초록

Background: Dentists bend needles prior to certain injections; however, there are concerns regarding needle fracture, lumen occlusion, and sharps handling. A previous study found that a 30-gauge needle fractures after four to nine 90° bends. This fatigue life study evaluated how many 90° bends a 30-gauge dental needle will sustain before fracture when bent using a needle guide. Methods: Two operators at Element Materials Technology, an independent testing, inspection, and certification company tested 48 30-gauge needles. After applying the needle guide, the operators bent the needle to a 90° angle and expressed the anesthetic from the tip. The needle was then bent back to a 0° angle, and the functionality was tested again. This process was repeated until the anesthetic failed to pass through the end of the needle due to fracture or obstruction. Each operator tested 24 needles (12 needles from each lot), and the number of sustained bends before the needle fracture was recorded. Results: The average number of sustained bends before needle failure was 40.33 (95% confidence interval = 37.41-43.26), with a minimum of 20, median of 40, and a maximum of 54. In each trial, the lumen remained patent until the needle fractured. The difference between the operators was statistically significant (P < 0.001). No significant differences in performance between needle lots were observed (P = 0.504). Conclusion: Our results suggest that using a needle guide increases the number of sustained bends before needle fracture (P < 0.000001) than those reported in previous studies. Future studies should further evaluate the use of needle guides with other needle types across a variety of operators. Furthermore, additional opportunities lie in exploring workplace safety considerations and clinical applications of anesthetic delivery using a bent needle.

키워드

과제정보

Testing was performed at Element Materials Technology, an independent testing, inspection, and certification company. This study was financially supported by Septodont. Septodont provided materials for this study. Septodont and Element Materials Technology expressed permission for the release of the study data and were not involved in writing the report or its submission for publication. Gregory K. Tuttle, a coauthor of this study, holds a patent interest in the TNN Needle Guide. None of the other coauthors have any financial or proprietary interests to disclose.

참고문헌

  1. Malamed SF. Handbook of Local Anesthesia. 7th edition. Amsterdam, Elsevier Health Sciences. 2019, pp 106.
  2. Zimmerman RK. 27-gauge short. J - Acad Gen Dent 1975; 23: 37.
  3. Robison SF, Mayhew RB, Cowan RD, Hawley RJ. Comparative study of deflection characteristics and fragility of 25-, 27-, and 30-gauge short dental needles. J Am Dent Assoc 1984; 109: 920-4. https://doi.org/10.14219/jada.archive.1984.0246
  4. Oikarinen VJ, Perkki K. A metallurgic and bacteriological study of disposable injection needles in dental and oral surgery practice. Proc Finn Dent Soc 1975; 71: 147-61.
  5. Cooley RL, Robison SF. Comparative evaluation of the 30-gauge dental needle. Oral Surg Oral Med Oral Pathol 1979; 48: 400-4. https://doi.org/10.1016/0030-4220(79)90065-3
  6. Monteiro MAO, Antunes ANDG, Basting RT. Physical, chemical, mechanical, and micromorphological characterization of dental needles. J Dent Anesth Pain Med 2021; 21: 139-53. https://doi.org/10.17245/jdapm.2021.21.2.139
  7. Teja KV, Ramesh S, Battineni G, Vasundhara KA, Jose J, Janani K. The effect of various in-vitro and ex-vivo parameters on irrigant flow and apical pressure using manual syringe needle irrigation: Systematic review. Saudi Dent J 2022; 34: 87-99. https://doi.org/10.1016/j.sdentj.2021.12.001
  8. Smith GN, Walton RE, Abbott BJ. Clinical evaluation of periodontal ligament anesthesia using a pressure syringe. J Am Dent Assoc 1983; 107: 953-56. https://doi.org/10.14219/jada.archive.1983.0357