초록

Aortic aneurysms, caused by abnormal expansion of the aortic wall, pose critical risks upon rupture. Circular stapler anastomosis offers a faster and safer alternative to suturing. This study evaluates leakage risk and fixation strength in aortic graft circular stapler anastomosis using finite element analysis (FEA). A simplified FEA model was developed and validated using existing leakage pressure data. Key parameters, including inter-row distance, triple-row structure and triple-row spacing were analyzed. The Factor of Safety (FoS) was defined to quantitatively compare leakage risks under varying conditions. The inter-row distance critically influenced leakage risk, with wider gaps increasing leakage likelihood. Conversely, the triple-row structure significantly enhanced fixation strength, reducing leakage risk. Internal spacing within the triple-row exhibited minimal sensitivity. This study establishes a design framework for safer and more effective aortic graft circular stapler anastomosis. Findings provide foundational data for optimizing design and assessing leakage risk.

키워드

과제정보

본 연구는 식품의약품안전처의 규제과학인재양성 사업(R&D)(22183규제학367)으로부터 지원받아 수행되었음.

참고문헌

  1. Evangelista A. Aneurysm of the ascending aorta. Heart. 2010;96(12):979-985. https://doi.org/10.1136/hrt.2008.152751
  2. Mathur A, Mohan V, Ameta D, Bhardwaj G, Haranahalli P. Aortic aneurysm. Journal of Translational Internal Medicine. 2016;4(1):35-41. https://doi.org/10.1515/jtim-2016-0008
  3. Kent KC. Abdominal aortic aneurysms. New England Journal of Medicine. 2014;371(22):2101-2108. https://doi.org/10.1056/NEJMcp1401430
  4. Kouchoukos NT, Kulik A, Castner CF. Open thoracoabdominal aortic repair for chronic type B dissection. The Journal of Thoracic and Cardiovascular Surgery. 2015;149(2):S125-S129. https://doi.org/10.1016/j.jtcvs.2014.07.064
  5. Chalmers J, El-Massri S, Shanmugam G, Naqvi A, Pillai R. A need for speed? Bypass time and outcomes after isolated aortic valve replacement surgery. Interactive Cardiovascular and Thoracic Surgery. 2014;19(1):21-26. https://doi.org/10.1093/icvts/ivu102
  6. Chiesa R, Kahlberg A, Mascia D, Tshomba Y, Civilini E, Melissano G. Use of a novel hybrid vascular graft for sutureless revascularization of the renal arteries during open thoracoabdominal aortic aneurysm repair. Journal of Vascular Surgery. 2014;60(3):622-630. https://doi.org/10.1016/j.jvs.2014.03.256
  7. Martín-Arévalo J, Moro-Valdezate D, Pla-Martí V, García-Botello S, Espí-Macías A, Ortega-Serrano J, Roig-Vila JV, Espí A. Two-row, three-row or powered circular stapler, which to choose when performing colorectal anastomosis? A systematic review and meta-analysis. International Journal of Colorectal Disease. 2024;39(1):51. https://doi.org/10.1007/s00384-024-04625-8
  8. Fang AH, Chao W, Ecker M. Review of colonic anastomotic leakage and prevention methods. Journal of Clinical Medicine. 2020;9(12):4061. https://doi.org/10.3390/jcm9124061
  9. Herzig DO, Ogilvie JW, Chudzinski A, Ferrara A, Ashraf SQ, Jimenez-Rodriguez R, Van der Speeten K, Kinross J, Atallah S. Assessment of a circular powered stapler for creation of anastomosis in left-sided colorectal surgery: a prospective cohort study. International Journal of Surgery. 2020;84:140-146. https://doi.org/10.1016/j.ijsu.2020.11.001
  10. Scardino A, Lucchetti D, Saccomandi P, D'Ovidio V, Staderini F, Posabella A, Basso L, Parello A. Effect of powered circular stapler in colorectal anastomosis after left-sided colic resection: systematic review and meta-analysis. International Journal of Colorectal Disease. 2024;39(1):152. https://doi.org/10.1007/s00384-024-04729-1
  11. Takata M, Inoue Y, Fukada J, Kinoshita Y, Nakamura T. Automatic aortic anastomosis with an innovative computer-controlled circular stapler for surgical treatment of aortic aneurysm. The Journal of Thoracic and Cardiovascular Surgery. 2011;141(5):1265-1269. https://doi.org/10.1016/j.jtcvs.2010.05.048
  12. Idrees JJ, Rosinski MJ, Johnston TT, Eagleton MJ, Svensson LG, Pettersson GB, Blackstone EH. Outcomes after aortic graft-to-graft anastomosis with an automated circular stapler: a novel approach. The Journal of Thoracic and Cardiovascular Surgery. 2016;152(4):1052-1057. https://doi.org/10.1016/j.jtcvs.2016.06.039
  13. Nováček V, Šefl V, Horák M, Konvičková S, Růžička P, Páral J, Kašparová P. Finite element modelling of stapled colorectal end-to-end anastomosis: Advantages of variable height stapler design. Journal of Biomechanics. 2012;45(15):2693-2697. https://doi.org/10.1016/j.jbiomech.2012.07.021
  14. Liao-yuan A, Xing F, Zhou C, Song Z, Luo C. Finite element analysis and experiment on large intestine end-to-end anastomosis. Frontiers in Biomedical Devices. 2018;40789. American Society of Mechanical Engineers. https://doi.org/10.1115/DMD2018-6832
  15. Song P, He Y, Adeloye D, Zhu Y, Ye X, Yi Q, Rahimi K, Rudan I; on behalf of the Global Health Epidemiology Research Group (GHERG). The global and regional prevalence of abdominal aortic aneurysms: a systematic review and modeling analysis. Annals of surgery. 2023;277(6):912-919. https://doi.org/10.1097/SLA.0000000000005716
  16. Baker RS, Foote J, Kemmeter P, Brady R, Vroegop T, Serveld M. The science of stapling and leaks. Obesity Surgery. 2004;14(10):1290-1298. https://doi.org/10.1381/0960892042583888
  17. Strassner H, Fischer A, Weber T, Schmitt K, Lang R, Hoffmann M. Evaluating performance of circular staplers using comparative test methods for evidence-based surgery. Surgical Innovation. 2023;30(5):576-585. https://doi.org/10.1177/15533506231166447
  18. Kang BR, Park A, Kim KT, Kim ST, Lee DH. A study on the usability of the vascular bursting pressure test as a method of performance evaluation of vessel sealing. Journal of Biomedical Engineering Research. 2020;41(5):210-218. https://doi.org/10.9718/JBER.2020.41.5.210
  19. Rusakova YL, Ivanov P, Smirnov A, Malkova E, Efimov N, Chepeleva MV, Markov A. Pigs as models to test cardiovascular devices. Biomedicines. 2024;12(6):1245. https://doi.org/10.3390/biomedicines12061245
  20. de Beaufort HWL, Haulon S, Sobocinski J, Azzaoui R, Gökalp AL, Patterson BO, Karthikesalingam A, van Herwaarden JA, Moll FL, Wanhainen A. Comparative analysis of porcine and human thoracic aortic stiffness. European Journal of Vascular and Endovascular Surgery. 2018;55(4):560-567. https://doi.org/10.1016/j.ejvs.2017.12.014
  21. Jadidi M, Dens J, Shahmansouri N, Zeller T, Goeteyn W. Mechanical and structural changes in human thoracic aortas with age. Acta Biomaterialia. 2020;103:172-188. https://doi.org/10.1016/j.actbio.2019.12.024
  22. Shahbazian N, de Beule M, Mortier P, Wyers MC, Prendergast PJ, Segers P, Verhegghe B. A modeling framework for computational simulations of thoracic endovascular aortic repair. International Journal for Numerical Methods in Biomedical Engineering. 2023;39(4):e3578. https://doi.org/10.1002/cnm.3578
  23. Tremblay D, Cartier R, Pellerin É, Mongrain R, Leask RL. A comparison of mechanical properties of materials used in aortic arch reconstruction. The Annals of Thoracic Surgery. 2009;88(5):1484-1491. https://doi.org/10.1016/j.athoracsur.2009.07.023
  24. Soderberg CR. Factor of safety and working stress. Transactions of the American Society of Mechanical Engineers. 1930;52(2):13-21. https://doi.org/10.1115/1.4057307
  25. Gupta A, Koch P, Liu XC. Evaluate medical device design robustness by combining statistical and probabilistic tools with Finite Element Analysis. Journal of Medical Devices. 2012;6(1):017590. https://doi.org/10.1115/1.4026768