• Title/Summary/Keyword: 평행판 점도계

Search Result 1, Processing Time 0.016 seconds

A parallel plate viscometer for blood viscosity measurement (혈액점도 측정용 평행판 점도계)

  • Donggil Seo;Kyung Hyun Ahn;Jihoon Kang;Sangsoo Park
    • The Journal of the Convergence on Culture Technology
    • /
    • v.9 no.4
    • /
    • pp.331-335
    • /
    • 2023
  • As the viscosity of the blood increases, the blood becomes more sticky and difficult to flow, so the possibility of thrombosis increases and the probability of ischemic cerebral infarction increases. The importance of measuring blood viscosity has recently been emphasized for the prevention of circulatory system diseases, and the need for a viscometer capable of easily and accurately measuring blood viscosity has emerged. In this study, the measured values of a viscosity standard solution and an artificial blood by a parallel-plate viscometer ARS-Medi were compared with the those by Ares-G2 of TA instrument, which is internationally recognized for its accuracy and reliability. The viscosity of N44 standard solution, which is a Newtonian solution, was almost perfectly matched between the two instruments at all shear rates. In the case of an artificial blood, which is a non-Newtonian solution, the measured values between the two instruments showed a difference of about 10% at the lowest shear rate 1 rad/s; however, at a clinically significant shear rate of 10 rad/s or higher, the measured values between them were consistent within the error range. We expect that ARS-Medi, a newly developed parallel-plate viscometer for blood, using disposable plates, will be very useful in clinical practice as it improves the convenience and hygiene of blood viscosity measurement.