DOI QR코드

DOI QR Code

Comparative efficacy of different video laryngoscopy types in difficult tracheal intubation cases: a randomized crossover manikin study

  • Kyotaro Koshika (Department of Dental Anesthesiology, Tokyo Dental College) ;
  • Wataru Hashimoto (Department of Dental Anesthesiology, Tokyo Dental College) ;
  • Ai Nakakuki (Department of Dental Anesthesiology, Tokyo Dental College) ;
  • Kanako Yajima (Department of Dental Anesthesiology, Tokyo Dental College) ;
  • Tatsuya Ichinohe (Department of Dental Anesthesiology, Tokyo Dental College)
  • Received : 2024.11.05
  • Accepted : 2025.01.11
  • Published : 2025.02.01

Abstract

Background: Video laryngoscopy is beneficial in difficult airway intubation; however, various factors complicate the process. These devices come in different designs, and their usefulness may vary by type. In this study, we compared the effectiveness of several video laryngoscopic .instruments across three simulated difficult intubation scenarios using manikin models. Methods: Training simulators for tracheal intubation were set to four conditions: (i) Normal (mouth opening: 50 mm, normal neck); (ii) Head tilt disorder (mouth opening: 50 mm, rigid neck); (iii) Trismus (mouth opening: 20 mm, normal neck); and (iv) Head tilt disorder + trismus (mouth opening: 20 mm, rigid neck). Seventeen dental anesthesiologists attempted oral tracheal intubation using the following video laryngoscopes: Airway Scope; McGRATH (Normal blade [size 3]); McGRATH (X-blade); and i-view. Evaluated parameters included total intubation time, glottic visualization time, tube induction time, success rate, and difficulty grading of tracheal intubation (Cormack-Lehane classification and the Numerical Rating Scale [NRS]). Statistical analysis was conducted using mixed models, incorporating two-way ANOVA, Tukey's test, two-way ANOVA without repeated measures, and Kruskal-Wallis test, with P < 0.05 deemed statistically significant. Results: Intubation time using i-view was significantly longer for head tilt disorder and trismus compared to other video laryngoscopes (head tilt disorder: P < 0.001 for all, trismus: P = 0.021 vs. Airway Scope, P = 0.028 vs. X-blade). The Cormack-Lehane grade was notably high (P = 0.001) for tracheal intubation with i-view in the head tilt disorder scenario, with intubation failing in three cases. In the combined situation of head tilt disorder and trismus, intubation time with Airway Scope was shorter (P < 0.001 vs. X-blade), achieving a success rate of 100%. However, all attempts with i-view were unsuccessful. The NRS score was significantly higher for i-view compared to the other video laryngoscopes (P < 0.001). Conclusion: Video laryngoscopy effectiveness varies by type in difficult tracheal intubation cases. The Airway Scope or McGRATH instrument appears more suitable for such cases, as indicated by the metrics of intubation time, success rate, and difficulty level.

Keywords

References

  1. Asai T, Shingu K. Difficulty in advancing a tracheal tube over a fibreoptic bronchoscope: incidence, causes and solutions. Br J Anaesth 2004; 92: 870-81. https://doi.org/10.1093/bja/aeh136
  2. Fiadjoe JE, Litman RS. Difficult tracheal intubation: looking to the past to determine the future. Anesthesiology 2012; 116: 1181-2. https://doi.org/10.1097/ALN.0b013e318254d0a0
  3. Vargas M, Servillo G, Buonanno P, Iacovazzo C, Marra A, Putensen-Himmer G, et al. Video vs. direct laryngoscopy for adult surgical and intensive care unit patients requiring tracheal intubation: a systematic review and meta-analysis of randomized controlled trials. Eur Rev Med Pharmacol Sci 2021; 25: 7734-49.
  4. Hansel J, Rogers AM, Lewis SR, Cook TM, Smith AF. Videolaryngoscopy versus direct laryngoscopy for adults undergoing tracheal intubation. Cochrane Database Syst Rev 2022; 4: CD011136. https://doi.org/10.1002/14651858.CD011136.pub3
  5. Kuzmanovska B, Shosholcheva M, Kartalov A, Jovanovski-Srceva M, Gavrilovska-Brzanov A. Survey of current difficult airway management practice. Open Access Maced J Med Sci. 2019; 7: 2775-9. https://doi.org/10.3889/oamjms.2019.673
  6. Sato Boku A, Sobue K, Kako E, Tachi N, Okumura Y, Kanazawa M, et al. The usefulness of the McGrath MAC laryngoscope in comparison with Airway Scope and Macintosh laryngoscope during routine nasotracheal intubation: a randomaized controlled trial. BMC Anesthesiol 2017; 17: 160. https://doi.org/10.1186/s12871-017-0451-y
  7. Suzuki A, Toyama Y, Katsumi N, Kunisawa T, Sasaki R, Hirota K, et al. The Pentax-AWS((R)) rigid indirect video laryngoscope: clinical assessment of performance in 320 cases. Anaesthesia 2008; 63: 641-7. https://doi.org/10.1111/j.1365-2044.2008.05452.x
  8. Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S, et al. Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg 2010; 110: 1049-55. https://doi.org/10.1213/ANE.0b013e3181d2aad7
  9. Moritz A, Leonhardt V, Prottengeier J, Birkholz T, Schmidt J, Irouschek A. Comparison of Glidescope® GoTM, King VisionTM, Dahlhausen VL, I ViewTM and Macintosh laryngoscope use during difficult airway management simulation by experienced and inexperienced emergency medical staff: a randomized crossover manikin study. PLoS One 2020; 15: e0236474. https://doi.org/10.1371/journal.pone.0236474
  10. Phillips RA, Monaghan WP. Incidence of visible and occult blood on laryngoscope blades and handles. AANA J 1997;65: 241-6.
  11. Negri de Sousa AC, Levy CE, Freitas MI. Laryngoscope blades and handles as sources of cross-infection: an integrative review. J Hosp Infect 2013; 83: 269-75. https://doi.org/10.1016/j.jhin.2012.10.015
  12. Benumof JL, Dagg R, Benumof R. Critical hemoglobin desaturation will occur before return to an unparalyzed state following 1 mg/kg intravenous succinylcholine. Anesthesiology 1997; 87: 979-82. https://doi.org/10.1097/00000542-199710000-00034
  13. Ryoo SH, Park KN, Karm MH. The utilization of videolaryngoscopy in nasotracheal intubation for oral and maxillofacial surgical procedures: a narrative review. J Dent Anesth Pain Med 2024; 24: 1-17. https://doi.org/10.17245/jdapm.2024.24.1.1
  14. Dyson A, Harris J, Bhatia K. Rapidity and accuracy of tracheal intubation in a mannequin: comparison of the fibreoptic with the Bullard laryngoscope. Br J Anaesth 1990; 65: 268-70. https://doi.org/10.1093/bja/65.2.268
  15. Ray DC, Billington C, Kearns PK, Kirkbride R, Mackintosh K, Reeve CS, et al. A comparison of McGrath and Macintosh laryngoscopes in novice users: a manikin study. Anaesthesia 2009; 64: 1207-10. https://doi.org/10.1111/j.1365-2044.2009.06061.x