Comparative evaluation for leaf position accuracy according to gantry angle variation in MLC quality assurance using electronic portal imaging device(EPID) and GafChromic EBT3 film (전자포탈영상장치(EPID)와 GafChromic EBT3 film을 이용한 다엽콜리메이터 정도관리 시 갠트리 각도 변화에 따른 엽의 위치 정확성 비교 평가)

  • Yang, Myung Sic;Park, Ju Kyeong;Lee, Seung Hun;Lee, Sun Young;Kim, Jung Soo;Kwon, Hyoung Cheol;Kim, Yang Su
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.29 no.2
    • /
    • pp.83-91
    • /
    • 2017
  • Objectives: The purpose of this study was to evaluate the error of the leaf position accuracy of the MLC due to the gravity effect according to the gantry angle by using picket fence test using EPID and GafChromic EBT3 film. Materials and Methods: A 5 cm solid phantom was placed on the table and the SAD was set to 100 cm. The EBT3 film was placed exactly over the solid phantom and covered a 1.5 cm solid phantom and the picket fence test was performed. The EPID was measured under the same conditions as the EBT3 film at SID 100 cm. The gantry angles were measured at $0^{\circ}$, $90^{\circ}$, $180^{\circ}$ and $270^{\circ}$ in order to evaluate the position of the MLC according to the gantry angle. For the geometric evaluation of the MLC, the leaf position accuracy of the MLC was analyzed using the analysis program. Results: In case of EPID, when the gantry angle was changed to $0^{\circ}$, $90^{\circ}$, $180^{\circ}$, $270^{\circ}$, the difference of the position errors of the leaves was 0.18 mm, 0.31 mm, 0.20 mm, 0.26 mm on the average and the maximum values of the errors were respectively 0.44 mm, 0.54 mm, 0.34 mm, 0.44 mm. In case of EBT3 film, when the gantry angle was changed to $0^{\circ}$, $90^{\circ}$, $180^{\circ}$, $270^{\circ}$, the difference of the position errors of the leaves was 0.19 mm, 0.21 mm, 0.19 mm, 0.31 mm on the average and the maximum values of the errors were respectively 0.35 mm, 0.45 mm, 0.36 mm, 0.48 mm. Conclusion: In this study, we analyzed the position error of the leaf of the MLC according to the gantry angle, and confirmed the position error of the leaf by gravity effect. As a result of comparing the leaf position accuracy using EPID and EBT3 film according to the variation of gantry angle, a larger error occurred in the error analysis method using EPID than that of EBT3 film. Therefore, in the case of IMRT based on MLC, as well as verification of accurate dosimetry should be conducted, it is considered that the quality control and verification for the precise operation of the MLC will be needed. and it is necessary to compare and verify the method of analysis.

  • PDF

Evaluation of Clinical Risk according to Multi-Leaf Collimator Positioning Error in Spinal Radiosurgery (척추 방사선수술 시 다엽콜리메이터 위치 오차의 임상적 위험성 평가)

  • Dong‑Jin Kang;Geon Oh;Young‑Joo Shin;Jin-Kyu Kang;Jae-Yong Jung;Boram Lee
    • Journal of radiological science and technology
    • /
    • v.46 no.6
    • /
    • pp.527-533
    • /
    • 2023
  • The purpose of this study is to evaluate the clinical risk of spinal radiosurgery by calculating the dose difference due to dose calculation algorithm and multi-leaf collimator positioning error. The images acquired by the CT simulator were recalculated by correcting the multi-leaf collimator position in the dose verification program created using MATLAB and applying stoichiometric calibration and Monte Carlo algorithm. With multi-leaf collimator positioning error, the clinical target volume (CTV) showed a dose difference of up to 13% in the dose delivered to the 95% volume, while the gross tumor volume (GTV) showed a dose difference of 9%. The average dose delivered to the total volume showed dose variation from -8.9% to 9% and -10.1% to 10.2% for GTV and CTV, respectively. The maximum dose delivered to the total volume of the spinal cord showed a dose difference from -14.2% to 19.6%, and the dose delivered to the 0.35 ㎤ volume showed a dose difference from -15.5% to 19.4%. In future research, automating the linkage between treatment planning systems and dose verification programs would be useful for spinal radiosurgery.

Report on the External Audits Conducted by Korean Society of Medical Physics (한국의학물리학회 선형가속기 외부 품질관리 실시 현황보고)

  • Huh, Hyun Do;Cho, Kwang Hwan;Cho, Sam Ju;Choi, Sang Hyoun;Kim, Dong Wook;Hwang, Ui-Jung;Kim, Ki Hwan;Min, Chul Kee;Choi, Tae Jin;Oh, Young Kee;Lee, Seoung Jun;Park, Dahl;Park, Sung-Kwang;Ji, Young Hoon
    • Progress in Medical Physics
    • /
    • v.24 no.4
    • /
    • pp.315-322
    • /
    • 2013
  • The aim of this work is to verify the self-quality assurances in medical institutions in Korea through the external audits by the group of experts and have a mutual discussion of the systematic problems. In order to validate the external audits 30 of 80 medical institutions across the nation were picked out considering the regional distribution and the final 25 institutions applied voluntarily to take part in this work. The basic rules were setup that any information of the participants be kept secrete and the measurements be performed with the dosimetry system already verified through intercomparision. The outputs for 2 or more photon beams, the accuracy of gantry rotation and collimator rotation and the poistional accuracy of MLC movement were measured. The findings for the output measurement showed the differences of -0.8%~4.5%, -0.79%~3.01%, and -0.7%~0.07% with respect to that of the verified dosimetry system for the 6MV, 10MV, and 15MV, respectively. For the reference absorbed dose 8 (16%) of 50 photon beams in 25 medical institutions differed 2.0% or greater from the reference value. The coincidences of Field size with x-ray beam and radiation isocenters of Gantry roration and collimator rotation gave the results of within ${\pm}2$ mm for every institute except 2 institutions. The positional accuracy of MLC movement agreed to within ${\pm}1$ mm for every institute. For the beam qualities of 6 MV photon beams kQ values showed the distribution within 0.4% between maximum and minimum. For the protocols 21 institutions (84%) used absorbed dose to water based protocol while 4 insitutions (16%) used air kerma based one. 22 institutions employed the SSD technique while 3 institutions did the SAD one. External audit plays an important role in discovering the systematic problems of self-performing Quality Assurances and having in depth discussion for mutual complementation. Training experts of international level as well as national support system are required so that both the group of experts of medical physicists and government laboratory could perform together periodical and constant external audits.

The Accuracy Evaluation according to Dose Delivery Interruption and Restart for Volumetric Modulated Arc Therapy (용적변조회전 방사선치료에서 선량전달의 중단 및 재시작에 따른 정확성 평가)

  • Lee, Dong Hyung;Bae, Sun Myung;Kwak, Jung Won;Kang, Tae Young;Back, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.25 no.1
    • /
    • pp.77-85
    • /
    • 2013
  • Purpose: The accurate movement of gantry rotation, collimator and correct application of dose rate are very important to approach the successful performance of Volumetric Modulated Arc Therapy (VMAT), because it is tightly interlocked with a complex treatment plan. The interruption and restart of dose delivery, however, are able to occur on treatment by various factors of a treatment machine and treatment plan. If unexpected problems of a treat machine or a patient interrupt the VMAT, the movement of treatment machine for delivering the remaining dose will be restarted at the start point. In this investigation, We would like to know the effect of interruptions and restart regarding dose delivery at VMAT. Materials and Methods: Treatment plans of 10 patients who had been treated at our center were used to measure and compare the dose distribution of each VMAT after converting to a form of digital image and communications in Medicine (DICOM) with treatment planning system (Eclipse V 10.0, Varian, USA). We selected the 6 MV photon energy of Trilogy (Varian, USA) and used OmniPro I'mRT system (V 1.7b, IBA dosimetry, Germany) to analyze the data that were acquired through this measurement with two types of interruptions four times for each case. The door interlock and the beam-off were used to stop and then to restart the dose delivery of VMAT. The gamma index in OmniPro I'mRT system and T-test in Microsoft Excel 2007 were used to evaluate the result of this investigation. Results: The deviations of average gamma index in cases with door interlock, beam-off and without interruption on VMAT are 0.141, 0.128 and 0.1. The standard deviations of acquired gamma values are 0.099, 0.091, 0.071 and The maximum gamma value in each case is 0.413, 0.379, 0.286, respectively. This analysis has a 95-percent confidence level and the P-value of T-test is under 0.05. Gamma pass rate (3%, 3 mm) is acceptable in all of measurements. Conclusion: As a result, We could make sure that the interruption of this investgation are not enough to seriously affect dose delivery of VMAT by analyzing the measured data. But this investigation did not reflect all cases about interruptions and errors regarding the movement of a gantry rotation, collimator and patient So, We should continuously maintain a treatment machine and program to deliver the accurate dose when we perform the VMAT for the many kinds of cancer patients.

  • PDF