewed to find out suitable upgrade scheme of the modulators. In order to obtain electron beam of the consequently stability for XFEL linac, the pulse-to-pulse beam voltage regulation is less than +/-0.5%. To get the reliable stability of the modulator which is less than +/-0.2%, a charging section is improved in a modulator which has been operated with inverter power supply and de-Q'ing.

  • PDF

A Study on Effective Source-Skin Distance using Phantom in Electron Beam Therapy

  • Kim, Min-Tae;Lee, Hae-Kag;Heo, Yeong-Cheol;Cho, Jae-Hwan
    • Journal of Magnetics
    • /
    • v.19 no.1
    • /
    • pp.15-19
    • /
    • 2014
  • In this study, for 6-20 MeV electron beam energy occurring in a linear accelerator, the authors attempted to investigate the relation between the effective source-skin distance and the relation between the radiation field and the effective source-skin distance. The equipment used included a 6-20 MeV electron beam from a linear accelerator, and the distance was measured by a ionization chamber targeting the solid phantom. The measurement method for the effective source-skin distance according to the size of the radiation field changes the source-skin distance (100, 105, 110, 115 cm) for the electron beam energy (6, 9, 12, 16, 20 MeV). The effective source-skin distance was measured using the method proposed by Faiz Khan, measuring the dose according to each radiation field ($6{\times}6$, $10{\times}10$, $15{\times}150$, $20{\times}20cm^2$) at the maximum dose depth (1.3, 2.05, 2.7, 2.45, 1.8 cm, respectively) of each energy. In addition, the effective source-skin distance when cut-out blocks ($6{\times}6$, $10{\times}10$, $15{\times}15cm^2$) were used and the effective source-skin distance when they were not used, was measured and compared. The research results showed that the effective source-skin distance was increased according to the increase of the radiation field at the same amount of energy. In addition, the minimum distance was 60.4 cm when the 6 MeV electron beams were used with $6{\times}6$ cut-out blocks and the maximum distance was 87.2 cm when the 6 MeV electron beams were used with $20{\times}20$ cut-out blocks; thus, the largest difference between both of these was 26.8 cm. When comparing the before and after the using the $6{\times}6$ cut-out block, the difference between both was 8.2 cm in 6 MeV electron beam energy and was 2.1 cm in 20 MeV. Thus, the results showed that the difference was reduced according to an increase in the energy. In addition, in the comparative experiments performed by changing the size of the cut-out block at 6 MeV, the results showed that the source-skin distance was 8.2 cm when the size of the cut-out block was $6{\times}6$, 2.5 cm when the size of the cut-out block was $10{\times}10$, and 21.4 cm when the size of the cut-out block $15{\times}15$. In conclusion, it is recommended that the actual measurement is used for each energy and radiation field in the clinical dose measurement and for the measurement of the effective source-skin distance using cut-out blocks.