• Title/Summary/Keyword: 동시합산 보정

Search Result 2, Processing Time 0.015 seconds

Evaluate the usefulness of Coincidence Summing Correction Factors for Cylinder and Extended Source Penelope Simulation (실린더 및 확장 소스 PENELOPE 시뮬레이션에 대한 동시합성보정 계수 유용성 평가)

  • Jang, Eun-Sung;Chang, Bo-Seok
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.6
    • /
    • pp.821-831
    • /
    • 2021
  • In order to calibrate energy and efficiency using the PENELOPE Code, a PENELOPE simulation was performed using a volume source. Here, we want to verify peak efficiency and usefulness by performing simultaneous measurement and correction. calculate the coincident sum correction for all volumes, first subdivide the volumes of the cylinder and the four Marinelli beakers into three heights again. Therefore, the simultaneous measurement correction coefficient in three areas and the simultaneous measurement correction coefficient for the entire volume source are calculated as output. At low energies, the j value for each source volume (50-300 ml) is small and increases significantly in the high energy range. Simulation results showed good agreement within 2.5% for all source volumes except for 50 ml and 300 ml, which were up to 4%. This means that the correction for the simultaneous measurement effect during measurement is effective. In addition. Based on this, it can be confirmed that there is an advantage to improve the detection efficiency when measuring various sources and environmental samples.

The Effect of Applying Self-absorption and Coincidence Summation Correction when Measuring Environmental Samples (환경시료 측정 시 자체흡수 및 동시합성 보정 적용 효과)

  • Eun-Sung Jang;Byung-In Min
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.4
    • /
    • pp.531-539
    • /
    • 2023
  • Self-absorption is the most important factor affecting the accuracy of gamma spectroscopy measurements in environmental samples. In particular, it is affected by other factors such as the chemical composition of the sample, geometric shape, thickness, density, atomic number, distance between the sample and detector, energy of the emitted gamma photon, and humidity coefficient or percentage in the sample. To test the calibration method, a 450 ml CRM standard source (9 nuclide) Marinelli beaker was used. Five soil samples among environmental samples were measured by density by applying the corrected values. Therefore, it can be seen that the self-absorption value is more effective for somewhat large and low photon energy. In the case of environmental samples, it was confirmed that the overall energy peak efficiency through self-absorption of the source greatly depends on the density of the sample.