치과용 임플란트 지대주 나사 구조에 관한 연구 (Characteristics of Abutment Screw Structure for Dental Implant)

  • 송종법;최일경;정효경;권순홍;권순구;박종민;김종순;정성원;최원식
    • 한국산업융합학회 논문집
    • /
    • 제20권2호
    • /
    • pp.169-176
    • /
    • 2017
  • Dental implants are required to have biomechanical functions and biostability in order to perform authoring, pronunciation, and aesthetic functions in the oral cavity. In terms of biostability, pure titanium for medical have good biostability and no rejection in the alveolar bone. with appropriate strength in terms of strength as well as biocompatibility. In recent years, various surgical methods and devices have been developed to improve the convenience and safety of the procedure. However, as the number of procedures increases, the screw loosening of the abutment screw connecting the artificial root and the abutment There are many reports of artificial root and abutment fracture. Fig. 1 is an example of a case where the upper part of the abutment screw is arbitrarily modified to remove the abutment by the abutment fracture due to the loosening of the abutment screw. The fundamental cause of abduction of the abutment screw is caused by the slight movement due to the lowering of the retention force of the abutment screw. It is necessary to minimize loosening of the abutment screw to avoid problems such as fracture during the period of using the implant. The purpose of this study is to investigate the structure of the abutment screw to prevent the loosening of the abutment screw by forming 0.5mm slot.

치과용 임플란트 지대주나사의 조임체결력에 따른 지지골과 지대주나사의 유한요소법 응력 분석 (A finite element stress analysis on the supporting bone and abutment screw by tightening torque of dental implant abutment screw)

  • 이명곤
    • 대한치과기공학회지
    • /
    • 제42권2호
    • /
    • pp.99-105
    • /
    • 2020
  • Purpose: A study analysed the stress distribution of abutment screw and supporting bone of fixture by the tightening torque force of the abutment screw within clinical treatment situation for the stability of the dental implant prosthesis. Methods: The finite element analysis was targeted to the mandibular molar crown model, and the implant was internal type 4.0 mm diameter, 10.0 mm length fixture and abutment screw and supporting bone. The occlusal surface was modeled in 4 cusps and loaded 100 N to the buccal cusps. The connection between the abutment and the fixture was achieved by combining three abutment tightening torque forces of 20, 25, and 30 Ncm. Results: The results showed that the maximum stress value of the supporting bone was found in the buccal cortical bone region of the fixture in all models. The von Mises stress value of each model showed 184.5 MPa at the 20 Ncm model, 195.3 MPa in the 25 Ncm model, and 216.5 MPa in the 30 Ncm model. The contact stress between the abutment and the abutment screw showed the stress value in the 20 Ncm model was 201.2 MPa, and the 245.5 MPa in the 25 Ncm model and 314.0 MPa in the 30 Ncm model. Conclusion: The increase of tightening force within the clinical range of the abutment screw of the implant dental prosthesis was found to have no problem with the stability of the supporting bone and the abutment screw.

Clinical study on screw loosening in dental implant prostheses: a 6-year retrospective study

  • Lee, Ki-Young;Shin, Kyung Su;Jung, Ji-Hye;Cho, Hye-Won;Kwon, Kyung-Hwan;Kim, Yu-Lee
    • Journal of the Korean Association of Oral and Maxillofacial Surgeons
    • /
    • 제46권2호
    • /
    • pp.133-142
    • /
    • 2020
  • Objectives: In this study, we determined the incidence and pattern of screw loosening in patients who received dental implants. Materials and Methods: Patients who received implants between January 2008 and October 2013 and completed their prosthetic rehabilitation were evaluated for the incidence, frequency, and onset of screw loosening using dental charts and radiographs. The association between each factor and screw loosening was analyzed using the chi-square test and a multivariate analysis with binary logistic regression models (P<0.05). Results: Total 1,928 implants were placed in 837 patients (448 males, 389 females), whose follow-up period after loading varied from 0.25 to 70 months (mean period, 31.5 months). Screw loosening occurred in 7.2% of implants. Most cases occurred less than six months after loading. Among those, 22.3% experienced recurrent screw loosening. Screw loosening was most common in the molar region (8.5%) and frequently associated with an implant diameter of ≥5 mm (14.2%). External implant-abutment connections (8.9%) and screw-retained implant prostheses (10.1%) showed higher incidence of problems than internal implant-abutment connections and cement-retained implants, respectively. Screw loosening was most common in implant prostheses with single crowns (14.0%). Conclusion: Within the limits of the current study, we conclude that the incidence of screw loosening differs significantly according to the position of implant placement, the type of implant and manufacturer, implant diameter, the type of implant-abutment connection, the type of retention in the implant prosthesis, and the type of implant prosthesis.

지대주 나사 조임 토크가 맞춤형 지대주 임플란트 시스템의 연결부 안정성에 미치는 영향: 3차원 유한 요소 해석 (Effect of tightening torque on the connection stability of a custom-abutment implant system: 3D finite element analysis)

  • 홍민호
    • 대한치과기공학회지
    • /
    • 제43권3호
    • /
    • pp.99-105
    • /
    • 2021
  • Purpose: This study aims to examine the stress distribution effect of tightening torques of different abutment screws in a custom-abutment implant system on the abutment-fixture connection interface stability using finite element analysis. Methods: The custom-abutment implant system structures used in this study were designed using CATIA program. It was presumed that the abutment screws with a tightening torque of 10, 20, and 30 N·cm fixed the abutment and fixture. Furthermore, two external loadings, vertical loading and oblique loading, were applied. Results: When the screw tightening torque was 10 N·cm, the maximum stress value of the abutment screw was 287.2 MPa that is equivalent to 33% of Ti-6Al-4V yield strength. When the tightening torque was 20 N·cm, the maximum stress value of the abutment screw was 573.9 MPa that is equivalent to 65% of Ti-6Al-4V yield strength. When the tightening torque was 30 N·cm, the maximum stress value of the abutment screw was 859.6 MPa that is similar to the Ti-6Al-4V yield strength. Conclusion: As the screw preload rose when applying each tightening torque to the custom-abutment implant system, the equivalent stress increased. It was found that the tightening torque of the abutment influenced the abutment-fixture connection interface stability. The analysis results indicate that a custom-abutment implant system should closely consider the optimal tightening torque according to clinical functional loads.

Correlation between microleakage and screw loosening at implant-abutment connection

  • Sahin, Cem;Ayyildiz, Simel
    • The Journal of Advanced Prosthodontics
    • /
    • 제6권1호
    • /
    • pp.35-38
    • /
    • 2014
  • PURPOSE. This study aimed to evaluate the correlation between microleakage and screw loosening at different types of implant-abutment connections and/or geometries measuring the torque values before and after the leakage tests. MATERIALS AND METHODS. Three different abutment types (Intenal hex titanium, internal hex zirconium, morse tapered titanium) with different geometries were connected to its own implant fixture. All the abutments were tightened with a standard torque value then the composition was connected to the modified fluid filtration system. After the measurements of leakage removal torque values were re-measured. Kruskal-wallis test was performed for non-parametric and one-way ANOVA was performed for parametric data. The correlation was evaluated using Spearman Correlation Test (${\alpha}=0.05$). RESULTS. Significantly higher microleakage was found at the connection of implant-internal hex zirconium abutment. Observed mean torque value loss was also significantly higher than other connection geometries. Spearman tests revealed a significant correlation between microleakage and screw loosening. CONCLUSION. Microleakage may provoke screw loosening. Removing torque values rationally decrease with the increase of microleakage.

나사풀림과 나사파절이 동반된 실패한 임플란트 고정성 보철물의 회복증례 (Reconstruction of failed fixed implant prosthesis accompanied by abutment screw loosening and fracture : A case report)

  • 배은빈;신영근;전영찬;정창모;윤미정;이소현;허중보
    • 대한치과의사협회지
    • /
    • 제55권7호
    • /
    • pp.457-466
    • /
    • 2017
  • In the case of failed fixed implant prosthesis accompanied by abutment screw fracture, fractured screw fragment must be removed to use the existing implant fixtures. A 61-year-old male patient, who had a failed maxillary fixed implant prosthesis accompanied by three abutment screw fracture, hoped to reconstruct the maxillary implant prosthesis, while maintaining the existing implant fixtures. To use the existing implant fixtures, fractured screw fragments were removed. A maxillary implant overdenture using available existing implants was planned. Bar-attachment with Locator was used for implant splinting, denture stability, and retention. Final impression was taken after treatment of peri-implantitis. Jaw relation registration was taken to evaluate available interarch space for bar-attachment. After fabricating bar-attachment, centric relation was taken. Implant overdenture using bar-attachment with Locator was delivered after wax-denture evaluation. This case report showed that a satisfactory clinical result was achieved by implant overdenture using existing implant fixtures in a maxillary edentulous patient.

  • PDF

Evaluation of Fatigue Fracture Life for TiN Coated Abutment Screw in Dental Implant

  • Choe, Han-Cheol;Chung, Chae-Heon
    • 한국표면공학회:학술대회논문집
    • /
    • 한국표면공학회 2012년도 추계총회 및 학술대회 논문집
    • /
    • pp.115-116
    • /
    • 2012
  • In this study, fitness and fatigue test were performed to estimate the coating effects of abutment screw for implant system after fatigue test. The purpose of this study was to investigate fatigue fracture phenomena of dental implant fixture used titanium nitride coated abutment screw under cyclic load.

  • PDF

시멘트리스 나사 유지형 임플란트 시스템을 이용한 보철물 수복 증례 (Prosthetic restorations with cementless screw-retained implant systems: a case report)

  • 정영해;김임선
    • 대한치과기공학회지
    • /
    • 제45권4호
    • /
    • pp.124-130
    • /
    • 2023
  • This study aims to describe the clinical experience of single and bridge crowns fabricated using a cementless screw-retained implant prosthesis system. In the case of single crown (#37), regular link (HDL) was used, and bridge crowns (#15~#24), (#26~#27), (#17~#14) (#24~#26) were fabricated by selecting regular link and short link considering the vertical height. One abutment was hex shaped to ensure that it could be mounted while preventing insertion and prosthesis rotation. The advantages of cementless implant prosthesis include shorter chair time and periodic care, strong retention with LINK abutment, safety from inflammation, bacterial infection, and complications due to peri-implant cement, and high patient satisfaction. Dentists should double-check the position of the implant fixture and dental technicians should continuously manage the fit of the link and prosthesis with digital equipment to reduce screw loosening and fractures.

임플란트-지대주-나사의 적합에 관한 연구 (A STUDY ON THE FIT OF THE IMPLANT-ABUTMENT-SCREW INTERFACE)

  • 김낙형;정재헌;손미경;백대화
    • 대한치과보철학회지
    • /
    • 제41권4호
    • /
    • pp.503-518
    • /
    • 2003
  • Statement of problem : There have been previous studies about considerable variations in machining accuracy and consistency in the implant-abutment-screw interfaces. Purpose : The purpose of this study was to evaluate the machining accuracy and consistency of implant/abutment/screw combinations on two randomly selected implants from each of four manufactures. Material and methods : In this study, screws were respectively used to secure a cemented abutment, to a hexlock implant fixture ; teflon coated titanium alloy screw(Torq-Tite) and titanium alloy screw in Steri--Oss system, gold-plated gold-palladium alloy screw(Gold-Tite) and titanium alloy screw in 3i system gild screw ana titanium screw in AVANA Dental Implant system, and titanium screws in Paragon System. The implants were perpendicularly mounted in polymethyl methacrylate autopolymerizing acrylic resin block(Orthodontic resin, Densply International Inc. USA) by use of dental surveyer. Each abutment screw was secured to the implant with recommended torque value using a digital torque controller. Each screw was again tightened after 10 minutes. All samples were cross sectioned with grinder-polisher unit(Omnilap 2000 SBT Inc) after embeded in liquid unsaturated polyester (Epovia, Cray Valley Inc) Results : There were the largest gaps in the neck areas of screws in hexagonal extension implants which were examined in this study. The leading edge of the abutment screw thread (superior surface) was in contact with the implant body thread, and the majority of the contacting surfaces were localized to the middle portion of the mating threads. Considerable variation in the contacting surfaces was noted in the samples evaluated. Amounts of contact in the abutment screw thread were larger for assemblies with Gold-Tite screw, gold alloy screw. Torq-Tite screw than those with titanium screws. The findings of intimate contact between the screw and screw seat were seen in all samples, regardless of manufacturers. However, microgap between the head and lateral neck surface of the screw and the abutment could be dectected in all samples. The findings of intimate contact between the platform of the implant and the bottom of the abutment were consistent in all samples, regardless of manufacturers. However, microgaps between the lateral surface of external hex of the fixture and the abutment could be dectected in all samples. Conclusion : Considerable variations in machining accuracy and consistency were noted in the samples and the implant-abutment-screw interfaces were incomplete. From the results of this study, further development of the system will be required, including improvements in pattern design.

Comparison of implant component fractures in external and internal type: A 12-year retrospective study

  • Yi, Yuseung;Koak, Jai-Young;Kim, Seong-Kyun;Lee, Shin-Jae;Heo, Seong-Joo
    • The Journal of Advanced Prosthodontics
    • /
    • 제10권2호
    • /
    • pp.155-162
    • /
    • 2018
  • PURPOSE. The aim of this study was to compare the fracture of implant component behavior of external and internal type of implants to suggest directions for successful implant treatment. MATERIALS AND METHODS. Data were collected from the clinical records of all patients who received WARANTEC implants at Seoul National University Dental Hospital from February 2002 to January 2014 for 12 years. Total number of implants was 1,289 and an average of 3.2 implants was installed per patient. Information about abutment connection type, implant locations, platform sizes was collected with presence of implant component fractures and their managements. SPSS statistics software (version 24.0, IBM) was used for the statistical analysis. RESULTS. Overall fracture was significantly more frequent in internal type. The most frequently fractured component was abutment in internal type implants, and screw fracture occurred most frequently in external type. Analyzing by fractured components, screw fracture was the most frequent in the maxillary anterior region and the most abutment fracture occurred in the maxillary posterior region and screw fractures occurred more frequently in NP (narrow platform) and abutment fractures occurred more frequently in RP (regular platform). CONCLUSION. In external type, screw fracture occurred most frequently, especially in the maxillary anterior region, and in internal type, abutment fracture occurred frequently in the posterior region. placement of an external type implant rather than an internal type is recommended for the posterior region where abutment fractures frequently occur.