• Title/Summary/Keyword: $SEDIsystem^{TM}$

Search Result 1, Processing Time 0.097 seconds

Evaluation of Automated ESR Measuring System, $SEDIsystem^{TM}$ ($SEDIsystem^{TM}$을 이용한 적혈구 침강속도 측정의 평가)

  • Lee, Jung-Ee;Kim, Kyung-Dong;Lee, Chae-Hoon;Kim, Chung-Sook
    • Journal of Yeungnam Medical Science
    • /
    • v.13 no.1
    • /
    • pp.110-115
    • /
    • 1996
  • The ESR is one of the oldest laboratory test still in use. Although it lacks specificity in diagnosis, it can be effective for monitoring disease activity and following-up. The Westergren method is used for reference method, however coefficient of variation has been described 0.8% to 22.9% according to the literature. Since the ESR was invented in 1921, measurement technique has developed and automated measurement is introduced. We analyzed one hundred forty-three patient samples using $SEDIsystem^{TM}$ automated ESR measuring system and compared with modified Westergren and Wintrobe methods. Comparison between $SEDIsystem^{TM}$ and modified Westergren for ESR measurement yields the following regression equation; y = 0.863x - 1.69 (r=0.830), $SEDIsystem^{TM}$ and Wintrobe y = 1.14x - 14.7 (r=0.789), respectively. We repeated measurement to evaluate reliability, results are not significant in statistically. In conclusion, $SEDIsystem^{TM}$ automated ESR measurement correlated with modified Westergren and Wintrobe methods, reveal reliable results after 4 hours and can report rapidly for large samples. Thus, these results indicate that $SEDIsystem^{TM}$ automated ESR measurement may be useful tool for clinical practice.

  • PDF