주사기의 마찰 특성이 다단 연결된 시린지펌프의 구동 중 하중부하와 카테터 내부 유체압력에 미치는 영향 분석

Effect of the Frictional Characteristics of Syringe on the Driving Workload and the in-line Hydraulic Pressure of Simultaneously Operating Syringe Pumps

  • 윤성욱 (양산부산대학교병원 의공학과) ;
  • 신대현 (양산부산대학교병원 의공학과) ;
  • 김건호 (양산부산대학교병원 의공학과) ;
  • 남경원 (양산부산대학교병원 의공학과)
  • Sung Uk Yun (Department of Biomedical Engineering, Pusan National University Yangsan Hospital) ;
  • Daehyeon Shin (Department of Biomedical Engineering, Pusan National University Yangsan Hospital) ;
  • Gun Ho Kim (Department of Biomedical Engineering, Pusan National University Yangsan Hospital) ;
  • Kyoung Won Nam (Department of Biomedical Engineering, Pusan National University Yangsan Hospital)
  • 투고 : 2024.10.10
  • 심사 : 2024.10.30
  • 발행 : 2024.12.31

초록

Purpose: Accuracy and stability during drug administration via syringe pumps are vital in intensive care units. However, the effect of frictional characteristic of the pump-equipped syringe and connection protocol of syringe pumps remain elusive. Materials and Methods: Here, these effects, pump arrangement in multiple connection on workload, and hydraulic pressure at the outlet of each syringe pump were evaluated. Results: Statistical analysis demonstrated that the driving workload and hydraulic pressure were substantially affected by variations in the mechanical quality of syringe and arrangement of the connected syringe pumps. Conclusion: Therefore, to increase accuracy and stability, clinics should use high-quality syringes and arrange syringe pumps to minimize positional variations due to different infusion rates in intensive care units.

키워드

과제정보

본 연구는 2023년 양산부산대학교병원 임상연구비 지원으로 이루어졌음.

참고문헌

  1. Hayes C, Jackson D, Davidson PM, Power T. Medication errors in hospitals: a literature review of disruptions to nursing practice during medication administration. J Clin Nurs. 2015;24(21-22):3063-76. https://doi.org/10.1111/jocn.12944
  2. Kim GH, Yun SU, Ro JH, Nam KW. A new injection rate estimation technique for on-site screening test of medication infusion pump by nurses. P I Mech Eng H. 2020;234(4):370-6. https://doi.org/10.1177/0954411919894082
  3. Batliner M, Weiss M, Dual SA, Grass B, Meboldt M, Daners MS. Evaluation of a novel flow-controlled infusion pump for precise and continuous drug delivery at low flow rates: a laboratory study. Anaesthesia. 2019;74(11):1425-31. https://doi.org/10.1111/anae.14784
  4. Yun SU, Nam KW. Experimental study about the effect of the mechanical characteristics of a syringe on the erroneous occlusion alarm generation of syringe pump. J Biomed Eng Res. 2021;42(2):55-61.
  5. Kim JH, Nam KC, Kim H, Nam YG, Kim YJ, Jo EH, Kwon BS. A study on the cognizance of hospital and oriental hospital workers for medical device safety information monitoring. J Biomed Eng Res. 2021;42(3):86-93.
  6. Park SM, Ryu GH. ISO14971:2019 Detailed analysis and periodic safety update report establishment method for the single use medical device – focusing on medical device regulation 2017/745 requirements. J Biomed Eng Res. 2023;44(1):1-10.
  7. Weiss M, Neff T, Gerber A, Fischer J. Impact of infusion line compliance on syringe pump performance. Paediatr Anaesth. 2000;10(6):595-9. https://doi.org/10.1111/j.1460-9592.2000.566ab.x
  8. Kim DW, Steward DJ. The effect of syringe size on the performance of an infusion pump. Paediatr Anaesth. 1999;9(4):335-337. https://doi.org/10.1046/j.1460-9592.1999.00402.x
  9. Schmidt N, Saez C, Seri I, Maturana A. Impact of syringe size on the performance of infusion pumps at low flow rates. Pediatr Crit Care Med. 2010;11(2):282-6. https://doi.org/10.1097/PCC.0b013e3181c31848
  10. Neff SB, Neff TA, Gerber S, Weiss MM. Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps. Eur J Anaesthesiol. 2007;24(7):602-8. https://doi.org/10.1017/S0265021507000026
  11. Weiss M, Fischer J, Neff T, Baenziger O. The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia. 2000;55(11):1094-8. https://doi.org/10.1046/j.1365-2044.2000.01550.x
  12. Kim GH, Hwang YJ, Kim MJ, Nam KW. Real-time monitoring of the actual infusion rate of syringe pump using 2D image marker tracking. J Biomed Eng Res. 2023;44(1):92-98.