Dosimetric Properties of LiF:Mg,Cu,Na,Si TL pellets (LiF:Mg,Cu,Na,Si TL 소자의 선량계적 특성)

  • Nam, Young-Mi;Kim, Jang-Lyul;Chang, Si-Young
    • Journal of Radiation Protection and Research
    • /
    • v.26 no.1
    • /
    • pp.7-12
    • /
    • 2001
  • Sintered LiF:Mg,Cu,Na,Si thermoluminescence (TL) pellets were developed for application in radiation dosimetry. In the present study, the TL dosimetric properties of LiF:Mg,Cu,Na,Si TL pellets have been investigated for emission spectrum, dose response, energy response, and fading characteristics. LiF:Mg,Cu,Na,Si TL pellets were made by using a sintering process, that is, pressing and heat treatment from TL powders. Photon irradiations for the experiments were carried out using X-ray beams and a $^{137}Cs$ gamma source at the Korea Atomic Energy Research Institute (KAERI). The average energies and the dose were in the range of 20-662 keV and $10^{-6}-10^{-2}\;Gy$, respectively. The glow curves were measured with a manual type TLD reader(System 310, Teledyne) at a constant nitrogen flux and a linear heating rate. For a constant heating rate of $5^{\circ}C\;s^{-1}$, the main dosimetric peak of glow curve appeared at $234^{\circ}C$, the activation energy was 2.34 eV and frequency factor was $1.00{\times}10^{23}$. TL emission spectrum is appeared at the blue region centered at 410 nm. A linearity of photon dose response was maintained up to 100 Gy. The photon energy responses relative to $^{137}Cs$ response were within ${\pm}20%$ at overall photon energy region. The fading of TL sensitivity of the pellets stored at the room temperature was not found for one year.

  • PDF

A Study on X-Ray Fluorescence Analysis of Ta$_2O_5,\;Nb_2O_5,\;SnO_2$ and ZrO$_2$ in Tin-slag Samples (희석 파라미터법에 의한 주석슬랙중 Ta$_2O_5,\;Nb_2O_5,\;SnO_2$ 및 ZrO$_2$의 X-선형광분석에 관한 연구)

  • Young Sang Kim
    • Journal of the Korean Chemical Society
    • /
    • v.29 no.3
    • /
    • pp.265-270
    • /
    • 1985
  • Ta$_2$O$_5$, Nb$_2$O$_5$, SnO$_2$ and ZrO$_2$ in tin-slag samples were determined by X-ray fluorescence spectrometry using the dilution parameter method and the analytical results were compared with the data obtained by standard calibration curve method. Tin-slag samples and one standard sample having very similar composition to the tin-slags were diluted with a proper diluent (La$_2$O$_3$) to the ratio of 1: 1, 1 : 2, 1 : 3 and 1 : 4. After measuring the X-ray intensities of original and diluted samples, the values of dilution parameters were calculated by using the dilution parameter equation. Without any calibration curve, the analytical results were calculated from the equation including the dilution parameter term. The results were in good agreement with the reference data by the standard calibration curve method.

  • PDF

X-Ray Fluorescence Analysis of $Ta_2O_5,\;Nb_2O_5$ and ZrO_2$ in Tin-Slag Samples using Simple Dilution Method (주석 슬랙중 $Ta_2O_5,\;Nb_2O_5$ZrO_2$의 단일희석법을 이용한 X-선 형광분석)

  • Young-Sang Kim;Hak Je Sung
    • Journal of the Korean Chemical Society
    • /
    • v.28 no.5
    • /
    • pp.293-301
    • /
    • 1984
  • -ray fluorescence analysis of $Ta_2O_5$, $Nb_2O_5$ and $ZrO_2$ in tin-slag samples using the simple dilution method was studied. The method is to correct mathematically the calibration curve to the linear line by the dilution. One synthesized standard having similar composition to the sample and tin slag samples were diluted with anhydrous $Li_2B_4O_7$ at the level of 1%, 2% and 3% of the sample content respectively. The diluted samples were fused at $1150^{\circ}C$ for 30 minutes and these glass beads were finely ground and pelletized. Measuring the X-ray intensities with these pellets, analytical results were calculated by the equation derived from J. Scherman's equation for the characteristic X-ray intensity of an element. Analytical results agreed with the reference values obtained by the standard calibration method within allowable error range and were reproducible.

  • PDF

Determination of Eu(III) by Fluorescence Spectrometry (형광분광법에 의한 Eu(Ⅲ)의 정량)

  • Lee, Sang Hak;Han, Jong Hwan;Choi, Sang Seob
    • Journal of the Korean Chemical Society
    • /
    • v.42 no.3
    • /
    • pp.285-291
    • /
    • 1998
  • Methods to determine EU(Ⅲ) ion in aqueous solution by fluorescence spectrometry based upon the ligand sensitized fluorescence of Eu(Ⅲ)-terephthalic acid (TPA) complex ion have been studied. The effects of excitation wavelength, pH, concentration of TPA and emission wavelength on the fluorescence intensity were investigated. The fluorescence intensity of the Eu(Ⅲ) complex ion was further increased with addition of trioctylphosphine oxide (TOPO). In this case Triton X-100 was used to dissolve TOPO in aqueous solution. The calibration curve for Eu(Ⅲ) was linear over the range from $1.0{\times}10^{-6}M\;to\;4.0{\times}10^{-4}M$ and the detection limit was $1.0{\times}10^{-6}M$ under the experimental conditions of 256 nm, 5.6, $3.5{\times}10^{-4}$M$ and 615 nm for excitation wavelength, pH, concentration of TPA and emission wavelength, respectively. When TOPO was added to the Eu(Ⅲ)-TPA system, the concentration range of linear response and the detection limit were $1.0 {\times}10^{-9}M\;to\;1.0{\times}10^{-4}M,\;1.0{\times}10^{-7}M,$ respectively under the experimental conditions of 284 nm, 4.4 and $1.0{\times} 10^{-4}M$ for excitation wavelength, pH and concentration of TOPO, respectively. Effects of interferences from various cations for the determination of Eu(Ⅲ) ion were also investigated.

  • PDF

Determination of Eu(III) by Fluorescence Spectrometry using Fiber Optic Sensor (광섬유센서를 이용한 Eu(III)의 형광분광법적 정량)

  • Lee, Sang Hak;Lee, Yoon Hee;Yang, Seung Tae;Choi, Sang Seob
    • Analytical Science and Technology
    • /
    • v.11 no.5
    • /
    • pp.409-412
    • /
    • 1998
  • The analytical method to determine europium(III) ion in aqueous solution by fluorescence spectroscopy based upon the conformational change of calmodulin in the presence of the analyte has been studied. The fiber optic chemical sensor used in this study was constructed by entrapping a fluorescein-labeled calmodulin solution, EGTA, buffer solution at the common end of a bifurcated fiber optic bundle by means of a dialysis membrane. The calibration curve to determine europium(III) ion was obtained when concentration of calmodulin, concentration of EGTA, Tris-HCl buffer solution, pH, excitation wavelength and fluorescence wavelength were $5.0{\times}10^{-5}M$, 0.50 mM, 5.0 mM, 7.0, 495 nm and 520 nm, respectively. The detection limit was $1.0{\times}10^{-11}M$ and the working range of the calibration curve for the sensor was $1.0{\times}10^{-11}M{\sim}1.0{\times}10^{-9}M$. The response time was 15 minutes. For the determination of europium(III) ion by the present method, $Na^+$ and $K^+$ ions did not interfere but $Ca^{2+}$ ion seriously interfered.

  • PDF

Fabrication of $Li_2B_4O_7$ Series Single-Crystal TLDs and their TL properties ($Li_2B_4O_7$ 계열 단결정 TLD 소자의 제작과 특성)

  • Park, Myeong-Hwan;Park, Kang-Soo
    • Journal of radiological science and technology
    • /
    • v.28 no.1
    • /
    • pp.1-7
    • /
    • 2005
  • High-quality single crystals of pure $Li_2B_4O_7$ as well as $Li_2B_4O_7$ doped with Cu, Mn and Mg impurities (1.0mol%, respectively) have been grown from the melt of $Li_2CO_3+2B_2O_3$ by Czochralski method in platinum crucibles. To study the thermoluminescent properties, $Li_2B_4O_7$ series single crystal TLDs were made by cutting in the size of $4{\times}5{\times}1\;mm^3$. The glow curves show two or three peaks which can be easily deconvoluted. It is observed that room temperature($20{\sim}30^{\circ}C$) fadings of the dosimetric peaks of $Li_2B_4O_7$ series single crystal TLDs were about 10 % for 30 days. The relative photon energy response for $Li_2B_4O_7$ series single crystal TLDs were about 85 % when the responses were normalized to that measured with $^{60}Co\;(1.25\;MeV)\;{\gamma}-rays$. The measured data are in a good agreement with theoretical ones. The $Li_2B_4O_7$ series single crystal TLDs fabricated in this work can be used for monitoring personal and environmental radioactivity.

  • PDF

Influence of Temperature on the Photosynthetic Responses of Benthic Diatoms: Fluorescence Based Estimates (온도가 저서규조류 광합성 반응에 미치는 영향: 형광을 이용한 추정)

  • Yun, Mi-Sun;Lee, Choon-Hwan;Chung, Ik-Kyo
    • The Sea:JOURNAL OF THE KOREAN SOCIETY OF OCEANOGRAPHY
    • /
    • v.14 no.2
    • /
    • pp.118-126
    • /
    • 2009
  • Benthic diatoms are very important primary producers in understanding estuary ecosystems and their productions are largely varied by their photo-physiological characteristics. The short-term effects of increased temperature on the photosynthetic and photo-physiological characteristics of cultured different species of benthic diatoms (Navicula sp., Nitzschia sp., Cylindrotheca closterium, and Pleurosigma elongatum) were investigated by measuring their PSII-fluorescence kinetics using a Diving-PAM. Photosynthesis versus irradiance curves were measured every two hours at six different temperatures (10, 15, 20, 25, 30, and $35^{\circ}C$) for twenty-four hour. The effective quantum yield of PSII ($\Phi_{PSII}$) for most of the species showed a decreasing trend with increased temperature. The relative maximum electron transport rate (rETRmax) was significantly increased up to the optimum temperature level and then sharply decreased. Relative to the values of other parameters, the maximum light use coefficient ($\alpha$) was not substantially changed at lower temperature levels (<$30^{\circ}C$) but significantly decreased only at higher temperatures (30 and $35^{\circ}C$). The light saturation coefficient ($E_K$) mirrored the rETRmax temperature response. In regards to the temperature acclimation abilities of the four species with time, Navicula sp. and C. closterium acclimated to short-term changes in temperature through their photo-physiological adjustments.

Weighted Interpolation Method Using Supplementary Filter (보조필터를 이용한 가중치 보간방법)

  • Jang, In-Gul;Lee, Jae-Kyung;Chung, Jin-Gyun
    • Journal of the Institute of Electronics Engineers of Korea SP
    • /
    • v.48 no.3
    • /
    • pp.119-124
    • /
    • 2011
  • Interpolation filters are widely used in many communication and multimedia applications. Polynomial interpolation computes the coefficients of the polynomial according to the input information to obtain the interpolated value. Recently, FIR interpolation method using supplementary filters was proposed to improve the performances of polynomial interpolation methods. In this paper, by combining a weighting factor approach with the supplementary filter method, we propose a weighted interpolation method which can be efficiently used to compute the maximum or minimum values of a given curve using only a restricted number of sample values. With application to the interpolation of normal distribution curves used in XRF systems, it is shown that the proposed approach exhibits improved performances compared with conventional interpolation methods.

Study for Energy Transfer from Rhodamine 6G to Malachite Green Using Time Correlated Single Photon Counting Method (시간상관 단일광자 계수법에의한 Rhodamine 6G에서 Malachite Green으로의 에너지 전달 연구)

  • Kim, Hyun-Soo;Eom, Hyo-Soon;Choi, Gyu-Kwan;Jeong, Hong-Sik;Kim, Ung
    • Korean Journal of Optics and Photonics
    • /
    • v.2 no.4
    • /
    • pp.203-208
    • /
    • 1991
  • We investigated the nonradiative energy transfer process from Rhodamine 6G to Malachite Green in ethylen glycol solvent using time correlated single photon counting system equipped with a modelocked Ar ' laser. The reduced concentration and critical transfer distance for various acceptor concentration were obtained by using a full-fitting analysis of the fluorescence decay curves. We found that Huber model is more suitable than Forster model and the influence of energy migration through the dipole-dipole interaction becomes more significant for the low acceptor concentrations relative to the donor concentration($5\times 10^4$mol/l).

  • PDF

Evaluation the Output Dose of Linear Accelerator Photon Beams by Blind Test with Dose Characteristics of LiF:Mg,Cu,P TLD (LiF:Mg,Cu,P 열형광선량계의 선량특성을 이용한 눈가림법에 의한 출력선량 평가)

  • Choi, Tae-Jin;Lee, Ho-Joon;Yie, Ji-Won;Oh, Young-Gi;Kim, Jin-Hee;Kim, Ok-Bae
    • Progress in Medical Physics
    • /
    • v.20 no.4
    • /
    • pp.308-316
    • /
    • 2009
  • To achieve the accurate evaluation of given absorbed dose from output dose of linear accelerator photon beam through investigate the characteristics of LiF:Mg,Cu,P TLD powder. This experimental TL phosphor is performed with a commercial LiF:Mg,Cu,P powder (Supplied by PTW) and TL reader (LTM, France). The TLD was exposed to 6 MV X rays of linear accelerator photon beam with range 15 to 800 cGy in blind dose at two hospitals. The dose evaluation of TLD was through the experimental algorithms which were dose dependency, dose rate dependency, fading and powder weight dependency. The glow curve has shown the three peaks which are 110, 183 and 232 degrees of heating temperature and the main dosimetric peak showed highest TL response at 232 high temperature. In this experiments, the LiF:Mg,Cu,P phosphor has shown the 2.5 eV of electron trap energy with a second order. This experiments guided the dose evaluation accuracy is within 1% +2.58% of discrepancy. The TLD powder of LiF:Mg,Cu,P was analyzed to dosimetric characterists of electron captured energy and order by glow shape, and dose-TL response curve guided the accuracy within 1.0+2.58% of output dose discrepancy.

  • PDF