• Title/Summary/Keyword: 비접속형 측정장비

Search Result 1, Processing Time 0.014 seconds

Usefulness of Non-Invasive Measurement Tool on Performance Evaluation of Inverter Type X-ray Unit (인버터식 X선장치의 성능평가 시 비접속형 측정기의 유용성)

  • Kang, Se-Sik;Kim, Chang-Soo;Ko, Sung-Jin
    • Journal of radiological science and technology
    • /
    • v.31 no.2
    • /
    • pp.123-127
    • /
    • 2008
  • Purpose: As the demand of a simple and precise method increases to evaluate the performance of the inverter type x-ray unit, we evaluated the usefulness of the recently-introduced X-ray Multi-Function Test Device (moldel : Xi (unfors)-prestige). Method: We compared the performance of X-ray Multi-Function Test Device (XMFTD) which is non-inveasive type device with the performance of Dynalyzer III that has been most widely used inveasive type measure device. Result: X-ray output dose was increased a little in the XMFTD, but both devices were below the performance evaluation standard, 0.002 in the output reproducibility. Linearity of XMFTD were below 0.1 which means that Dynalyzer III showed more excellency in linearity. As the the accuracy of exposure factor, 1.8 and 2 tube voltage, 2.01 and 2.3 tube current were measured. The exposure time was also measured by 0.01 sec ${\pm}10%$. Both devices were within the acceptance of performance evaluatioin standard. Conclusion: We proved the usefulness of X-ray Multi-Function Test Device (model: Xi (unfors)-prestige) to evaluated the performance on reproductibility and linearity of X-ray output and accuracy of exposure factor of inverter type unit.

  • PDF