• Title/Summary/Keyword: Draft detecting sensor

Search Result 1, Processing Time 0.014 seconds

Experimental Research for Traction force Sensor Development on Drawing Exercise Medical Instrument (재활 및 교정을 위한 견인운동치료기의 견인측정센서 개발에 관한 실험적 연구)

  • Lee, Sang-sik;Park, Won-yeop;Lee, Choong-ho
    • The Journal of Korea Institute of Information, Electronics, and Communication Technology
    • /
    • v.2 no.2
    • /
    • pp.3-8
    • /
    • 2009
  • The traction system has been mainly used for rehabilitation and correction of patients with spine or gait diseases in orthopedics or at home. Some problems could occur in human body when patients forced their training using the traction system. So it needs to measure a traction force and control the training time. However, most of products on market have no sensor measuring traction force. Thus we designed and made a sensor detecting traction force using strain gauge, amplifier for transition to output signal and experiment devices for performance test. We carried out experiment of a sensor detecting a traction force and measured electric responses of it with respect to traction loads. Maximum error was within about 1% for experiments in static condition and the average error was about 0.7% for experiments in dynamic condition. We concluded that it is possible to use the developed sensor for measurement of traction force since the maximum output variation of a sensor detecting a traction force was about 0.3% in $0^{\circ}C-60^{\circ}C$ temperature condition.

  • PDF