DOI QR코드

DOI QR Code

Relative Humidity Transducer Proficiency Test for KOLAS Humidity Calibration Laboratories

KOLAS 교정기관 간 측정 동등성 확립을 위한 상대습도 변환기 숙련도 시험

  • Sang-Wook Lee (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Young-Suk Lee (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Byung-Il Choi (Division of Physical Metrology, Korea Research Institute of Standards and Science)
  • 이상욱 (한국표준과학연구원 물리표준본부) ;
  • 이영석 (한국표준과학연구원 물리표준본부) ;
  • 최병일 (한국표준과학연구원 물리표준본부)
  • Received : 2023.10.16
  • Accepted : 2023.11.14
  • Published : 2023.11.30

Abstract

The Korea Laboratory Accreditation Scheme (KOLAS) operates accreditation programs for ensuring measurement traceability with the International System (SI) of Units - the highest calibration standard that measurements can be tested against. As of September 2023, there are 70 KOLAS-accredited laboratories in the Republic of Korea that specialize in humidity calibration. Among them, 32 KOLAS laboratories, along with one laboratory not affiliated with KOLAS, participated in the proficiency test (PM 2023-11) for relative humidity transducers in 2023. This proficiency test was conducted within a relative humidity range of 20-90% at a temperature of approximately 20 ℃, taking into consideration the calibration and measurement capability (CMC) of the participating laboratories. The primary objective of the proficiency test was to establish the measurement equivalence between each participating laboratory and the reference laboratory, by calculating the number of equivalence (En). When |En| was less than 1, the measurements from the participating and reference laboratory were equivalent. Out of the 33 participating laboratories, 32 successfully met this criterion and passed the proficiency test.

Keywords

Acknowledgement

이 연구는 수탁사업인 숙련도시험 프로그램 시료 개발 사업(IP2023-0005)의 지원을 받아 수행되었다.

References

  1. P. Mecenas, R. T. R. M. Bastos, A. C. R. Vallinoto, and D. Normando, "Effects of temperature and humidity on the spread of COVID-19: A systematic review", PLoS one. Vol. 15, No. 9, p. e0238339, 2020.
  2. H. H. Funke, B. L. Grissom, C. E. McGrew, and M. W. Raynor, "Techniques for the measurement of trace moisture in high-purity electronic specialty gases", Rev. Sci. Instrum., Vol. 74, No. 9, pp. 3909-3933, 2003. https://doi.org/10.1063/1.1597939
  3. J. S. Park, H. Chae, H. K. Chung, and S. I. Lee, "Thin film encapsulation for flexible AM-OLED: a review", Semicond Sci Technol, Vol. 26, No. 3, p. 034001, 2011.
  4. R. M. Syamaladevi, J. Tang, R. Villa-Rojas, S. Sablani, B. Carter, and G. Campbell, "Influence of water activity on thermal resistance of microorganisms in low-moisture foods: a review", Compr. Rev. Food Sci. Food Saf., Vol 15, No. 2, pp 353-70, 2016. https://doi.org/10.1111/1541-4337.12190
  5. C. Cardinali, L. Isaksen, and E. Andersson, "Use and impact of automated aircraft data in a global 4DVAR data assimilation system", Mon. Weather Rev., Vol. 131, No. 8, pp. 1865-1877, 2003. https://doi.org/10.1175//2569.1
  6. R. J. Dirksen, M. Sommer, F. J. Immler, D. F. Hurs, R. Kivi, and H. Vomel, "Reference quality upper-air measurements: GRUAN data processing for the Vaisala RS92 radiosonde", Atmos. Meas. Tech., Vol. 7, No. 12, pp. 4463-4490, 2014. https://doi.org/10.5194/amt-7-4463-2014
  7. World Meteorological Organization (WMO), Guide to Instruments and Methods of Observation-Chapter 4. Measurement of humidity, WMO, Geneva, 2021.
  8. S. W. Lee, B. I. Choi, S. B. Woo, J. C. Kim, and Y. G. Kim, "Calibration of a radiosonde humidity sensor at low temperature and low pressure, Metrologia, Vol. 56, No. 5, pp. 055008, 2019.
  9. International Standard Organization, "General requirements for the competence of testing and calibration laboratories", ISO/IEC 17025, 01 Dec., 2017.
  10. International Standard Organization, "Conformity Assessment-General Requirement for Proficiency Testing", ISO/IEC 17043, 17 Mar., 2010.
  11. Korea Association of Standards & Testing Organizations, "Standard Calibration Procedure of Dew-Point Hygrometers", KASTO 18-50201-370, 14 Dec., 2018.
  12. Korea Association of Standards & Testing Organizations, "Standard Calibration Procedure of Relative Humidity transducers", KASTO 16-50305-408, 30 Nov., 2016.