Usefulness of CTAC Shift Revision Method of Artifact by Diaphragm in PET/CT (PET/CT 검사에서 횡격막에 의한 인공물의 CTAC Shift 보정방법의 유용성)

  • Ham, Jun Cheol;Kang, Chun Koo;Cho, Seok Won;Bahn, Young Kag;Lee, Seung Jae;Lim, Han Sang;Kim, Jae Sam;Lee, Chang Ho
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.17 no.1
    • /
    • pp.71-75
    • /
    • 2013
  • Purpose: Currently, decrement revision using LDCT is used in PET/CT. But cold artifacts are often found in decrement revision image by mismatch between LDCT image and Emission image near diaphragm due to patient's respiration. This research studied reduction of cold artifact by patient's respiration using CTAC Shift among revision methods. Materials and Methods: From March to September in 2012, 30 patients who had cold artifacts by respiration were targeted using PET/CT Discovery 600 (GE Healthcare, MI, USA) equipment. Patients with cold artifacts were additionally scan in diaphragm area, and the image shown cold artifacts at whole body test were revised using CTAC Shift. Cold artifacts including image, additional scan image and CTAC Shift revision image were evaluated as 1~5 points with naked eye by one nuclear medicine expert, 4 radiotechnologists with over 5 year experience. Also, standard uptake value of 3 images was compared using paired t-test. Results: Additional scan image and CTAC Shift revision image received relatively higher score in naked eye evaluation than cold artifacts including image. The additional scan image and CTAC Shift revision image had high correlation as the results of ANOVA test of standard uptake value and did not show significant difference. Conclusion: When cold artifacts are appeared by patient's respiration at PET/CT, it causes not only patient inconvenience but troubles in test schedule due to extra radiation exposure and time consumption by additional scan. But if CTAC Shift revision image can be acquired with out additional scan, it is considered to be helped in exact diagnosis without unnecessary extra radiation exposure and additional scan.

  • PDF

The Research of Comparison Evaluation on the Decline in Artifact Using Respiratory Gating System in PET-CT (PET-CT 검사 시 호흡동조 시스템을 이용한 인공물 감소에 대한 비교 평가)

  • Kim, Jin-Young;Lee, Seung Jae;jung, Suk;Park, Min-Soo;Kang, Chun-Goo;Im, Han-Sang;Kim, Jae-Sam
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.19 no.2
    • /
    • pp.63-67
    • /
    • 2015
  • Purpose Among various causes that influence image quality degradation, various methods for decrease in Artifact occurred by respiration of patients are being used. Among them, this study intended to evaluate CTAC Shift correction method and additional scan compare to the Scan(Q static scan) using respiratory gated system. Materials and Methods This study was conducted on 10 patients, and used PET-CT Discovery 710 (GE Healthcare, MI, USA) and Varian's RPM system. 5.18 Mbq per kg of $^{18}F$-FDG was injected on patients, asked them to take a rest for 1 hour in the bed, and conducted test after urination. Images were visualized through Q static scan, CTAC Shift correction method, Additional scan based on the Whole body scan(WBS) with Artifact. Decrease in Artifact was compared in each image, conducted Gross Evalution, and measured changes of SUVmax. Results For image obtained through the CTAC Shift correction method through WBS with Artifact, 12~56%, Q static scan image showed 17~54% of change rate and Additional Scan showed -27~46% of change rate. In Blind Test, the CTAC Shift correction image showed the highest point with 4 points, Q static scan image showed 3.5 points, and Additional scan image showed 3.4 points. The standardized WBS scan through Oneway ANOVA and three types of Scan method showed significant difference(p<0.05), and did not show significant difference between the three Scan methods(p>0.05). However, the three Scan methods showed significant difference in Blind test. Conclusion Additional scan and Q static scan require more time than the CTAC Shift correction method, there is concern about excessive exposure to patients by CT rescan and Q static scan is difficult to apply on patients with inconsistent respiration or irregular respiration cycle due to pain. For CTAC Shift correction method, limited correction is possible and the range is limited as well. It is considered as a useful method of improving diagnostic value when hospitals use the system appropriately and develop various advantageous factors of each method.

  • PDF