Purpose:By evaluating the dependence of the tray transmission factor (tray factor) on collimator setting and tray thickness, we determined the validity of the clinically used single tray factor for standard radiation field size (10${\times}$10 $\textrm{cm}^2$). Methods and Materials:For each X ray energies (6 and 10 MV), outputs were measured by using 5 steps of tray thickness (0, 6, 8, 10, 12 mm) and 7 steps of radiation field size (5${\times}$5, 10${\times}$10, 15${\times}$15, 20${\times}$20, 25${\times}$25, 30${\times}$30, 35${\times}$35 $\textrm{cm}^2$) at 10 cm phantom depth. Outputs were measured in both 'with tray' and 'without tray' conditions by using radiation with the same monitor units, and the tray factors were determined by the ratios of the two outputs. To evaluate the validity of a single tray factor obtained for standard radiation field, we analyzed the pattern of the field sizes in cases treated at our hospital in 2002. Results : In the 6 MV X-ray, the increases in the tray factor between the standard field (l0${\times}$10 $\textrm{cm}^2$) and the largest field (35${\times}$35 $\textrm{cm}^2$) were 0.517%, 0.835%, 1.058%, 1.066% in 6, 8, 10, and 12 mm thickness tray, respectively. In the 10 MV X-ray, the increases in the fray factor between the standard field (10${\times}$10 $\textrm{cm}^2$) and the largest field (35${\times}$35 $\textrm{cm}^2$) were 0.517%, 0.836%, 1.058%, 1.066% in 6, 8, 10, 12 mm thickness tray, respectively. In a major portion of clinical cases, when the field size was smaller than 20${\times}$20 $\textrm{cm}^2$, the tray factor was in good agreement with the standard tray factor. However, in cases where the field sizes were 30${\times}$30 $\textrm{cm}^2$ and 35${\times}$35 $\textrm{cm}^2$, the error could exceed 1.0%. Conclusion:The tray factor increased with increasing field size or decreasing tray thickness. The difference of tray factor between the small field and the large field increased with increasing tray thickness. Furthermore, the standard tray factor was valid in most clinical cases except for when the field size was greater than 30${\times}$30 $\textrm{cm}^2$, wherein the error could exceed 1.0%.