Long-Term Extrinsic Uniformity Analysis of Gamma Cameras

감마카메라의 중장기 외인성 균일도 분석

  • Yoon-Jae Kim (Department of Nuclear Medicine, Asan Medical Center) ;
  • Woo-Young Jung (Department of Nuclear Medicine, Asan Medical Center) ;
  • Dong-Hoon Lee (Department of Nuclear Medicine, Asan Medical Center)
  • 김윤재 (서울아산병원 핵의학과) ;
  • 정우영 (서울아산병원 핵의학과) ;
  • 이동훈 (서울아산병원 핵의학과)
  • Published : 2023.05.20

Abstract

Purpose The long-term trend of extrinsic uniformity by type of gamma camera was analyzed, and the factors affecting uniformity were investigated. Based on this, the purpose was to predict the life of the gamma camera, pay attention to factors that affect uniformity, and obtain better quality images. Materials and methods Four of the gamma cameras in operation at a senior general hospital in Seoul were selected and the trend of extrinsic uniformity from the first operation date to the present was analyzed. In order to minimize various factors affecting uniformity, a detailed analysis was conducted by calculating the monthly and annual average of the uniformity values. Results Two Symbia E gamma cameras from SIEMENS, one Symbia Evo Excel gamma camera, and one Symbia Intevo16 gamma camera were selected and analyzed. The uniformity of Symbia E1 (2012 warehousing) changed unevenly, and the uniformity of Symbia E2 (2014 warehousing) changed according to the replacement cycle of 57Co sheet sources. The uniformity of Symbia Evo Excel (received in 2017) and Symbia Intevo 16 (received in 2017) was constant compared to Symbia E. Conclusion The extrinsic uniformity of the gamma camera gradually increased over time. However, there was a difference in uniformity for each type of gamma camera, and there was a change in uniformity in which the cause could not be accurately identified. In order to improve the quality of the image, it is necessary to periodically check changes in uniformity and minimize factors that affect uniformity.

Keywords

References

  1. O'Connor MK. Quality control of scintillation cameras (Planar and SPECT). Eur J Nucl Med. 1999. p. 314-9 
  2. 이광철, 조철우, 장지선, 윤상혁, 이규찬, 신규설, 외2명, 감마카메라의 내인성 장균일도 정도관리 평가, 핵의학기술. 2000;5(2):239-47. 
  3. Randy B, Jody B, Vesper G. Extrinsic versus intrinsic uniformity correction for γ-cameras. J Nucl Med Technol. 2011;39(3):208-12.  https://doi.org/10.2967/jnmt.110.084814
  4. 이종훈, 조성길, 심동오. 57Co 밀봉선원의 노후도에 따른 균일도의 변동과 중첩사용에 관한 유용성 평가. 핵의학기술. 2017;21(1):50-3.