The purpose of this study was to analyze the effect on the selectivity on of high-voltage rectification device that measured the performance of the grid, and the contrast improvement ability (K factor) by measuring the scattered radiation content of the transmitted X-rays. The scattered radiation generated when the X-ray flux comes from the diagnostic X-ray generator that passes through an object. Targeting four different rectifications of X-ray generators, the mean value of the tube voltage and the tube current was measured in order to maximize the accuracy of the generating power dose within the same exposure condition. Using fluorescence meter, the content of the scattered rays that are transmitted through the acrylic was measured depending on the grid usage. When grid is not used, the content of the scattered rays was the lowest (34.158%) with the single-phase rectifier, was increased with the inverter rectifier (37.043%) and the three-phase 24-peak rectification method (37.447%). The difference of the scattered radiation content of each device was significant from the lowest 0.404% to the highest 3.289% while using 8:1 grid, the content of the scattered ray was the lowest with the single content of the scattered ray was the lowest with the single-phase rectifier (18.258%), was increased with the rectifier (25.502%) and the 24-peaks rectification (24.217%). Furthermore, there was difference up to content 7.244% to the lowest content 1.285% within three-phase 24-peaks rectification, inverter rectifications, and single-phase rectifier depending on the selectivity of the grid. Drawn from the statistical analysis, there was a similar relationship between the contrast improvement factor and the K factor. As a result, the grid selectivity and the contrast were increased within the single-phase rectifier rather than the constant voltage rectifier.
To evaluation of patients who have shoulder impingement syndrome is by diagnostic radiography. Shoulder impingement is a problem which occurs in young, active individuals as well as older individuals. In fact, the pain is probably caused by repetitive stress placed on the shoulder joint either through recreational activities of your occupation. Impingement series approach to radiographic examination of the shoulder is take five projections. First anteroposterior oblique projection. Second standard anteroposterior projection. Third superoinferior axial projection. Fourth supraspinatus outlet projection offers a view of the outlet of the supraspinatus tendon unit as it passes under the coraacromial arch. Fifth anteroposterior $30^{\circ}$ caudal projection will adequately demonstrate the anterior acromial spur or ossification in the coraacromial ligament and more reliable to demonstrate spurring of the anterior acromion than supraspinatus outlet projection. This decreased the need for additional radiographic veiws, reduces the patient's exposure to x - ray radiation and decreases use of film. This can lower the cost of the evaluation and improve patient satisfaction.
This study was carried out to investigate radiographical and operating conditions of X-ray units and exposure doses to patients during chest radiography, so that the results could provide basic data used for reducing the exposure dose and for providing the diagnostic information with better quality. The conditions and exposure doses of 100 X-ray units mainly used for chest radiography were examined and also 100 radiological technologists mainly handling those apparatus at 76 medical facilities in Pusan were surveyed using a questionnaire from October 1 to December 31 in 1995. The following results were obtained from the study : 1. It was found that most units were capable of taking a high tube voltage radiography by showing 67% of the units equipped with the maximum tube voltage of 150 kV, 94% with more than 500 mA for the rating capacity and 85% with the full wave type of a signal phase. 2. For actual chest radiographical conditions, however, 80% of the units were operated at $60{\sim}100\;kVp$ and only 14% at 100 kVp and over for the high tube voltage. 3. The average exposure time was less than 0.1 second, and eighty four percent of the units adapted the X-ray tube currents ranging from 200 to 300 mA, 80% the focus-film distances between 180 and 210 cm, and 63% the focus sizes of more than 2.0 mm. 4. Most units(98%) employed additional filters made of aluminum, 75% the thickness of filters less than 2.0 mm, and only 2 units the compound filters. 5. Ortho chromatic system was only adopted in 13% of screen film system for the units, and 73% used the grid ratio at 8 : 1 for the low tube voltage during chest radiography. 6. The average exposure dose of all X-ray units during chest radiography was $371\;{\mu}Sv$ with a difference of about 16 times between the minimum to the maximum, and $386\;{\mu}Sv$ both at hospitals and at health centers, followed by $380\;{\mu}Sv$ at general hospitals and $263\;{\mu}Sv$ at university hospitals without showing any statistically significant differences. In conclusion, since patients during chest radiography at medical facilities in Pusan exposed to high levels of radiation, it is recommended that appropriate added filters and grids necessary for the high tube voltage radiography and high-speed screen systems should be adopted and used as soon as possible in order to reduce exposure dose to the patients.
Ji, Hoon;Han, Su Chul;Baek, Jong Hyeun;Lee, Dong Hoon;Park, Seungwoo
Journal of Electrical Engineering and Technology
/
v.13
no.2
/
pp.936-942
/
2018
The diagnostic multi-leaf collimator preventing unnecessary dose from entering into patients during the diagnostic examination was made in this study. The movement of the entire 50 leaves was embodied with the group of 25 ones thereof configured in a pair facing each other on the left and right of the median line. Dimensions of the length, width, and height of each shielding leaf were $5{\times}0.5{\times}0.5cm^3$ resulting in the maximum boost field of $10{\times}10cm^2$. The material of multi-leaf collimator had the excellence on the machinability with the use of the SKD-11 alloy tool steel having the high wear resistance against frequent movement, and it was devised to control both-side's shielding leaves by moving 2 motors unlike existing remedial multi-leaf collimator that use as many motors as the number of 50 shielding leaves. Thereafter, the transmission dose of leaves, cross-leaf leakage dose, and inter-leaf leakage dose were measured by the developed multi-leaf collimator attached to X-ray equipment. An ionization chamber was used to detect doses there from, and the comparative analysis was carried out by means of the radiographic film that was easy to detect the dose leakage in between each leaf. Results obtained from the test conducted in comparative analysis yielded approximately 98%, 96%, and 94% of shielding efficiency realized at each level of energy of 80kV, 100kV, and 120kV it was confirmed there was no dose leakage resulted from the varied level of irradiation energy. Thus the multi-leaf collimator to be developed based on this study is thought that it could fully reduce the unnecessary dose to patients in the diagnostic test and the shielding efficiency thereof is expected to be increasing if it is made in a miniaturized form with a way of increasing the thickness of each leaf later for an extended application to general diagnostic purposes.
Journal of Korean Academy of Oral and Maxillofacial Radiology
/
v.22
no.1
/
pp.87-94
/
1992
The purpose of this study was to evaluate the effect of scatter radiation to dental x-ray film with long time-exposure in the different structures of the tooth, by using pinhole camera. For this study, pinhole camera, skull with tooth, and pocket dosimeter were used. The radiation with 70 and 90kVp and exposure time (minimum: 2.5 min., maximum 10 hrs.) was projected to the film in the pinhole camera. And density of the obtained x-ray film was measured with densitometer. In the intra-oral film taking, the amount of exposure of the scatter radiation affecting the thyroid gland area was measured with the dosimeter at the thyroid gland. The density of radiographs was compared in radiation projected with or without the metal cone of dental machine. The effect of the back scatter radiation to the film was also evaluated when the lead foil was removed. The obtained results were as follows: 1. A pinhole camera was a valuable device for locating the source of x-ray. 2. The scatter radiation affected the dental x-ray film when the radiation source was exposed. more than 5 hours'. In that case, the density of the scatter radiation could be observed visually. 3. The scatter radiation caused by short exposure of dental radiation didn't affect the diagnostic quality of the dental x-ray film. 4. The differences of densities between the tooth and the soft tissue according to exposure time showed 0.16 in 5 hours' exposure & 0.17 in 10 hours' exposure at 70 kVp & 0.12 in 5 hours' exposure & 0.13 in 10 hours' exposure at the 90kVp. 5. The differences of densities between the tooth and the soft tissue according to kVp showed no difference between 5 hours' exposure of tooth at 70 kVp and soft tissue at 90 kVp, but showed 0.05 high density in tooth when 10 hours' exposure at 90 kVp. 6. No difference of density was on radiographs taken with or without dental machine cone. 7. Back scatter radiation was recorded image of radiographs for only 3 min. 8. The amounts of the scatter radiation exposed to the thyroid gland in intraoral film taking were 1.12 mr in upper anterior, 0.55 mr in upper posterior, 2.75 mr in lower anterior, and 1.92 mr in lower posterior teeth.
Kim, Hyun-jin;Joo, Yeong-chul;Choi, Jae-ho;Lim, Woo-taek
Journal of radiological science and technology
/
v.42
no.6
/
pp.423-428
/
2019
The Precise measurement of the knee's Congruence angle is required for diagnosis of patella dislocation. The purpose of this study is to consider the distortion diagram and usefulness about the test of the bilateral side and one side through the evaluating congruence angle from Merchant view of patellofemoral joint to central X-ray. We've evaluated the sulcus angle and Congruence angle following central X-ray's changes when we took Merchant view of patellofemoral joint aimed to right lower limb from human whole body phantom. The subject of the evaluation was classified as normal group and varus group, and the varus group has experimented with External rotation of legs as 15° and 30°. When normal groups result as 0, 7.5, 15 cm, it was measured as 17.25 ± 1.34°, 19.60 ± 1.41°, 20.55 ± 1.77° each. The gap between minimal and maximum angle was 3.3°, and the value was shown as getting bigger when it further away from the centeral X-ray. When Congruence angle in 15° varus group was 0, 7.5, 15 cm, it was each 16.45 ± 1.34°, 17.10 ± 0.99°, 17.80 ± 1.13°. And when Congruence angle in 30° varus group was 0, 7.5, 15 cm, it was measured each 18.35 ± 1.63°, 18.95 ± 1.06°, 19.60 ± 1.41°. The difference between minimum and maximum of angle in 15° varus group and 30° varus group was each 1.35° and 1.25°, the angles have shown as increasing the further away from the center. The patellofemoral joint showed that the congruence angle increases as the further away from Central X-ray, and also it is judged that is possible to change the congruence angle by a degree of varus knee. Thus, accurate measurement of the congruence angle is shown to advantage that methods for examining by split each side at two times than methods for examining both sides at a time. Therefore, it is expected to helpful on the diagnostic side of patella dislocation and subluxation.
In diagnostic radiology, the imaging system has been changed from film/screen to digital system. However, the method for removing scatter radiation such as anti-scatter grid has not kept pace with this change. Therefore, authors have devised the indirect flat panel detector (FPD) system with net-like lead in substrate layer which can remove the scattered radiation. In clinical context, there are many radiographic examinations with angulated incident X-ray. However, our proposed FPD has net-like lead foil so the vertical lead foil to the angulate incident X-ray would have bad effect on its performance. In this study, we identified the effect of vertical/horizontal lead foil component on the novel system's performance and improved the structure of novel system for clinical usage with angulated incident X-ray. Grid exposure factor and image contrast were calculated to investigate various structure of novel system using Monte Carlo simulation software when the incident X-ray was tilted ($0^{\circ}$, $15^{\circ}$, and $30^{\circ}$ from the detector plane). More photons were needed to obtain same image quality in the novel system with vertical lead foil only then the system with horizontal lead foil only. An optimal structure of novel system having different heights of its vertical and horizontal lead foil component showed improved performance compared with the novel system in a previous study. Therefore, the novel system will be useful in a clinical context with the angulated incident X-ray if the height and direction of lead foil in the substrate layer are optimized as the condition of conventional radiography.
Kim, Dong-Jin;Chae, Jong-Sang;Yoo, Chae-Min;Lee, Bae-Won
Journal of radiological science and technology
/
v.41
no.1
/
pp.7-12
/
2018
Patients who visit the emergency room with urinary stones have difficulty lying down in a supine position due to severe pain when performing the KUB test. The purpose of this study was to find methods to reduce the patients' pain and image distortion, and obtain medical images with high diagnostic values. After checking the standard classification of disease and cause of death, the target group consisted of 121 patients who had clearly distinguished stones from computed tomography. Patients with stones in the ureteralvesical junction were excluded. Qualitative image evaluation was performed by confirming the location of the stone in the computed tomography images. and evaluated the rate of visual discrimination of stones possible through KUB and abdominal plain X-ray. Quantitative image evaluation was performed on the KUB, abdominal plain X-ray images. The transverse process of the first lumbar vertebrae served as the standard point, and the length from this point to the lower part of the stone was measured. Results from looking at the rate of visual discrimination of stones possible through KUB and abdominal plain X-ray showed: 94 patients (77.6%) for KUB images and 91 patients (75.2%) for computed tomography images. The standard deviation for KUB and abdominal X-ray was 3 (2.4%). Comparing and analyzing the location from KUB images and abdominal plain X-ray images, the stone position was 10.1 mm in the kidney, 10.5 mm in the ureteropelvic junction, and 9.7 mm in the ureters. It was shown that the stone moved 10 mm on average with significant statistical difference (P<0.05). In cases where the pain is so severe that it is impossible to perform the test in the supine position, an alternative may be to check the stone position by performing a modified KUB test by having the patient stand in a vertical position. In the future, this will provide convenience to both the examiner and the patient when performing the examination, and it will contribute with its reproducibility.
In veterinary medicine, most radiographic images are obtained by restraining patients, inevitably exposing the restrainer to secondary scattered radiation. Radiation exposure can result in stochastic reactions such as cancer and genetic effects, as well as deterministic reactions such as skin burns, cataracts, and bone marrow suppression. Radiation-shielding equipment, including aprons, thyroid shields, eyewear, and gloves, can reduce radiation exposure. However, the risk of radiation exposure to the upper arms, face, and back remains, and lead aprons and thyroid shields are heavy, restricting movement. We designed a new radiation-shielding system and compared its shielding ability with those of conventional radiation-shielding systems. We hypothesized that the new shielding system would have a wider radiation-shielding range and similar shielding ability. The radiation exposure dose differed significantly between the conventional and new shielding systems in the forehead, chin, and bilateral upper arm areas (p < 0.001). When both systems were used together, the radiation-shielding ability was better than when only one system was used at all anatomical locations (p < 0.01). This study suggests that the new radiation-shielding system is essential and convenient for veterinary radiation workers because it is a step closer to radiation safety in veterinary radiography.
In this pauper, the back, forward, side and $45^{\circ}$ oblique scatter dose were measured the X-ray exposure conditions 60, 80, 100, 120kV, FFD 100cm, FS $20\times20cm$, toward the $25\times25cm\times10\sim20cm$ of solid water, paraffin and MiX-DP phantom, and Pb, Cu, Al, and styrofoam meterials, by the electrometer and 5.3 cc ionization chamber. The obtained results are summarized as following. 1. The percentage depth dose(PDD) at the range of the diagnostic x-ray energy were appeared 50 % depth dose at the 2 cm depth with 60 kV, and 5 cm depth with 120 kV X-ray, 10 % depth dose at the 10 cm depth with 60 kV and 14 cm depth with 120 kV X-ray, 5 % below depth dose at the 20 cm depth. 2. The back scatter dose which were generated the surface of Pb, Cu and Al metal plates were 10 % below, and than the back scatter dose at the Pb plate were a most amount of these which were about 10 %, and were appeared the order of Cu and Al. 3. The percentage forward scatter were appeared from 50 % to 65 %, and the more phantom thicknees become, the more forward scatter were increased with the ratio of 5 % per 5 cm thickness. 4. The percentage back scatter which were generated the tissue equivalence meterials solid water, paraffin and MiX-DP were from 20 % to 40 %, and than the back scatter dose at the solid water were a mest amount of those, and paraffin and MiX-DP were appeared with the next values. 5. The percentage $90^{\circ}$ lateral and $45^{\circ}$ oblique side scatter dose were measured from 4 % to 12 %. a most amount of scatter dose which were generated from the patient in radiography were the forward scatter, the next values were the back scatter, the third values were the $90^{\circ}$ lateral scatter.
본 웹사이트에 게시된 이메일 주소가 전자우편 수집 프로그램이나
그 밖의 기술적 장치를 이용하여 무단으로 수집되는 것을 거부하며,
이를 위반시 정보통신망법에 의해 형사 처벌됨을 유념하시기 바랍니다.
[게시일 2004년 10월 1일]
이용약관
제 1 장 총칙
제 1 조 (목적)
이 이용약관은 KoreaScience 홈페이지(이하 “당 사이트”)에서 제공하는 인터넷 서비스(이하 '서비스')의 가입조건 및 이용에 관한 제반 사항과 기타 필요한 사항을 구체적으로 규정함을 목적으로 합니다.
제 2 조 (용어의 정의)
① "이용자"라 함은 당 사이트에 접속하여 이 약관에 따라 당 사이트가 제공하는 서비스를 받는 회원 및 비회원을
말합니다.
② "회원"이라 함은 서비스를 이용하기 위하여 당 사이트에 개인정보를 제공하여 아이디(ID)와 비밀번호를 부여
받은 자를 말합니다.
③ "회원 아이디(ID)"라 함은 회원의 식별 및 서비스 이용을 위하여 자신이 선정한 문자 및 숫자의 조합을
말합니다.
④ "비밀번호(패스워드)"라 함은 회원이 자신의 비밀보호를 위하여 선정한 문자 및 숫자의 조합을 말합니다.
제 3 조 (이용약관의 효력 및 변경)
① 이 약관은 당 사이트에 게시하거나 기타의 방법으로 회원에게 공지함으로써 효력이 발생합니다.
② 당 사이트는 이 약관을 개정할 경우에 적용일자 및 개정사유를 명시하여 현행 약관과 함께 당 사이트의
초기화면에 그 적용일자 7일 이전부터 적용일자 전일까지 공지합니다. 다만, 회원에게 불리하게 약관내용을
변경하는 경우에는 최소한 30일 이상의 사전 유예기간을 두고 공지합니다. 이 경우 당 사이트는 개정 전
내용과 개정 후 내용을 명확하게 비교하여 이용자가 알기 쉽도록 표시합니다.
제 4 조(약관 외 준칙)
① 이 약관은 당 사이트가 제공하는 서비스에 관한 이용안내와 함께 적용됩니다.
② 이 약관에 명시되지 아니한 사항은 관계법령의 규정이 적용됩니다.
제 2 장 이용계약의 체결
제 5 조 (이용계약의 성립 등)
① 이용계약은 이용고객이 당 사이트가 정한 약관에 「동의합니다」를 선택하고, 당 사이트가 정한
온라인신청양식을 작성하여 서비스 이용을 신청한 후, 당 사이트가 이를 승낙함으로써 성립합니다.
② 제1항의 승낙은 당 사이트가 제공하는 과학기술정보검색, 맞춤정보, 서지정보 등 다른 서비스의 이용승낙을
포함합니다.
제 6 조 (회원가입)
서비스를 이용하고자 하는 고객은 당 사이트에서 정한 회원가입양식에 개인정보를 기재하여 가입을 하여야 합니다.
제 7 조 (개인정보의 보호 및 사용)
당 사이트는 관계법령이 정하는 바에 따라 회원 등록정보를 포함한 회원의 개인정보를 보호하기 위해 노력합니다. 회원 개인정보의 보호 및 사용에 대해서는 관련법령 및 당 사이트의 개인정보 보호정책이 적용됩니다.
제 8 조 (이용 신청의 승낙과 제한)
① 당 사이트는 제6조의 규정에 의한 이용신청고객에 대하여 서비스 이용을 승낙합니다.
② 당 사이트는 아래사항에 해당하는 경우에 대해서 승낙하지 아니 합니다.
- 이용계약 신청서의 내용을 허위로 기재한 경우
- 기타 규정한 제반사항을 위반하며 신청하는 경우
제 9 조 (회원 ID 부여 및 변경 등)
① 당 사이트는 이용고객에 대하여 약관에 정하는 바에 따라 자신이 선정한 회원 ID를 부여합니다.
② 회원 ID는 원칙적으로 변경이 불가하며 부득이한 사유로 인하여 변경 하고자 하는 경우에는 해당 ID를
해지하고 재가입해야 합니다.
③ 기타 회원 개인정보 관리 및 변경 등에 관한 사항은 서비스별 안내에 정하는 바에 의합니다.
제 3 장 계약 당사자의 의무
제 10 조 (KISTI의 의무)
① 당 사이트는 이용고객이 희망한 서비스 제공 개시일에 특별한 사정이 없는 한 서비스를 이용할 수 있도록
하여야 합니다.
② 당 사이트는 개인정보 보호를 위해 보안시스템을 구축하며 개인정보 보호정책을 공시하고 준수합니다.
③ 당 사이트는 회원으로부터 제기되는 의견이나 불만이 정당하다고 객관적으로 인정될 경우에는 적절한 절차를
거쳐 즉시 처리하여야 합니다. 다만, 즉시 처리가 곤란한 경우는 회원에게 그 사유와 처리일정을 통보하여야
합니다.
제 11 조 (회원의 의무)
① 이용자는 회원가입 신청 또는 회원정보 변경 시 실명으로 모든 사항을 사실에 근거하여 작성하여야 하며,
허위 또는 타인의 정보를 등록할 경우 일체의 권리를 주장할 수 없습니다.
② 당 사이트가 관계법령 및 개인정보 보호정책에 의거하여 그 책임을 지는 경우를 제외하고 회원에게 부여된
ID의 비밀번호 관리소홀, 부정사용에 의하여 발생하는 모든 결과에 대한 책임은 회원에게 있습니다.
③ 회원은 당 사이트 및 제 3자의 지적 재산권을 침해해서는 안 됩니다.
제 4 장 서비스의 이용
제 12 조 (서비스 이용 시간)
① 서비스 이용은 당 사이트의 업무상 또는 기술상 특별한 지장이 없는 한 연중무휴, 1일 24시간 운영을
원칙으로 합니다. 단, 당 사이트는 시스템 정기점검, 증설 및 교체를 위해 당 사이트가 정한 날이나 시간에
서비스를 일시 중단할 수 있으며, 예정되어 있는 작업으로 인한 서비스 일시중단은 당 사이트 홈페이지를
통해 사전에 공지합니다.
② 당 사이트는 서비스를 특정범위로 분할하여 각 범위별로 이용가능시간을 별도로 지정할 수 있습니다. 다만
이 경우 그 내용을 공지합니다.
제 13 조 (홈페이지 저작권)
① NDSL에서 제공하는 모든 저작물의 저작권은 원저작자에게 있으며, KISTI는 복제/배포/전송권을 확보하고
있습니다.
② NDSL에서 제공하는 콘텐츠를 상업적 및 기타 영리목적으로 복제/배포/전송할 경우 사전에 KISTI의 허락을
받아야 합니다.
③ NDSL에서 제공하는 콘텐츠를 보도, 비평, 교육, 연구 등을 위하여 정당한 범위 안에서 공정한 관행에
합치되게 인용할 수 있습니다.
④ NDSL에서 제공하는 콘텐츠를 무단 복제, 전송, 배포 기타 저작권법에 위반되는 방법으로 이용할 경우
저작권법 제136조에 따라 5년 이하의 징역 또는 5천만 원 이하의 벌금에 처해질 수 있습니다.
제 14 조 (유료서비스)
① 당 사이트 및 협력기관이 정한 유료서비스(원문복사 등)는 별도로 정해진 바에 따르며, 변경사항은 시행 전에
당 사이트 홈페이지를 통하여 회원에게 공지합니다.
② 유료서비스를 이용하려는 회원은 정해진 요금체계에 따라 요금을 납부해야 합니다.
제 5 장 계약 해지 및 이용 제한
제 15 조 (계약 해지)
회원이 이용계약을 해지하고자 하는 때에는 [가입해지] 메뉴를 이용해 직접 해지해야 합니다.
제 16 조 (서비스 이용제한)
① 당 사이트는 회원이 서비스 이용내용에 있어서 본 약관 제 11조 내용을 위반하거나, 다음 각 호에 해당하는
경우 서비스 이용을 제한할 수 있습니다.
- 2년 이상 서비스를 이용한 적이 없는 경우
- 기타 정상적인 서비스 운영에 방해가 될 경우
② 상기 이용제한 규정에 따라 서비스를 이용하는 회원에게 서비스 이용에 대하여 별도 공지 없이 서비스 이용의
일시정지, 이용계약 해지 할 수 있습니다.
제 17 조 (전자우편주소 수집 금지)
회원은 전자우편주소 추출기 등을 이용하여 전자우편주소를 수집 또는 제3자에게 제공할 수 없습니다.
제 6 장 손해배상 및 기타사항
제 18 조 (손해배상)
당 사이트는 무료로 제공되는 서비스와 관련하여 회원에게 어떠한 손해가 발생하더라도 당 사이트가 고의 또는 과실로 인한 손해발생을 제외하고는 이에 대하여 책임을 부담하지 아니합니다.
제 19 조 (관할 법원)
서비스 이용으로 발생한 분쟁에 대해 소송이 제기되는 경우 민사 소송법상의 관할 법원에 제기합니다.
[부 칙]
1. (시행일) 이 약관은 2016년 9월 5일부터 적용되며, 종전 약관은 본 약관으로 대체되며, 개정된 약관의 적용일 이전 가입자도 개정된 약관의 적용을 받습니다.