• Title/Summary/Keyword: Analytical Spectral Devices (ASD)

Search Result 1, Processing Time 0.015 seconds

Development of relative radiometric calibration system for in-situ measurement spectroradiometers (현장관측용 분광 광도계의 상대 검교정 시스템 개발)

  • Oh, Eunsong;Ahn, Ki-Beom;Kang, Hyukmo;Cho, Seong-Ick;Park, Young-Je
    • Korean Journal of Remote Sensing
    • /
    • v.30 no.4
    • /
    • pp.455-464
    • /
    • 2014
  • After launching the Geostationary Ocean Color Imager (GOCI) on June 2010, field campaigns were performed routinely around Korean peninsula to collect in-situ data for calibration and validation. Key measurements in the campaigns are radiometric ones with field radiometers such as Analytical Spectral Devices FieldSpec3 or TriOS RAMSES. The field radiometers must be regularly calibrated. We, in the paper, introduce the optical laboratory built in KOSC and the relative calibration method for in-situ measurement spectroradiometer. The laboratory is equipped with a 20-inch integrating sphere (USS-2000S, LabSphere) in 98% uniformity, a reference spectrometer (MCPD9800, Photal) covering wavelengths from 360 nm to 1100 nm with 1.6 nm spectral resolution, and an optical table ($3600{\times}1500{\times}800mm^3$) having a flatness of ${\pm}0.1mm$. Under constant temperature and humidity maintainance in the room, the reference spectrometer and the in-situ measurement instrument are checked with the same light source in the same distance. From the test of FieldSpec3, we figured out a slight difference among in-situ instruments in blue band range, and also confirmed the sensor spectral performance was changed about 4.41% during 1 year. These results show that the regular calibrations are needed to maintain the field measurement accuracy and thus GOCI data reliability.