A formalism for the absorbed dose evaluation of the glass dosimeter

  • Ka-Young Park (Department of Physics, Inha University) ;
  • Hyun-Chul Kim (Department of Physics, Inha University) ;
  • Byoung-Chul Kim (Ionizing Radiation Metrology Group, Korea Research Institute of Standards and Science)
  • 투고 : 2023.01.26
  • 심사 : 2023.05.18
  • 발행 : 2023.06.25

초록

We propose in the present work how the reference glass dosimeters can be introduced, which reflects the user irradiation condition. The reference glass dosimeters are used for correcting the reader fluctuation by reading it with sample glass dosimeters at the same time. Since they can be used without annealing after irradiation for long periods, one should consider both the fading effect and the natural background dose accumulation quantitatively. We construct an empirical but practical formalism of evaluating the absorbed dose on the glass dosimeter with the fading effect and the natural background dose accumulation considered.

키워드

과제정보

The present work is supported by Inha University Grant in 2023.

참고문헌

  1. T. Yamamoto, RPL dosimetry: principles and applications, AIP Conf. Proc. 1345 (2011) 217-230.
  2. F. Araki, N. Moribe, T. Shimonobou, Y. Yamashita, Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and Cyber-Knife, Med. Phys. 31 (2004) 1980-1986. https://doi.org/10.1118/1.1758351
  3. K. Son, H. Jung, S.H. Shin, H.-H. Lee, M.-S. Kim, Y.H. Ji, K.B. Kim, Evaluation of the dosimetric characteristics of a radiophotoluminescent glass dosimeter for high-energy photon and electron beams in the field of radiotherapy, Radiat. Meas. 46 (2011) 1117-1122. https://doi.org/10.1016/j.radmeas.2011.08.021
  4. N. Kadoya, K. Shimomura, S. Kitou, Y. Shiota, Y. Fujita, S. Dobashi, K. Takeda, K. Jingu, H. Matsushita, Y. Namito, S. Ban, S. Koyama, K. Tabushi, Dosimetric properties of radiophotoluminescent glass detector in low-energy photon beams, Med. Phys. 39 (2012) 5910-5916. https://doi.org/10.1118/1.4747261
  5. F. Sato, N. Zushi, T. Nagai, T. Tanaka, Y. Kato, T. Yamamoto, T. Iida, Development of radiophotoluminescence glass dosimeter useable in high temperature environment, Radiat. Meas. 53e54 (2013) 8-11. https://doi.org/10.1016/j.radmeas.2013.03.016
  6. J. Perks, M. Gao, V. Smith, S. Skubic, S. Goetsch, Glass rod detectors for small field, stereotactic radiosurgery dosimetric audit, Med. Phys. 32 (2005) 726-732. https://doi.org/10.1118/1.1861155
  7. W.K. Chung, D.W. Kim, Characteristic study of a radio-photoluminescence glass rod detector for clinical usages: skin and inner body in-vivo verification, J. Kor. Phys. Soc. 62 (2013) 670-676. https://doi.org/10.3938/jkps.62.670
  8. Y. Hoshi, T. Nomura, T. Oda, T. Iwasaki, K. Fujita, T. Ishikawa, A. Kato, T. Ikegami, K. Sakai, H. Tanooka, T. Yamada, Application of a newly developed photoluminescence glass dosimeter for measuring the absorbed dose in individual mice exposed to low-dose rate 137Cs γ-rays, J. Radiat. Res. 41 (2000) 129-137. https://doi.org/10.1269/jrr.41.129
  9. Asahi Techno Glass Co., Ltd., RPL Glass dosimeter/Small Element System Basic Characteristic Data, 2000.
  10. AGC Techno Glass Co., Ltd., Glass Dosimetry Reader FGD-1000 Instruction Manual (Ver.2, 2012.
  11. J.-E. Rah, J.-Y. Hong, G.-Y. Kim, Y.-L. Kim, D.-O. Shin, T.-S. Suh, A comparison of the dosimetric characteristics of a glass rod dosimeter and a thermoluminescent dosimeter for mailed dosimeter, Radiat. Meas. 44 (2009) 18-22. https://doi.org/10.1016/j.radmeas.2008.10.010
  12. International Organization for Standardization (ISO), Dosimetry with Radiophotoluminescent Glass Dosimeters for Dosimetry Audit in MV X-Ray Radiotherapy, 2019. ISO 22127.
  13. P.E. Wesolowska, A. Cole, T. Santos, T. Bokulic, P. Kazantsev, J. Izewska, Characterization of three solid state dosimetry systems for use in high energy photon dosimetry audits in radiotherapy, Radiat. Meas. 106 (2017) 556-562. https://doi.org/10.1016/j.radmeas.2017.04.017
  14. IAEA TRS-398, Absorbed Dose Determination in, External Beam Radiotherapy: an International Code of Practice on Dosimetry Based on the Standards of Absorbed Dose to Water, International Atomic Energy Agency, 2000.